令和 5年10月 6日作成
一般的名称及び販売名
一般的名称: 汎用超音波画像診断装置 販売名 : 汎用超音波画像診断装置 Venue Go
対象ロット、数量及び出荷時期
製造番号: VNU000056、VNU000057、VNU000058、VNU000059、VNU000060、VNU000067、VNU000068 VNU000061、VNU000062、VNU000069、VNU000063、VNU000070、VNU000064、VNU000065 VNU000066、VNU000071、VNU000093、VNU000096、VNU000094、VNU000097、VNU000095 VNU000103、VNU000104、VNU000105、VNU000128、VNU000123、VNU000124、VNU000130 VNU000121、VNU000122、VNU000137、VNU000138、VNU000142、VNU000150、VNU000149 VNU000151、VNU000157、VNU000153、VNU000152、VNU000154、VNU000155、VNU000156 VNU000158、VNU000148、VNU000212、VNU000211、VNU000209、VNU000210、VNU000192 VNU000195、VNU000196、VNU000214、VNU000198、VNU000213、VNU000200、VNU000199 VNU000197、VNU000220、VNU000219、VNU000217、VNU000221、VNU000218、VNU000216 VNU000215、VNU000232、VNU000228、VNU000227、VNU000229、VNU000235、VNU000233 VNU000234、VNU000240、VNU000238、VNU000504、VNU000244、VNU000245、VNU000251 VNU000258、VNU000250、VNU000254、VNU000241、VNU000249、VNU000236、VNU000242 VNU000239、VNU000237、VNU000262、VNU000263、VNU000261、VNU000260、VNU000264 VNU000259、VNU000265、VNU000267、VNU000270、VNU000269、VNU000268、VNU000330 VNU000276、VNU000275、VNU000277、VNU000266、VNU000285、VNU000284、VNU000297 VNU000568、VNU000296、VNU000288、VNU000293、VNU000294、VNU000301、VNU000583 VNU000300、VNU000302、VNU000308、VNU000306、VNU000603、VNU000305、VNU000304 VNU000303、VNU000316、VNU000315、VNU000586、VNU000321、VNU000320、VNU000323 VNU000322、VNU000327、VNU000326、VNU000331、VNU000346、VNU000328、VNU000329 VNU000340、VNU000342、VNU000343、VNU000341、VNU000352、VNU000350、VNU000348 VNU000349、VNU000347、VNU000351、VNU000356、VNU000354、VNU000362、VNU000363 VNU000355、VNU000361、VNU000366、VNU000357、VNU000359、VNU000365、VNU000360 VNU000358、VNU000373、VNU000421、VNU000372、VNU000419、VNU000367、VNU000384 VNU000374、VNU000425、VNU000426、VNU000422、VNU000434、VNU000439、VNU000420 VNU000435、VNU000430、VNU000429、VNU000431、VNU000441、VNU000445、VNU000447 VNU000450、VNU000364、VNU000443、VNU000449、VNU000505、VNU000442、VNU000448 VNU000444、VNU000468、VNU000480、VNU000469、VNU000472、VNU000470、VNU000465 VNU000523、VNU000466、VNU000467、VNU000483、VNU000489、VNU000487、VNU000495 VNU000481、VNU000486、VNU000484、VNU000485、VNU000482、VNU000503、VNU000506 VNU000501、VNU000498、VNU000499、VNU000502、VNU000508、VNU000500、VNU000507 VNU000510、VNU000509、VNU000539、VNU000540、VNU000542、VNU000541、VNU000535 VNU000533、VNU000544、VNU000543、VNU000538、VNU000546、VNU000545、VNU000551 VNU000548、VNU000550、VNU000547、VNU000536、VNU000549、VNU000563、VNU000593 VNU000561、VNU000591、VNU000564、VNU000558、VNU000556、VNU000557、VNU000572 VNU000579、VNU000573、VNU000589、VNU000575、VNU000559、VNU000585、VNU000584 VNU000574、VNU000576、VNU000580、VNU000581、VNU000582、VNU000592、VNU000588 VNU000610、VNU000570、VNU000590、VNU000578、VNU000594、VNU000595、VNU000587 VNU000596、VNU000600、VNU000627、VNU000644、VNU000624、VNU000630、VNU000625 VNU000602、VNU000623、VNU000626、VNU000632、VNU000629、VNU000633、VNU000628 VNU000635、VNU000636、VNU000634、VNU000650、VNU000638、VNU000641、VNU000640 VNU000651、VNU000637、VNU000639、VNU000645、VNU000643、VNU000646、VNU000642 VNU000647、VNU000760、VNU000663、VNU000661、VNU000662、VNU000812、VNU000666 VNU000669、VNU000668、VNU000667、VNU000671、VNU000670、VNU000678、VNU000679 VNU000813、VNU000821、VNU000726、VNU000806、VNU000826、VNU000808、VNU000807 VNU000919、VNU000880、VNU000810、VNU000809、VNU000816、VNU000881、VNU000822 VNU000689、VNU000879、VNU000882、VNU000883、VNU000884、VNU000683、VNU000684 VNU000685、VNU000817、VNU000682、VNU000681、VNU000778、VNU000698、VNU000696 VNU000697、VNU000700、VNU000701、VNU000709、VNU000699、VNU000702、VNU000710 VNU000714、VNU000713、VNU000719、VNU000717、VNU000715、VNU000718、VNU000716 VNU000720、VNU000721、VNU000712、VNU000733、VNU000735、VNU000768、VNU000775 VNU000732、VNU000765、VNU000731、VNU000730、VNU000734、VNU000776、VNU000777 VNU000738、VNU000736、VNU000737、VNU000780、VNU000764、VNU000769、VNU000767 VNU000771、VNU000779、VNU000770、VNU000766、VNU000773、VNU000772、VNU000761 VNU000781、VNU000782、VNU000787、VNU000763、VNU000792、VNU000783、VNU000784 VNU000788、VNU000791、VNU000789、VNU000793、VNU000794、VNU000785、VNU000803 VNU000762、VNU000786、VNU000774、VNU000790、VNU000805、VNU000804、VNU000802 VNU000795、VNU000796、VNU000797、VNU000799、VNU000798、VNU000801、VNU000800 VNU000665、VNU000830、VNU000829、VNU000825、VNU000827、VNU000820、VNU000832 VNU000831、VNU000867、VNU000869、VNU000874、VNU000834、VNU000876、VNU000863 VNU000871、VNU000864、VNU000868、VNU000870、VNU000872、VNU000865、VNU000873 VNU000866、VNU000877、VNU000890、VNU000878、VNU000889、VNU000903、VNU000904 VNU000888、VNU000912、VNU000902、VNU000897、VNU000885、VNU000905、VNU000911 VNU000908、VNU000913、VNU000906、VNU000909、VNU000914、VNU000910、VNU000907 VNU000916、VNU000915、VNU000918、VNU000923、VNU000924、VNU000920、VNU000928 VNU000927、VNU000926、VNU000922、VNU000921、VNU000941、VNU000937、VNU000932 VNU000936、VNU000934、VNU000931、VNU000944、VNU000938、VNU000943、VNU000939 VNU000933、VNU000935、VNU000940、VNU000942、VNU001024、VNU001025、VNU001043 VNU000958、VNU000957、VNU000961、VNU000959、VNU000965、VNU000956、VNU000969 VNU000979、VNU000972、VNU000966、VNU000973、VNU000970、VNU000974、VNU000976 VNU000971、VNU000981、VNU000997、VNU000980、VNU000990、VNU000987、VNU000984 VNU000991、VNU000982、VNU000986、VNU000989、VNU000983、VNU000985、VNU001007 VNU000996、VNU001008、VNU001001、VNU001002、VNU001000、VNU001009、VNU001006 VNU000995、VNU001005、VNU001003、VNU001022、VNU001004、VNU000998、VNU001010 VNU000999、VNU001023、VNU001027、VNU001029、VNU001032、VNU001028、VNU001031 VNU001030、VNU001026、VNU001033、VNU001034、VNU001036、VNU001038、VNU001037 VNU001039、VNU001035、VNU001041、VNU001040、VNU001042、VNU001050、VNU001044 VNU001049、VNU001047、VNU001045、VNU001046、VNU001048、VNU001051、VNU001056 VNU001058、VNU001054、VNU001057、VNU001053、VNU001055、VNU001052、VNU000992 VNU001076、VNU001080、VNU001079、VNU001078、VNU001077、VNU001082、VNU001083 VNU001085、VNU001081、VNU001084、VNU001124、VNU001112、VNU001118、VNU001123 VNU001113、VNU001108、VNU001107、VNU001132、VNU001119、VNU001109、VNU001128 VNU001129、VNU001120、VNU001131、VNU001127、VNU001130、VNU001135、VNU001134 VNU001136、VNU001133、VNU001144、VNU001148、VNU001147、VNU001141、VNU001145 VNU001143、VNU001146、VNU001142、VNU001166、VNU001154、VNU001168、VNU001167 VNU001152、VNU001153、VNU001150、VNU001163、VNU001162、VNU001169、VNU001187 VNU001180、VNU001179、VNU001177、VNU001173、VNU001176、VNU001178、VNU001175 VNU001174、VNU001188、VNU001181、VNU001195、VNU001189、VNU001186、VNU001185 VNU001190、VNU001183、VNU001182、VNU001184、VNU001208、VNU001202、VNU001203 VNU000988 数量:617台 出荷時期 : 令和 元年12月〜令和 5年 3月
製造販売業者等名称
製造販売業者の名称 : GEヘルスケア・ジャパン株式会社 製造販売業者の所在地: 東京都日野市旭が丘4‐7‐127 許可の種類 : 第一種医療機器製造販売業 許可番号 : 13B1X00150 輸入先製造業者 : GE Medical Systems Israel Ltd.
改修理由
製造元からの報告により、当該装置の固定台付きカートにおいて、チルト機構の不具合により、装置がカートか ら落下する可能性のあることが分かりました。このため、製造元の情報を基に、お客様に情報提供を行うことで 注意喚起を実施し、問題が修正されたチルト機構の部品に交換する改修作業を実施します。
危惧される具体的な健康被害
本事象が装置使用中に発生し、落下した装置が患者または使用者と接触した場合、打撲などの怪我につながるお それがあります。しかしながら、装置の使用に際しては、医療従事者により常に監視されており、異常が発生し た場合においても直ちに適切な処置をとれることから、本事象により患者様へ重篤な健康被害が発生することは 無いと考えております。なお、国内において本事象による健康被害の発生は報告されておりません。
改修開始年月日
令和5年10月6日
効能・効果又は用途等
超音波を用いて体内の形状、性状、又は動態を可視化し、画像情報を診断のために提供する装置です。
その他
納入いたしました医療機関は弊社が把握しておりますので、連絡の上、改修を実施します。