一般的名称及び販売名
一般的名称: (1)〜(5)据置型デジタル式汎用X線診断装置,(6),(7)天井取り付け式X線管支持器 販売名 : (1)デジタル式X線一般撮影システム RADspeed safire (2)診断用X線装置 RADspeed Pro (3)デジタル式X線一般撮影システム DR-XD100 (4)診断用X線装置 DR-XD101 (5)診断用X線装置 FUJIFILM DR BENEO-Fx (6)天井走行式X線管懸垂器 CH-200 (7)天井走行式X線管懸垂器 CH-200M
対象ロット、数量及び出荷時期
(1)デジタル式X線一般撮影システム RADspeed safire 製造番号:011B715701,071B717101,071B720201,071B720301,071B720401,071B720501,071M225101,081B736401 081B736501,081B736601,081B740801,081C111701,081K486401,081K486501,081K486601,081K486701 081K486801,081K486901,081K716201,081M234601,081M234701,081Q683301,081S121101,091B806301 091B829502,091B829602,091B829702,091B829802,091H746001,091H778601,091K595501,091K899001 101B855202,101B856203,101B862003,101H799301,101K885901,101K886001,101K886101,101M300203 101Q698304,101R752201,40D0BE728001,40D0BE741001,40D0BED0C001,40D0BED12001,40D0BF123001 出荷時期:平成19年5月 〜 平成26年1月 出荷台数:47台 (2)診断用X線装置 RADspeed Pro 製造番号:101A050106,101A063307,101A069206,101A081309,101A082208,101A084601,101A087701,101A087801 101A089501,101A090301,101A093201,101A093607,101A094701,101A095801,101A097401,101A098401 101A102401,101A102601,101A105901,101A108901,101A111801,101A112001,101A112501,101A114801 101A118301,101A121801,101A122201,101A124201,101A124801,101A126001,101A127901,101A128701 101A132301,101A132801,101A133901,101B863228,101B870001,101B872601,101B872701,101B872801 101B873001,101B873501,101B875101,101C176101,101C176201,101C177201,101C177301,101C178401 101C178701,101C179801,101C179901,101C180701,101C180801,101C181001,101C182501,101C183201 101F338301,101F338501,101F343401,101G294001,101H001017,101H001114,101H009001,101H009201 101H009701,101H010801,101H011401,101K014701,101K015601,101K152301,101K255501,101K255601 101K309801,101K309901,101K887501,101L696210,101L697101,101L697401,101L698101,101M303501 101M303601,101O687401,101O689401,101P094301,101P096101,101Q709701,101Q709801,101Q711101 101Q712101,101R755201,101S154401,101S154501,101S154601,101S646801,101S646901,101S647001 410560113001,410560311001,410560313001,410560318001,410560318002,410560321001,410560321003 410560334001,410560334002,410560341001,410560341002,410560351001,410560351002,410560354001 410560438001,410560443001,410561013001,410561014001,410561111001,410561111002,410561111003 410561111004,410561112001,410561112002,410561112003,410561112004,410561112005,410561214001 410561214002,410561214003,410561214004,410561214005,410561215001,410561215002,410561315001 410561316001,410561316002,410561911001,410561911002,410561911003,410561911004,410561911005 410561912001,410561912002,410561912003,410561912004,410561912005,410561912006,410561913001 410561913002,410561913003,410561913004,410561913005,410561913006,410561913007,410561913008 410561914001,410561914002,410561914003,410561914004,410561914005,410561914007,410561915001 410561915002,410561915003,410561915004,410561915005,410561916001,410561916002,410562112001 410562112002,410562115001,410562116001,410562116002,410562211001,410562213001,410562213003 410562214001,410562214002,410562215001,410562218001,410562323001,410562423001,410562426001 410562429001,410562434001,410562441001,418030042001,41055F10B001,41055F10B003,41055F10B004 41055F10B005,41055F10B006,41055F10C001,41055F10C002,41055F111001,41055F111002,41055F111003 41055F112001,41055F112002,41055F112003,41055F112004,41055F112005,41055F114001,41055F20B001 41055F20B002,41055F20B003,41055F20C002,41055F211001,41055F211002,41055F211003,41055F211004 41055F212001,41055F212002,41055F212003,41055F212004,41055F212005,41055F212006,41055F213001 41055F213003,41055F213004,41055F213005,41055F216001,41055F216002,41055F50B001,41055F50C001 41055F513001,41055F514002,41055F514003,41055F515001,41055F70C001,41055F70C002,41055F70C003 41055F712002,41055F712003,41055F713002,41055F713003,41055F71A002,41055F71B001,41055F71B002 41055F71C001,41055F71C002,41055F724002,41055F726001,41055F726002,41055F726003,41055F726004 41055F72A001,41055F72A002,41055F733001,41055F733002,41055F734001,41055F734002,41055F734003 41055F73B001,41055F73B002,41055F73C001,41055F73C002,41055F742001,41055F742002,41055F743001 41055F743002,41055F743003,41055F743004,41055F749001,41055F749002,41055F74A001,41055F74A002 41055F74B001,41055F74B002,41055F751001,41055F751002,41055F755001,41055F755002,41055F758001 41055F759001,41055F817001,41055F819001,41055F823001,41055F827001,41055F829001,41055F847001 41056010A001,41056030B001,41056031B001,41056031B002,410560321002,41056034B001,41056034B002 41056040C001,41056044C001,410560F0A001,410560F0C001,410560F0C002,410560F0C003,410560F0C004 410560F13001,410560F14001,41056100A001,41056110A001,41056110C001,41056110C002,41056120B001 41056120C001,41056140B001,41056140C001,41056190A001,41056190A002,41056190B001,41056190B002 41056190B003,41056190B004,41056190B005,41056190C001,41056190C002,41056190C003,41056190C004 41056190C005,41056190C006,41056190C007,410561A0A001,410561A0A002,410561A0A003,410561A0B001 410561A0B002,410561A0B003,410561A0B004,410561A0C001,410561A0C002,410561A0C003,410561A11001 410561A12001,410561A12002,410561A13001,410561A13002,410561A13003,410561A13004,410561A13005 410561A13007,410561A15001,410561A16001,410561A18001,410561B0B001,410561B0C001,410561B11001 410561B13001,410561B13003,410561B15001,410561B16001,410561B16002,410561B16003,410561B16004 410561B17001,410561B18001,410561B1A001,410561B1A002,410561B1C001,410561B22001,410561B22002 410561B22003,410561B23001,410561B24001,410561B24002,410561B25001,410561B26001,410561B26002 410561B27001,410561B27002,410561B27003,410561B28001,410561B29001,410561B2A001,410561B2A002 410561B2A003,410561B2C001,410561B31001,410561B32001,410561B32002,410561B33001,410561B33002 410561B33003,410561B35001,410561B35002,410561B36001,410561B37001,410561B39001,410561B3B001 410561B3B002,410561B3C001,410561B41001,410561B44001,410561B47001,410561B47002,410561B49001 410561B4B001,410561B4C001,410561B51001,410561B51002,410561B52001,410561B52002,410561B53001 410561B53002,410561B59001,410561B59002,410561B59003,410561C12001,410561C12002,410561C12003 410561C13001,410561C14001,410561C15001,410561C16001,410561C18001,410561C1B001,410561C1C001 410561C21001,410561C22001,410561C23001,410561C24001,410561C24002,410561C25001,410561C26001 410561C26002,410561C28001,410561C28002,410561C29001,410561C29002,410561C29003,410561C31001 410561C32001,410561C33001,410561C33002,410561C33003,410561C34001,410561C34002,410561C36001 410561C36002,410561C3B001,410561C41001,410561C42001,410561C43001,410561C43002,410561C45001 410561C46001,410561C48001,410561C4B001,410561C53001,410561C53002,410561C53003,410561C53004 410561C55001,410561C58001,410561C59001,41056210C001,41056210C002,41056220A001,41056220C001 41056220C002,41056220C003,41056220C004,41056232A001,41056232C001,41056232C002,41056241A001 41056242A001,410562D52001,410562D52002,410562D53001,410562D57001,410562D58001,410562E45001 410562E52001,410562E55001,410562E59001,41802C917001,41802C917002,41802C918001,41802C918002 41802C919003,41802C919004,41802C922001,41802C922002,41802C922003,41802C922004,41802C922005 41802C922006,41802C922007,41802C923001,41802C923002,41802C923003,41802C926001,41802C926002 41802C926003,41802C929001,41802C929002,41802C929003,41802C92A001,41802C92A002,41802C92B001 41802C931001,41802C931002,41802C931003,41802C931004,41802C931005,41802C931006,41802C932001 41802C932002,41802C932003,41802C932004,41802C932005,41802C933001,41802C933002,41802C933003 41802C933004,41802C933005,41802C933006,41802C933007,41802C934001,41802C934002,41802C934003 41802C934004,41802C934005,41802C937001,41802C938001,41802C938002,41802C938003,41802C938004 41802C938005,41802C939001,41802C939002,41802C939003,41802C939004,41802C939005,41802C939006 41802C93A001,41802C93A002,41802C93B001,41802C93B002,41802C93B003,41802C93C001,41802C93C002 41802C93C003,41802C93C004,41802C93C005,41802C93C006,41802C941001,41802C941002,41802C941003 41802C941004,41802C942001,41802C942002,41802C942003,41802C942004,41802C942005,41802C942006 41802C942007,41802C942008,41802C942009,41802C942010,41802C943001,41802C943002,41802C943003 41802C943004,41802C943005,41802C943006,41802C943007,41802C945001,41802C945002,41802C945003 41802C946001,41802C946002,41802C946003,41802C947001,41802C948001,41802C948002,41802C948003 41802C948004,41802C948005,41802C948006,41802C948007,41802C948008,41802C948009,41802C948010 41802C948011,41802C949001,41802C949002,41802C949003,41802C949004,41802C94A001,41802C94A002 41802C94A003,41802C94A004,41802C94A005,41802C94A006,41802C94A007,41802C94B002,41802C94B003 41802C94B004,41802C94C001,41802C94C002,41802C94C003,41802C94C004,41802C94C005,41802C94C006 41802C951001,41802C951002,41802C951003,41802C951004,41802C951005,41802C951006,41802C951007 41802C952001,41802C952002,41802C952003,41802C952004,41802C952005,41802C952006,41802C952007 41802C952008,41802C952009,41802C952010,41802C952011,41802C952012,41802C952014,41802C952015 41802C953001,41802C953002,41802C953003,41802C953004,41802C953005,41802C953006,41802C953008 41802C953009,41802C953010,41802C954001,41802C954002,41802C954003,41802C954004,41802C955001 41802C955002,41802C955003,41802C955004,41802C956001,41802C957001,41802C957003,41802C957004 41802C958001,41802C958002,41802C958003,41802C958004,41802C958005,41802C959001,41802C959002 41802C959003,41802C959005,41802C959006,41802CA16002,41802CA17001,41802CA17002,41802CA17003 41802CA18002,41802CA18003,41802CA19001,41802CA19002,41802CA19003,41802CA19004,41802CA19005 41802CA1A001,41802CA1A002,41802CA1B001,41802CA1B003,41802CA1C001,41802CA1C002,41802CA1C003 41802CA1C004,41802CA21001,41802CA22001,41802CA23001,41802CA23002,41802CA23003,41802CA23004 41802CA24001,41802CA24002,41802CA25001,41802CA25002,41802CA25003,41802CA25004,41802CA26001 41802CA26002,41802CA26003,41802CA26004,41802CA26005,41802CA26006,41802CA27001,41802CA29001 41802CA29002,41802CA29003,41802CA29004,41802CA2C001,41802CA2C002,41802CA2C003,41802CA31001 41802CA31002,41802CA31003,41802CA32001,41802CA32002,41802CA32003,41802CA32004,41802CA33001 41802CA33002,41802CA33003,41802CA33004,41802CA33005,41802CA33006,41802CA33007,41802CA33008 41802CA33009,41802CA33010,41802CA34001,41802CA34002,41802CA34003,41802CA34004,41802CA36001 41802CA36002,41802CA36003,41802CA37001,41802CA37002,41802CA3B001,41802CA3B002,41802CA3C001 41802CA3C002,41802CA3C003,41802CA3C004,41802CA3C005,41802CA3C006,41802CA3C007,41802CA3C008 41802CA3C009,41802CA3C010,41802CA3C011,41802CA3C012,41802CA41001,41802CA41002,41802CA41003 41802CA41004,41802CA42001,41802CA42002,41802CA42003,41802CA42004,41802CA42005,41802CA43001 41802CA43002,41802CA43003,41802CA43004,41802CA43005,41802CA43006,41802CA44001,41802CA44002 41802CA44003,41802CA45001,41802CA45002,41802CA45003,41802CA47001,41802CA47002,41802CA49001 41802CA49002,41802CA49003,41802CA49004,41802CA49005,41802CA49006,41802CA49007,41802CA4A001 41802CA4A002,41802CA4A003,41802CA4A004,41802CA4A005,41802CA4A006,41802CA4C001,41802CA4C002 41802CA4C003,41802CA4C004,41802CA4C005,41802CA4C006,41802CA4C007,41802CA4C008,41802CA51001 41802CA51002,41802CA51003,41802CA52001,41802CA52002,41802CA52003,41802CA53001,41802CA55001 41802CA56001,41802CA57001,41802CA59001,41802CA59002,41802CA59003,41802CA59004,41802CA59005 41802CA59006,41802CA59007,41802CA59008,41802CA59009,41802CA59010,41802CA59011,41802CA59012 41802CA59013,41802CB1A001,41802CB23001,41802CB41001,41802CB42001,41802CC22001,41802CC22002 41802CC34001,41802CD17001,41802CD18001,41802CD18002,41802CD19001,41802CD1A001,41802CD1A003 41802CD1A004,41802CD1A005,41802CD1B001,41802CD1B002,41802CD1B003,41802CD1B005,41802CD22001 41802CD24001,41802CD2C001,41802CD2C002,41802CD2C003,41802CD31001,41802CD31002,41802CD33001 41802CD34001,41802CD3C001,41802CD3C002,41802CD3C003,41802CD41001,41802CD43001,41802CD43002 41802CD44001,41802CD46001,41802CD48001,41802CD59001,41802CE33001,41802CE39001,41802CE3A001 41802CE43001,41802CE4B001,41802CE53002,41802CE53003,41802CE58001,41802D817001,41802D819001 41802D819002,41802D819003,41802D819004,41802D81A001,41802D821001,41802D823001,41802D823002 41802D826001,41802D829001,41802D829002,41802D82B001,41802D82C001,41802D833001,41802D837001 41802D838001,41802D838002,41802D838003,41802D839001,41802D839002,41802D839003,41802D83C001 41802D841001,41802D846001,41802D849001,41802D852001,41802D852002,41802D853001,41802D853002 41802D854001,41802D854002,41802D854004,41802D854005,41802D856001,41802D856002,41802D856003 41802D918001,41802D91A001,41802D91B001,41802D921001,41802D932001,41802D934002,41802D935001 41802D937001,41802D939001,41802D93A001,41802D942002,41802D946001,41802D949001,41802D959001 41802DA17001,41802DA18001,41802DA19001,41802DA1C001,41802DA37001,41802DA3B001,41802DA3B002 41802DA42001,41802DA42002,41802E719001,41802E719002,41802E71A001,41802E71A002,41802E71B001 41802E71B002,41802E71C001,41802E71C002,41802E721001,41802E721002,41802E722001,41802E722002 41802E722003,41802E723001,41802E723002,41802E723003,41802E723004,41802E726001,41802E726002 41802E726003,41802E727001,41802E727002,41802E728001,41802E728002,41802E729001,41802E729002 41802E729003,41802E729004,41802E72B001,41802E72B002,41802E731001,41802E731002,41802E731003 41802E731004,41802E731005,41802E732001,41802E732002,41802E733001,41802E73A001,41802E73A002 41802E742001,41802E742002,41802E742003,41802E742004,41802E742005,41802E742006,41802E742007 41802E743001,41802E743002,41802E743003,41802E743004,41802E743005,41802E743006,41802E743007 41802E743008,41802E744001,41802E744002,41802E744003,41802E744004,41802E744005,41802E747001 41802E74C001,41802E753001,41802E753002,41802E753003,41802E755001,41802E755002,41802E755003 41802E756001,41802E758001,41802E758002,41802E758003,41802E817001,41802E817002,41802E819001 41802E81A001,41802E81B001,41802E81C001,41802E81C002,41802E81C003,41802E822001,41802E822002 41802E826001,41802E826002,41802E827001,41802E828001,41802E828002,41802E829001,41802E829002 41802E82B001,41802E82B002,41802E82C001,41802E82C002,41802E82C003,41802E82C004,41802E82C005 41802E82C006,41802E831001,41802E832001,41802E832002,41802E832003,41802E832004,41802E832005 41802E832006,41802E832007,41802E832008,41802E832009,41802E832010,41802E833001,41802E833002 41802E839001,41802E839002,41802E83C001,41802E83C002,41802E841001,41802E842001,41802E842002 41802E842003,41802E842004,41802E842005,41802E842006,41802E842007,41802E842008,41802E842009 41802E843001,41802E843002,41802E843003,41802E843004,41802E843005,41802E843006,41802E843007 41802E843008,41802E847001,41802E847002,41802E847003,41802E847004,41802E847005,41802E847006 41802E847007,41802E847008,41802E847009,41802E847010,41802E847011,41802E849001,41802E849002 41802E849003,41802E84B001,41802E84B002,41802E84B003,41802E84C001,41802E84C002,41802E84C003 41802E851001,41802E852001,41802E852002,41802E852003,41802E852004,41802E852005,41802E853001 41802E853002,41802E853003,41802E853004,41802E853005,41802E856001,41802E859001,41802E859002 41802E917001,41802E917002,41802E917003,41802E91A001,41802E91B001,41802E91B002,41802E921001 41802E921002,41802E921003,41802E921004,41802E921005,41802E921006,41802E922001,41802E922002 41802E922003,41802E922004,41802E922005,41802E922006,41802E923001,41802E923002,41802E923003 41802E923004,41802E923005,41802E923006,41802E924001,41802E924002,41802E924003,41802E927001 41802E92B001,41802E92B002,41802E931001,41802E931002,41802E931003,41802E931004,41802E931005 41802E931006,41802E932001,41802E932002,41802E932003,41802E933001,41802E939001,41802E93B001 41802E941001,41802E941002,41802E942001,41802E942002,41802E942003,41802E943001,41802E945001 41802E945002,41802E946001,41802E948001,41802E948002,41802E94B001,41802E94B002,41802E94B003 41802E94B004,41802E94B005,41802E951001,41802E951002,41802E951003,41802E953001,41802E954001 41802E958001,41802EA17001,41802EA1A001,41802EA1B001,41802EA1C001,41802EA1C002,41802EA1C003 41802EA21001,41802EA22001,41802EA22002,41802EA23001,41802EA23002,41802EA23003,41802EA23004 41802EA23005,41802EA23006,41802EA26001,41802EA26002,41802EA28001,41802EA28002,41802EA29001 41802EA29002,41802EA29003,41802EA29004,41802EA2C001,41802EA32001,41802EA33001,41802EA38001 41802EA39001,41802EA39002,41802EA39003,41802EA39004,41802EA3B001,41802EA41001,41802EA41002 41802EA42001,41802EA43001,41802EA43002,41802EA43003,41802EA43004,41802EA43005,41802EA43006 41802EA47001,41802EA48001,41802EA49001,41802EA49002,41802EA49003,41802EA4A001,41802EA4A002 41802EA4C001,41802EA4C002,41802EA51001,41802EA58001,41802EA58002,41802EA58003,41802EA58004 41802EA59001,41802EA59002,41802EB17001,41802EB17002,41802EB1C001,41802EB22001,41802EB23001 41802EB28001,41802EB2A001,41802EB2A002,41802EB2B001,41802EB33001,41802EB34001,41802EB37001 41802EB37002,41802EB3B001,41802EB42001,41802EB43001,41802EB44001,41802EB49001,41802EB4A001 41802EB53001,41802EB59001,41802EC17001,41802EC19001,41802EC19002,41802EC19003,41802EC1B001 41802EC2B001,41802EC2C001,41802EC33001,41802EC34001,41802EC39001,41802EC42001,41802EC42002 41802EC49001,41802EC4B001,41802EC4B002,41802EC4C001,41802EC54001,41802EC54002,41802EC54003 41802EC55001,41802EC59001,41802F117001,41802F117002,41802F117003,41802F117004,41802F117005 41802F117006,41802F118001,41802F119001,41802F119002,41802F119003,41802F119004,41802F119005 41802F119006,41802F11A001,41802F11A002,41802F11A003,41802F11B001,41802F11C001,41802F11C002 41802F11C003,41802F11C004,41802F11C005,41802F11C006,41802F121001,41802F121002,41802F121003 41802F121004,41802F121005,41802F121006,41802F122001,41802F122002,41802F122003,41802F122004 41802F122005,41802F123001,41802F123002,41802F123003,41802F123004,41802F124001,41802F124002 41802F124003,41802F124004,41802F124005,41802F124006,41802F124007,41802F125001,41802F125002 41802F125003,41802F125004,41802F125005,41802F126001,41802F126002,41802F126003,41802F126004 41802F126005,41802F127001,41802F127002,41802F127003,41802F128001,41802F128002,41802F128003 41802F128004,41802F128005,41802F129001,41802F129002,41802F129003,41802F12A001,41802F12A002 41802F12A003,41802F12A004,41802F12A005,41802F12A006,41802F12C001,41802F12C002,41802F12C003 41802F131001,41802F131002,41802F131003,41802F131004,41802F131005,41802F132001,41802F132002 41802F132003,41802F132004,41802F133001,41802F133002,41802F133003,41802F133004,41802F133005 41802F133006,41802F134001,41802F134002,41802F135001,41802F135002,41802F135003,41802F135004 41802F136001,41802F136002,41802F138001,41802F138002,41802F138003,41802F138004,41802F138005 41802F138006,41802F139001,41802F139002,41802F139003,41802F139004,41802F13A001,41802F13A002 41802F13A003,41802F13A004,41802F13B001,41802F13B002,41802F13B003,41802F13B004,41802F13B005 41802F13B006,41802F13C001,41802F13C002,41802F13C003,41802F141001,41802F141002,41802F142001 41802F142002,41802F142003,41802F142004,41802F143001,41802F143002,41802F143003,41802F143004 41802F143005,41802F143006,41802F143007,41802F143008,41802F144001,41802F144002,41802F144003 41802F144004,41802F145001,41802F145002,41802F145003,41802F145004,41802F147001,41802F147002 41802F14A001,41802F14A002,41802F14A003,41802F14A004,41802F14C001,41802F14C002,41802F151001 41802F151002,41802F152001,41802F152002,41802F152003,41802F153001,41802F153002,41802F153003 41802F153004,41802F153005,41802F153006,41802F153007,41802F153008,41802F154001,41802F154002 41802F154003,41802F154004,41802F154005,41802F155001,41802F155002,41802F158001,41802F158002 41802F158003,41802F158004,41802F159001,41802F159002,41802F159003,41802F217001,41802F217002 41802F218001,41802F218002,41802F218003,41802F218004,41802F218005,41802F219001,41802F219002 41802F219003,41802F219004,41802F219005,41802F21A001,41802F21A002,41802F21C001,41802F21C002 41802F221001,41802F221002,41802F221003,41802F222001,41802F222002,41802F222003,41802F222004 41802F222005,41802F222006,41802F222007,41802F222008,41802F222009,41802F223001,41802F223002 41802F223003,41802F224001,41802F224002,41802F224003,41802F225001,41802F225002,41802F225003 41802F226001,41802F226002,41802F226003,41802F227001,41802F227002,41802F227003,41802F227004 41802F228001,41802F228002,41802F228003,41802F228004,41802F228005,41802F228006,41802F229001 41802F231001,41802F231002,41802F231003,41802F231004,41802F231005,41802F231006,41802F232001 41802F232002,41802F232003,41802F233001,41802F233002,41802F233003,41802F233004,41802F234001 41802F234002,41802F234003,41802F235001,41802F235002,41802F235003,41802F235004,41802F235005 41802F237001,41802F237002,41802F238001,41802F238002,41802F238003,41802F238004,41802F238005 41802F238006,41802F238007,41802F238008,41802F238009,41802F238010,41802F238011,41802F238012 41802F239001,41802F239002,41802F239003,41802F239004,41802F239005,41802F239006,41802F23A001 41802F23A002,41802F23A003,41802F23A004,41802F23B001,41802F23B002,41802F23B003,41802F23B004 41802F23B005,41802F23B006,41802F23C001,41802F23C002,41802F241001,41802F241002,41802F242001 41802F242002,41802F242003,41802F242004,41802F243001,41802F243002,41802F243003,41802F243004 41802F243005,41802F243006,41802F243007,41802F243008,41802F243009,41802F243010,41802F244001 41802F244002,41802F245001,41802F245002,41802F247001,41802F247002,41802F247003,41802F247004 41802F247005,41802F247006,41802F248001,41802F248002,41802F248003,41802F248004,41802F249001 41802F249002,41802F24A001,41802F24A002,41802F24A003,41802F24A004,41802F24A005,41802F24A006 41802F24A007,41802F24A008,41802F24B001,41802F24B002,41802F24C001,41802F24C002,41802F251001 41802F251002,41802F252001,41802F252002,41802F253001,41802F253002,41802F253003,41802F253004 41802F253005,41802F253006,41802F253007,41802F253008,41802F253009,41802F253010,41802F255001 41802F255002,41802F255003,41802F255004,41802F256001,41802F256002,41802F257001,41802F257002 41802F257003,41802F257004,41802F258001,41802F258002,41802F259001,41802F259002,41802F259003 41802F259004,41802F259006,41802F259007,41802F917001,41802F918001,41802F919001,41802F919002 41802F921001,41802F922001,41802F923001,41802F924001,41802F928001,41802F928002,41802F929001 41802F92A001,41802F92C001,41802F92C002,41802F932001,41802F935001,41802F935002,41802F936001 41802F937001,41802F938001,41802F93C001,41802F942001,41802F942002,41802F942003,41802FA19001 41802FA19002,41802FA19003,41802FA19004,41802FA1C001,41802FA21001,41802FA22001,41802FA22002 41802FA22003,41802FA22004,41802FA23001,41802FA23002,41802FA24001,41802FA28001,41802FA2B001 41802FA2C001,41802FA31001,41802FA31002,41802FA32001,41802FA32002,41802FA33001,41802FA34001 41802FA34002,41802FA34003,41802FA35001,41802FA35002,41802FA36001,41802FA36002,41802FA37001 41802FA37002,41802FA38001,41802FA38002,41802FA38003,41802FA39001,41802FA39002,41802FA41001 41802FA41002,41802FA42001,41802FA42002,41803003A001,418030A47001,418030A53001,418030A56001 418030A59001,418030A59002,418030B43001,418030B43002,418030B43003,418030B43004,418030B44001 418030B44002,418030B45001,418030B46001,418030B47001,418030B47002,418030B47003,418030B47004 418030B48001,418030B49001,418030B49002,418030B49003,418030B4A001,418030B4A002,418030B4A003 418030B4B001,418030B4C001,418030B4C002,418030B4C003,418030B51001,418030B51002,418030B52001 418030B53001,418030B53002,418030B53003,418030B54001,418030B54002,418030B54003,418030B55001 418030B55002,418030B55003,418030B56001,418030B58001,418030B58002,418030B58003,418030B59001 418030B59002,418030B59003,418030B59004,418030B59005,41B853943001,41B853943002,41B853943003 41B853949001,41B853A49001,41E1D4961001,41E1D4961002,41E1D4962001,41E1D4963001,41E1D4963002 41E1D4963003,41E1D4963004,41E1D4963005,41E1D4965001,41E1D4965002,41E1D4966001,41E1D4966002 41E1D4966003,41E1D4968001,41E1D4969001,41E1D4969002,41E1D496A001,41E1D496A002,41E1D496C001 41E1D496C002,41E1D4971001,41E1D4975001,41E1D4979001,41E1D497A001,41E1D4983001,41E1D4983002 41E1D4983003,41E1D4983004,41E1D4986001,41E1D4987001,41E1D4989001,41E1D4989002,41E1D4989003 41E1D498B001,41E1D498B002,41E1D498B003,41E1D4991001,41E1D4994001,41E1D4A5A001,41E1D4A5B001 41E1D4A5B002,41E1D4A5C001,41E1D4A5C002,41E1D4A5C003,41E1D4A61001,41E1D4A62001,41E1D4A62002 41E1D4A63001,41E1D4A63002,41E1D4A63003,41E1D4A63004,41E1D4A63005,41E1D4A64001,41E1D4A6C001 41E1D4A71001,41E1D4A72001,41E1D4A72002,41E1D4A73001,41E1D4A76001,41E1D4A79001,41E1D4A7C001 41E1D4A7C002,41E1D4A7C003,41E1D4A7C004,41E1D4A7C005,41E1D4A81001,41E1D4A81002,41E1D4A84002 41E1D4A84003,41E1D4A84004,41E1D4A84005,41E1D4A84006,41E1D4A85001,41E1D4A87001,41E1D4A88001 41E1D4A88002,41E1D4A89001,41E1D4A8C001,41E1D4A8C002,41E1D4A91001,41E1D4A92001,41E1D4A92002 41E1D4A93001,41E1D4A94001,41E1D4A94002,41E1D535C001,41E1D5361001,41E1D5361002,41E1D5362001 41E1D5363001,41E1D5365001,41E1D5366001,41E1D536B001,41E1D5371001,41E1D5373001,41E1D5373002 41E1D5373003,41E1D5379001,41E1D537A001,41E1D537A002,41E1D5381001,41E1D5382001,41E1D5382002 41E1D5388001,41E1D5389001,41E1D5389002,41E1D538A001,41E1D538A002,41E1D538B001,41E1D538B002 41E1D538B003,41E1D538B004,41E1D538C001,41E1D5391001,41E1D5391002,41E1D5391003,41E1D5392001 41E1D5392002,41E1D5392003,41E1D5393001,41E1D5396001,41E1D5396002,41E1D5462001,41E1D5463001 41E1D546B001,41E1D5472001,41E1D5472002,41E1D5479001,41E1D547C001,41E1D547C002,41E1D547C003 41E1D5482001,41E1D5482002,41E1D5482003,41E1D5482004,41E1D5483001,41E1D548C001,41E1D5492001 41E1D5493001,41E1D5496001,41E1D5497001,41E1D5497002,41E1D5D5A001,41E1D5D5A002,41E1D5D5A003 41E1D5D5B001,41E1D5D5B002,41E1D5D5C001,41E1D5D5C002,41E1D5D5C003,41E1D5D61001,41E1D5D61002 41E1D5D61003,41E1D5D61004,41E1D5D61005,41E1D5D61006,41E1D5D61007,41E1D5D61008,41E1D5D61009 41E1D5D62001,41E1D5D62002,41E1D5D62003,41E1D5D62004,41E1D5D67001,41E1D5D68001,41E1D5D68002 41E1D5D69001,41E1D5D69002,41E1D5D72001,41E1D5D72002,41E1D5D72003,41E1D5D72004,41E1D5D72005 41E1D5D72006,41E1D5D73001,41E1D5D7B001,41E1D5D7B002,41E1D5D7C001,41E1D5D82001,41E1D5D82002 41E1D5D82003,41E1D5D82004,41E1D5D82005,41E1D5D83001,41E1D5D83002,41E1D5D87001,41E1D5D87002 41E1D5D87003,41E1D5D8C001,41E1D5D8C002,41E1D5D91001,41E1D5D92001,41E1D5D93001,41E1D5D93002 41E1D5D94001,41E1D5D94002,41E1D5D94003,41E1D5E5B001,41E1D5E5B002,41E1D5E5C001,41E1D5E5C002 41E1D5E5C003,41E1D5E5C004,41E1D5E63001,41E1D5E6B001,41E1D5E6B002,41E1D5E6C001,41E1D5E6C002 41E1D5E6C003,41E1D5E72001,41E1D5E73001,41E1D5E78001,41E1D5E7A001,41E1D5E81001,41E1D5E82001 41E1D5E82002,41E1D5E89001,41E1D5E89002,41E1D5E89003,41E1D5E93001,41E1D5F5A001,41E1D5F5B001 41E1D5F62001,41E1D5F63001,41E1D5F68001,41E1D5F6B001,41E1D5F6B002,41E1D5F6B003,41E1D5F71001 41E1D5F71002,41E1D5F71003,41E1D5F71004,41E1D5F73001,41E1D5F73002,41E1D5F77001,41E1D5F77002 41E1D5F78001,41E1D5F7C001,41E1D5F84001,41E1D5F85001,41E1D5F86001,41E1D5F87001,41E1D5F89001 41E1D5F89002,41E1D5F8B001,41E1D5F92001,41E1D5F93001,41E1D5F94001,41E1D5F94002,41E1D606A001 41E1D606A002,41E1D606B001,41E1D606B002,41E1D606C001,41E1D607C001,41E1D6081001,41E1D6083001 41E1D6083002,41E1D6089001,41E1D6093001,41E1D6269001,41E1D6269002,41E1D626A001,41E1D6271001 41E1D6271002,41E1D6272001,41E1D6272002,41E1D6272003,41E1D6272004,41E1D6273001,41E1D6275001 41E1D627B001,41E1D6282001,41E1D6282002,41E1D6282003,41E1D6282004,41E1D6283001,41E1D6283002 41E1D6284001,41E1D6284002,41E1D6284003,41E1D6289001,41E1D6289002,41E1D628A001,41E1D628A002 41E1D6291001,41E1D6291002,41E1D6291003,41E1D6294001,41E1D6296001,41E1D6296002,41E1D6296003 41E1D6373001,41E1D6375001,41E1D6375002,41E1D6377001,41E1D6378001,41E1D6378002,41E1D6378003 41E1D6381001,41E1D638A001,41E1D638A002,41E1D638A003,41E1D638C001,41E1D6391001,41E1D6391002 41E1D6391003,41E1D6392001,41E1D6392002,41E1D6392003,41E1D6468001,41E1D6468002,41E1D6468003 41E1D6478001,41E1D6479001,41E1D6479002,41E1D647C001,41E1D6482001,41E1D6482002,41E1D6482003 41E1D6482004,41E1D6482005,41E1D6483001,41E1D6486001,41E1D6488001,41E1D6491001,41E1D6492001 41E1D6493001,41E1D6493002,41E1D656B001,41E1D656C001,41E1D6571001,41E1D6574001,41E1D6574002 41E1D657A001,41E1D657C001,41E1D658B001,41E1D658B002,41E1D658C001,41E1D675A001,41E1D675A002 41E1D675B001,41E1D675B002,41E1D6761001,41E1D6761002,41E1D6761003,41E1D6763001,41E1D6763002 41E1D6763003,41E1D6769001,41E1D6769002,41E1D6769003,41E1D676C001,41E1D6771001,41E1D6771002 41E1D6771003,41E1D6771004,41E1D6772001,41E1D6772002,41E1D6772003,41E1D6774001,41E1D6777001 41E1D6779001,41E1D6779002,41E1D6781001,41E1D6785001,41E1D6789001,41E1D6789002,41E1D6789003 41E1D6789004,41E1D6789005,41E1D678C001,41E1D678C002,41E1D678C003,41E1D6791001,41E1D6791002 41E1D6791003,41E1D6791004,41E1D6792001,41E1D6792002,41E1D6792003,41E1D6792004,41E1D6796001 41E1D6796002,41E1D6796003,41E1D685B001,41E1D685B002,41E1D685C001,41E1D6862001,41E1D6862002 41E1D6862003,41E1D6863001,41E1D6863002,41E1D6863003,41E1D6863004,41E1D6868001,41E1D6868002 41E1D6869001,41E1D6869002,41E1D686B001,41E1D686B002,41E1D6871001,41E1D6873001,41E1D6873002 41E1D6873003,41E1D6874001,41E1D6876001,41E1D6878001,41E1D6878002,41E1D6878003,41E1D6879001 41E1D6879002,41E1D687B001,41E1D687C001,41E1D687C002,41E1D687C003,41E1D6882001,41E1D6882002 41E1D6882003,41E1D6882004,41E1D6882005,41E1D6883001,41E1D6883002,41E1D6883003,41E1D6883004 41E1D6887001,41E1D6887002,41E1D6887003,41E1D6887004,41E1D6887005,41E1D6887006,41E1D6887007 41E1D6888001,41E1D6888002,41E1D6888003,41E1D6888004,41E1D6889001,41E1D6889002,41E1D688A001 41E1D688A002,41E1D688A003,41E1D688B001,41E1D688C001,41E1D688C002,41E1D688C003,41E1D688C004 41E1D688C005,41E1D688C006,41E1D688C007,41E1D6891001,41E1D6891002,41E1D6891003,41E1D6891004 41E1D6891005,41E1D6892001,41E1D6893001,41E1D6894001,41E1D6894002,41E1D6896001,41E1D695A001 41E1D695B001,41E1D695C001,41E1D695C002,41E1D695C003,41E1D6962001,41E1D6962002,41E1D6963001 41E1D6963002,41E1D6963003,41E1D6963004,41E1D6963005,41E1D6963006,41E1D6967001,41E1D6967002 41E1D6967003,41E1D6968001,41E1D6968002,41E1D6968003,41E1D6969001,41E1D696C001,41E1D696C002 41E1D6973001,41E1D6973002,41E1D6974001,41E1D6976001,41E1D6976002,41E1D6976003,41E1D6976004 41E1D6977001,41E1D6979001,41E1D697B001,41E1D697C001,41E1D697C002,41E1D6982001,41E1D6985001 41E1D6986001,41E1D6986002,41E1D6987001,41E1D6989001,41E1D6989002,41E1D6991001,41E1D6991002 41E1D6992001,41E1D6992002,41E1D6993001,41E1D6993002,41E1D6993003,41E1D6994001,41E1D6997001 41E1D6A62001,41E1D6A62002,41E1D6A62003,41E1D6A64001,41E1D6A64002,41E1D6A66001,41E1D6A6B001 41E1D6A6B002,41E1D6A6B003,41E1D6A6C001,41E1D6A6C002,41E1D6A6C003,41E1D6A6C004,41E1D6A71001 41E1D6A72001,41E1D6A72002,41E1D6A73001,41E1D6A73002,41E1D6A73003,41E1D6A73004,41E1D6A73005 41E1D6A73006,41E1D6A73007,41E1D6A73008,41E1D6A74001,41E1D6A74002,41E1D6A74003,41E1D6A77001 41E1D6A79001,41E1D6A79002,41E1D6A7B001,41E1D6A7C001,41E1D6A7C002,41E1D6A81001,41E1D6A81002 41E1D6A81003,41E1D6A82001,41E1D6A82002,41E1D6A82003,41E1D6A87001,41E1D6A89001,41E1D6A89002 41E1D6A8C001,41E1D6A8C002,41E1D6A91001,41E1D6A91002,41E1D6A92001,41E1D6A92002,41E1D6A92003 41E1D6A92004,41E1D6A93001,41E1D6A93002,41E1D6A94001,41E1D6A94002,41E1D6A94003,41E1D6A96001 41E1D6A97001,41E1D6A97002,41E1D6A97003,41E21315A001,41E21315A002,41E21315B001,41E21315C001 41E213161001,41E213163001,41E213163002,41E213163003,41E213163004,41E213163005,41E213163006 41E213163007,41E213169001,41E213169002,41E213169004,41E213172001,41E213174001,41E213175001 41E213175002,41E213175003,41E213178001,41E213179001,41E213183001,41E213183002,41E213183003 41E213184001,41E213184002,41E213187001,41E213189001,41E21318A001,41E21318B001,41E21318C001 41E213191001,41E213193001,41E213194001,41E213194002,41E213194003,41E213259001,41E21325A001 41E21325A002,41E21325A003,41E21325A004,41E21325B001,41E21325C001,41E21325C002,41E21325C003 41E21325C004,41E213261001,41E213262001,41E213262002,41E213262003,41E213263001,41E213263002 41E213263003,41E213263004,41E213264001,41E213264002,41E213268001,41E213269001,41E21326A001 41E21326C001,41E21326C002,41E213271001,41E213271002,41E213272001,41E213273001,41E213273002 41E213273003,41E213273004,41E213276001,41E21327A001,41E21327A002,41E21327A003,41E21327A004 41E21327A005,41E21327B001,41E21327C001,41E21327C002,41E21327C003,41E213281001,41E213282001 41E213282002,41E213282003,41E213283001,41E213283002,41E213284001,41E213287001,41E213287002 41E213288001,41E213288002,41E213288003,41E213289001,41E213289002,41E213289003,41E213289004 41E213289005,41E21328A001,41E21328B001,41E21328B002,41E21328B003,41E21328C001,41E21328C002 41E213291001,41E213292001,41E213293001,41E213293002,41E213293003,41E213293004,41E213294001 41E213294002,41E213297001,41E213371001,41E21337B001,41E213385001,41E21338B001,41E21338B002 41E213392001,41E213494001,41E21355C001,41E21355C002,41E213561001,41E213561002,41E213561003 41E213562001,41E213563001,41E213563002,41E213563003,41E213564002,41E213565001,41E213566001 41E213568001,41E213571001,41E213578001,41E213579001,41E21357B001,41E21357C001,41E21357C002 41E213581001,41E213583001,41E213584001,41E213585001,41E21358C001,41E213591001,41E213591002 41E213593001,41E213593002,41E213669001,41E21366C001,41E213691001,41E214064001,41E214064002 41E214064003,41E214068001,41E214071001,41E214072001,41E214072002,41E214074001,41E214078001 41E214079001,41E214079002,41E214079003,41E21407A001,41E214081002,41E214084001,41E214084002 41E214085001,41E214087001,41E214089001,41E21408A001,41E21408B001,41E214093001,41E214094001 41E214094002,41E214094003,41E214094004,41E214094005,41E21415A001,41E21415C001,41E214162001 41E214163001,41E214166001,41E214167001,41E214178001,41E214179001,41E214182001,41E214182002 41E214186001,41E214186002,41E214189001,41E214267001,41E214268001,41E21426B001,41E21428C001 41E214297001,41E21535A001,41E215362001,41E215363001,41E215369001,41E21536B001,41E21536B002 41E215373001,41E215374001,41E215378001,41E215385001,41E215386001,41E215391001,41E215391002 41E215461001,41E215461002,41E215462001,41E215462002,41E215462003,41E215469001,41E215469002 41E21546A001,41E21546B001,41E215473001,41E215478001,41E215481001,41E215482001,41E215482002 41E215483001,41E215484001,41E21548A001,41E21548B001,41E215493001,41E215494001,41E215495001 41E21595A001,41E21595A002,41E21595B001,41E21595B002,41E21595B003,41E21595B004,41E21595C001 41E21595C002,41E21595C003,41E21595C004,41E21595C005,41E21595C006,41E215961001,41E215961002 41E215961003,41E215961004,41E215962001,41E215962002,41E215962003,41E215962004,41E215962005 41E215962006,41E215964001,41E215964002,41E215964003,41E215964004,41E215964005,41E215964006 41E215965001,41E215965002,41E215965003,41E215965004,41E215966001,41E215966002,41E215966003 41E215966004,41E215967001,41E215967002,41E215967003,41E215967004,41E215969001,41E215969002 41E215969003,41E21596B001,41E215973001,41E215973002,41E215974001,41E215974002,41E215975001 41E215975002,41E215975003,41E215975004,41E215976001,41E215976002,41E215976003,41E215976004 41E215976005,41E215976006,41E215977001,41E215977002,41E215977003,41E215977004,41E215977005 41E215977006,41E215977007,41E215978001,41E215978002,41E215978003,41E215978004,41E215978005 41E215978006,41E215978007,41E215978008,41E215978009,41E215978010,41E21597A001,41E21597A002 41E21597A003,41E21597A004,41E21597B001,41E21597B002,41E21597B003,41E215982001,41E215983001 41E215984001,41E215984002,41E215985001,41E215985002,41E215985003,41E215985004,41E215985005 41E215986001,41E215986002,41E215986003,41E215986004,41E215987001,41E215987002,41E215988001 41E215989001,41E215989002,41E215989003,41E21598A001,41E21598B001,41E21598B002,41E21598B003 41E21598C001,41E215991001,41E215991002,41E215991003,41E215992001,41E215992002,41E215992003 41E215992004,41E215992005,41E215992006,41E215992007,41E215993001,41E215993002,41E215993003 41E215993004,41E215993005,41E215994001,41E215994002,41E215995001,41E215996001,41E215996002 41E215996003,41E215A5A001,41E215A5A002,41E215A5A003,41E215A5A004,41E215A5B001,41E215A5B002 41E215A5B003,41E215A5B004,41E215A5C001,41E215A5C002,41E215A5C003,41E215A5C004,41E215A61001 41E215A61002,41E215A61003,41E215A61004,41E215A61005,41E215A61006,41E215A62001,41E215A62002 41E215A62003,41E215A62004,41E215A62005,41E215A62006,41E215A62007,41E215A62008,41E215A63001 41E215A63002,41E215A63003,41E215A63004,41E215A63005,41E215A63006,41E215A63007,41E215A63008 41E215A63009,41E215A64001,41E215A64002,41E215A64003,41E215A64004,41E215A64005,41E215A64006 41E215A64007,41E215A64008,41E215A64009,41E215A64010,41E215A64011,41E215A64012,41E215A64013 41E215A64014,41E215A65001,41E215A65002,41E215A65003,41E215A65004,41E215A65005,41E215A65006 41E215A65007,41E215A66001,41E215A66002,41E215A66003,41E215A66004,41E215A66005,41E215A66006 41E215A67001,41E215A67002,41E215A67003,41E215A67004,41E215A67005,41E215A67006,41E215A68001 41E215A68002,41E215A68003,41E215A68004,41E215A69001,41E215A69002,41E215A69003,41E215A69004 41E215A69005,41E215A69006,41E215A69007,41E215A69008,41E215A69009,41E215A6A001,41E215A6A002 41E215A6A003,41E215A6A004,41E215A6B001,41E215A6B002,41E215A6B003,41E215A6B004,41E215A6C001 41E215A6C002,41E215A6C003,41E215A6C004,41E215A6C005,41E215A6C006,41E215A6C007,41E215A6C008 41E215A71001,41E215A71002,41E215A71003,41E215A71004,41E215A71005,41E215A71006,41E215A72001 41E215A72002,41E215A72003,41E215A72004,41E215A72005,41E215A72006,41E215A72007,41E215A73001 41E215A73002,41E215A73003,41E215A73004,41E215A73005,41E215A73006,41E215A74001,41E215A74002 41E215A74003,41E215A74004,41E215A74005,41E215A74006,41E215A74007,41E215A74008,41E215A75001 41E215A75002,41E215A75003,41E215A75004,41E215A75005,41E215A75006,41E215A76001,41E215A76002 41E215A76003,41E215A76004,41E215A76005,41E215A76006,41E215A77001,41E215A77002,41E215A77003 41E215A77004,41E215A77005,41E215A77006,41E215A77007,41E215A77008,41E215A78001,41E215A78002 41E215A78003,41E215A78004,41E215A78005,41E215A78006,41E215A79001,41E215A79002,41E215A7A001 41E215A7A002,41E215A7A003,41E215A7A004,41E215A7B001,41E215A7B002,41E215A7B003,41E215A7B004 41E215A7B005,41E215A7B006,41E215A7C001,41E215A7C002,41E215A7C003,41E215A7C004,41E215A7C005 41E215A7C006,41E215A81001,41E215A81002,41E215A81003,41E215A81004,41E215A81005,41E215A81006 41E215A82001,41E215A82002,41E215A82003,41E215A82004,41E215A82005,41E215A82006,41E215A82007 41E215A83001,41E215A83002,41E215A83003,41E215A84001,41E215A84002,41E215A84003,41E215A84004 41E215A85001,41E215A85002,41E215A85003,41E215A85004,41E215A85005,41E215A85006,41E215A86001 41E215A86002,41E215A86003,41E215A86004,41E215A87001,41E215A87002,41E215A87003,41E215A87004 41E215A88001,41E215A88002,41E215A88003,41E215A88004,41E215A88005,41E215A89001,41E215A89002 41E215A8A001,41E215A8A002,41E215A8A003,41E215A8A004,41E215A8B001,41E215A8B002,41E215A8B003 41E215A8B004,41E215A8B005,41E215A8B006,41E215A8C001,41E215A8C002,41E215A8C003,41E215A8C004 41E215A91001,41E215A91002,41E215A91003,41E215A91004,41E215A91005,41E215A92001,41E215A92002 41E215A92003,41E215A92004,41E215A92005,41E215A92006,41E215A92007,41E215A92008,41E215A92009 41E215A92010,41E215A92011,41E215A92012,41E215A92013,41E215A92014,41E215A92015,41E215A93001 41E215A93002,41E215A93003,41E215A93004,41E215A93005,41E215A93006,41E215A93007,41E215A93008 41E215A93009,41E215A93010,41E215A93011,41E215A93012,41E215A93013,41E215A94001,41E215A94002 41E215A94003,41E215A94004,41E215A94005,41E215A94006,41E215A94007,41E215A94008,41E215A94009 41E215A95001,41E215A95002,41E215A95003,41E215A95004,41E215A95005,41E215A95006,41E215A96001 41E215A96002,41E215A96003,41E215A96004,41E215A96005,41E215A96008,41E215A97007,41E215A97008 41E216977001,41E216977002,41E216977003,41E21697B001,41E21697B002,41E21697C001,41E216983001 41E216984001,41E216984002,41E216985001,41E216985002,41E216987001,41E216987002,41E216988001 41E216988002,41E216988003,41E216988004,41E216989001,41E216989002,41E216989003,41E21698A001 41E21698A002,41E21698A003,41E21698A004,41E21698A005,41E21698B001,41E21698B002,41E21698B003 41E21698C001,41E21698C002,41E21698C003,41E216991001,41E216991002,41E216991003,41E216992001 41E216993001,41E216993002,41E216993003,41E216993004,41E216993005,41E216994001,41E216994002 41E216994003,41E216994004,41E216995001,41E216995002,41E216995003,41E216995004,41E216996001 41E216996002,41E217269001,41E217273001,41E217278001,41E21727A001,41E21727C001,41E217283001 41E217289001,41E217294001,41E217294002,41E217363001,41E217363002,41E217363003,41E217363004 41E217364001,41E217364002,41E217364003,41E217364004,41E217364005,41E217364006,41E217364007 41E217365001,41E217367001,41E217367002,41E217367003,41E217368001,41E217369001,41E217369002 41E217369003,41E217369004,41E217369005,41E21736A001,41E21736A002,41E21736C001,41E217371001 41E217371002,41E217371003,41E217372001,41E217372002,41E217372003,41E217373001,41E217373002 41E217373003,41E217373004,41E217373005,41E217373006,41E217373007,41E217373008,41E217374001 41E217374002,41E217374003,41E217374004,41E217374005,41E217375001,41E217375002,41E217375003 41E217375004,41E217376001,41E217377001,41E217377002,41E217377003,41E217378001,41E217378002 41E217378003,41E217379001,41E217379002,41E217379003,41E217379004,41E21737C001,41E21737C002 41E21737C003,41E217381001,41E217382001,41E217382002,41E217383001,41E217383002,41E217383003 41E217383004,41E217383005,41E217383006,41E217383007,41E217384001,41E217384002,41E217384003 41E217386001,41E217387001,41E217387002,41E21738B001,41E217394001,41E217394002,41E251F5B001 41E251F5C002,41E251F61001,41E251F61002,41E251F6C001,41E251F6C002,41E251F72001,41E251F72002 41E251F73001,41E251F73002,41E251F74001,41E251F84001,41E251F86001,41E251F87001,41E251F88001 41E251F8A001,41E251F8A002,41E251F8C001,41E251F92002,41E251F92004,41E251F92005,41E251F93001 41E251F94002,41E25205B001,41E252B5A001,41E252B74001,41E252B7A001,41E252B81001,41E252C93001 41E25435B001,41E254361002,41E254363001,41E254364001,41E254364002,41E254364003,41E254366001 41E254367001,41E254368001,41E254368002,41E25436A001,41E25436A002,41E25436B001,41E25436B002 41E25436C001,41E254372001,41E254374001,41E254374002,41E254376001,41E254376002,41E254377001 41E254377002,41E25437C001,41E25437C002,41E254382001,41E254384001,41E254385001,41E254387001 41E25438A001,41E25438B001,41E25438B002,41E25438C001,41E254391001,41E254391002,41E254394001 41E254395001,41E254395002,41E25445A001,41E25445B001,41E254461001,41E254462001,41E254463001 41E254463002,41E254463003,41E254464001,41E254466001,41E254468001,41E254469001,41E254469002 41E25446A001,41E25446B001,41E254471001,41E254474001,41E254479001,41E254479002,41E25447B001 41E254481001,41E254482001,41E254482002,41E254483001,41E254483002,41E254484001,41E254484002 41E254485001,41E254486001,41E254488001,41E25448B001,41E25448B002,41E254491001,41E254492002 41E254492003,41E254493001,41E254493002,41E255677001,41E25567B001,41E255686001,41E255687001 41E255689001,41E255692001, 41E1D4998001(*),41E1D4999001(*),41E1D4999002(*),41E1D4A97001(*),41E1D4A98001(*) 41E1D4A99001(*),41E1D4A99002(*),41E1D4A99003(*),41E1D4A99004(*),41E1D4A99006(*) 41E1D5399001(*),41E1D5399002(*),41E1D5399003(*),41E1D5498001(*),41E1D5499001(*) 41E1D5499002(*),41E1D5D98001(*),41E1D5F99002(*),41E1D6298001(*),41E1D6298002(*) 41E1D6399001(*),41E1D6597001(*),41E1D6599001(*),41E1D6599003(*),41E1D6599004(*) 41E1D6599005(*),41E1D6798001(*),41E1D6798002(*),41E1D6898001(*),41E1D6898002(*) 41E1D6A97004(*),41E1D6A97005(*),41E1D6A97006(*),41E1D6A98001(*),41E1D6A98002(*) 41E1D6A98003(*),41E1D6A98004(*),41E1D6A98005(*),41E1D6A98006(*),41E1D6A99002(*) 41E1D6A99003(*),41E213199001(*),41E213199002(*),41E213298001(*),41E213298002(*) 41E213298003(*),41E213298004(*),41E213298005(*),41E213298006(*),41E213298007(*) 41E213298008(*),41E213298009(*),41E213598001(*),41E213599001(*),41E213599002(*) 41E214098001(*),41E214199001(*),41E215498001(*),41E215498002(*),41E215499002(*) 41E215997001(*),41E215997002(*),41E215997003(*),41E215997004(*),41E215997005(*) 41E215998001(*),41E215998002(*),41E215999001(*),41E215999002(*),41E215999003(*) 41E215999005(*),41E215A96006(*),41E215A96007(*),41E215A96009(*),41E215A96010(*) 41E215A97001(*),41E215A97002(*),41E215A97003(*),41E215A97004(*),41E215A97005(*) 41E215A97006(*),41E215A98001(*),41E215A98002(*),41E215A98003(*),41E215A98005(*) 41E215A98006(*),41E215A99001(*),41E215A99002(*),41E216996003(*),41E216996004(*) 41E216997001(*),41E216998001(*),41E216998002(*),41E216999001(*),41E216999002(*) 41E251F98002(*),41E251F99001(*),41E252B99001(*),41E254397001(*),41E254398001(*) 41E254399001(*),41E254399002(*),41E254399003(*),41E254399004(*),41E254499001(*) 41E255697001(*) 出荷時期:平成22年5月 〜 令和元年9月(*) 出荷台数:3,270台(*) (3)デジタル式X線一般撮影システム DR-XD 100 製造番号:071B300101,071BB30801,071BB30901,081BB31201,081BB31301,081BB31601,081BB31901,081BB32001 081BB32101,081BB32201,081BB34401,081BB34501,081BB37301,081BB37401,091BB39101,091BB39201 091BB39301,091BB39401,091BB40501,091BB40601,091BB42201,091BB42901,091BB50303,091BB51503 091BB51603,101BB53203,101BB55303,101BB56903,40D59F80C001,40D59F811001,40D59F813001 40D59F813002,40D59F813003,40D59F813004,40D59F813005,40D59F813006,40D59F816001,40D59F817001 40D59F818001,40D59F81B001,40D59F81B002,40D59F81C001,40D59F81C002,40D59F81C003,40D59F81C004 40D59F821001,40D59F821002,40D59F821003,40D59F822001,40D59F822002,40D59F822003,40D59F823001 40D59F823002,40D59F823003,40D59F823004,40D59F825001,40D59F825002,40D59F825003,40D59F825004 40D59F826001,40D59F828001,40D59F82A001,40D59F82A002,40D59F82A003,40D59F82B001,40D59F82C001 40D59F82C002,40D59F832001,40D59F832002,40D59F832003,40D59F833001,40D59F833002,40D59F841001 40D59F841002,40D59F842001,40D59F842002,40D59F842003,40D59F847001,40D59F84A001,40D59F857001 出荷時期:平成19年9月 〜 平成27年7月 出荷台数:80台 (4)診断用X線装置 DR-XD 101 製造番号:416FD252A001,416FD2531001,416FD2531002,416FD2532001,416FD2532002,416FD2533001,416FD2541001 416FD2669001,416FD2673001,416FD267B001,416FD2687001,416FD268B001,416FD268B002,416FD268B003 416FD268B004,416FD2691001,416FD2692001,416FD2694001,416FD2694002,416FD2F29001,416FD2F3B001 416FD3947001,416FD3959001,416FD3A6B001,416FD3A71001,416FD3A75001,416FD3A7C001,416FD3A82001 416FD3A88001,416FD3A88002,416FD3A88003,416FD3A92001,416FD3A97001,416FD3A97002,416FD3A97003 出荷時期:平成24年8月 〜 令和元年8月 出荷台数:35台 (5)診断用X線装置 FUJIFILM DR BENEO-Fx 製造番号:MP95A8A52001,MP95A8A52002,MP95A8A52003,MP95A8A53001,MP95A8A53002,MP95A8A53003,MP95A8A53004 MP95A8A53005,MP95A8C59001,MP95A935C001,MP95A9368001,MP95A9368002,MP95A936C001,MP95A936C002 MP95A9371001,MP95A9372001,MP95A9375001,MP95A9375002,MP95A937C001,MP95A937C002,MP95A9381001 MP95A9381002,MP95A9382001,MP95A9382002,MP95A9383001,MP95A9388001,MP95A938C001,MP95A938C002 MP95A938C003,MP95A9392001,MP95A9393001,MP95A9394001,MP95A9396001,MP95A9397001,MP95A9462001 MP95A9463001,MP95A9467001,MP95A9468001,MP95A9468002,MP95A9468003,MP95A9469001,MP95A946B001 MP95A946C001,MP95A9472001,MP95A9472002,MP95A9472003,MP95A9472004,MP95A9473001,MP95A947B001 MP95A947C001,MP95A9482001,MP95A9482002,MP95A9482003,MP95A9482004,MP95A9483001,MP95A9483002 MP95A9483003,MP95A9486001,MP95A9487001,MP95A9487002,MP95A9489001,MP95A948A001,MP95A948C001 MP95A948C002,MP95A9491001,MP95A9493001,MP95A9493002,MP95A9493003,MP95A9493004,MP95A9493005 MP95A9493006,MP95A9493007,MP95A9494001,MP95A9494002,MP95A9495001,MP95A9495002,MP95A955A001 MP95A955C001,MP95A955C002,MP95A955C003,MP95A9564001,MP95A9567001,MP95A9567002,MP95A9592001 MP95A9593001,MP95A9595001,MP95A9398001(*) 出荷時期:平成27年2月 〜 令和元年8月 出荷台数:87台 (6)天井走行式X線管懸垂器 CH-200 製造番号:0162M45146,0162M45147,0162M45148,0162M45149,0162M45150,0162M45151,0162M45152,0162M45153 0162M45154,0162M45155,0162M45201,0162M45202,0162M45203,0162M45204,0162M45205,0162M45206 0162M45207,0162M45208,0162M45209,0162M45210,0162M45301,0162M45302,0162M45304,0162M45305 0162M45307,0162M45308,0162M45309,0162M45310,0162M45401,0162M45402,0162M45403,0162M45404 0162M45407,0162M45410,0162M45501,0162M45502,0162M45503,0162M45504,0162M45505,0162M45506 0162M45507,0162M45508,0162M45509,0162M45510,0162M45601,0162M45602,0162M45603,0162M45604 0162M45605,0162M45606,0162M45607,0162M45608,0162M45609,0162M45610,0162M45701,0162M45702 0162M45703,0162M45704,0162M45705,0162M45706,0162M45707,0162M45801,0162M45802,0162M45803 0162M45804,0162M45805,0162M45806,0162M45807,0162M45808,0162M45904,0162M45905,0162M45906 0162M45907,0162M45908,0162M45909,0162M45910,0162M46002,0162M46003,0162M46004,0162M46005 0162M46006,0162M46007,0162M46008,0162M46009,0162M46010,0162M46405,0162M46406,0162M46409 0162M46410,0162M46506,0162M46507,0162M46901,0162M47006,0162M47007,0162M47008,0162M47009 0162M47010,0162M47201,0162M47202,0162M47203,0162M47204,0162M47205,0162M47206,0162M47207 0162M47208,0162M47210,0162M47303,0162M47304,0162M47408,0162M47409,0162M47410,0162M47501 0162M47603,0162M47604,0162M47605,0162M47606,0162M47608,0162M47609,0162M47610,0162M47701 0162M47702,0162M47703,0162M47801,0162M47802,0162M47803,0162M47804,0162M47805,0162M47806 0162M47901,0162M48004,0162M48005,0162M48006,0162M48007,0162M48008,0162M48009,0162M48010 0162M48101,0162M48102,0162M48103,0162M48104,0162M48105,0162M48106,0162M48107,0162M48108 0162M48109,0162M48110,0162M48201,0162M48202,0162M48203,0162M48204,0162M48205,0162M48206 0162M48207,0162M48208,0162M48209,0162M48210,0162M48301,0162M48302,0162M48303,0162M48304 0162M48305,0162M48306,0162M48307,0162M48308,0162M48309,0162M48310,0162M48401,0162M48402 0162M48403,0162M48404,0162M48405,0162M48408,0162M48409,0162M48410,0162M48501,0162M48502 0162M48503,0162M48504,0162M48507,0162M48508,0162M48509,0162M48701,0162M48709,0162M48710 0162M48804,0162M48805,0162M48806,0162M48809,0162M49008,0162M49009,0162M49101,0162M49102 0162M49103,0162M49109,0162M49110,0162M49208,0162M49209,0162M49505,0162M49506,0162M49509 0162M49510,0162M49604,0162M49605,0162M49608,0162M49609,0162M49610,0162M49701,0162M49702 0162M49703,0162M49705,0162M49706,0162M49707,0162M49708,0162M49709,0162M49710,0162M49801 0162M49901,0162M49902,0162M49903,0162M49904,0162M49905,0162M49906,0162M49907,0162M49908 0162M49910,0162M50002,0162M50010,018S132101,0262M45201,0262M45509,0262M45510,0262M45606 0262M45607,0262M45608,0262M45609,0262M45701,0262M45702,0262M45805,0262M45806,0262M45809 0262M45810,0262M45907,0262M45908,0262M45909,0262M46001,0262M46002,0262M46101,0262M46106 0262M46107,0262M46108,0262M46207,0262M46208,0262M46210,0262M46306,0262M46307,0262M46308 0262M46409,0262M46501,0262M46502,0262M46503,0262M46509,0262M46510,0262M46601,0262M46602 0262M46603,0262M46604,0262M46605,0262M46606,0262M46607,0262M46608,0262M46609,0262M46610 0262M46709,0262M46710,0262M46904,0262M46905,0262M46906,0262M46907,0262M46908,0262M46909 0262M46910,0262M47101,0262M47103,0262M47106,0262M47107,0262M47108,0262M47109,0262M47110 0262M47210,0262M47301,0262M47303,0262M47304,0262M47305,0262M47306,0262M47307,0262M47308 0262M47309,0262M47310,0262M47401,0262M47402,0262M47403,0262M47404,0262M47410,0262M47501 0262M47502,0262M47601,0262M47602,0262M47603,0262M47604,0262M47605,0262M47607,0262M47608 0262M47609,0262M47610,0262M47701,0262M47702,0262M47703,0262M47704,0262M47709,0262M47710 0262M47902,0262M47903,0262M48008,0262M48209,0262M48210,0262M48309,0262M48310,0262M48407 0262M48408,0262M48409,0262M48501,0262M48502,0262M48503,0262M48504,0262M48505,0262M48506 0262M48507,0262M48508,0262M48801,0262M48803,0262M48810,0262M48906,0262M49110,0262M49308 0262M49309,0262M49401,0262M49402,0262M49403,0262M49404,0262M49405,0262M49406,0262M49407 0262M49501,0262M49504,0262M49505,0262M49602,0262M49603,0262M49604,0262M49605,0262M49606 0262M49607,0262M49608,0262M49609,0262M49810,0262M50001,0262M50002,0262M50003,0262M50004 0262M50009,0262M50010,0362M45102,0362M45103,0362M45104,0362M45105,0362M45106,0362M45107 0362M45110,0362M45201,0362M45202,0362M45203,0362M45207,0362M45208,0362M45209,0362M45210 0362M45301,0362M45302,0362M45310,0362M45505,0362M45506,0362M45508,0362M45509,0362M45510 0362M45601,0362M45602,0362M45605,0362M45606,0362M45608,0362M45707,0362M45709,0362M45710 0362M45808,0362M45809,0362M45902,0362M45906,0362M46005,0362M46204,0362M46205,0362M46210 0362M46306,0362M46309,0362M46310,0362M46401,0362M46507,0362M47304,0362M47305,0362M47306 0362M47307,0362M47308,0362M47309,0362M47310,0362M47401,0362M47402,0362M47403,0362M47404 0362M47505,0362M47510,0362M47602,0362M47603,0362M47604,0362M47605,0362M47606,0362M47607 0362M47608,0362M47705,0362M47706,0362M47910,0362M48005,0362M48006,0362M48007,0362M48008 0362M48010,0362M48201,0362M48202,0362M48208,0362M48209,0362M48210,0362M48306,0362M48307 0362M48406,0362M48407,0362M48501,0362M48801,0362M49002,0362M49003,0362M49004,0362M49005 0362M49006,0362M49007,0362M49009,0362M49010,0362M49310,0362M49409,0362M49501,0362M49502 0362M49503,0362M49601,0362M49602,0362M49609,0362M49610,0362M49710,0362M49804,0362M49805 0362M49907,0362M49908,0362M49910,0362M50008,0362M50010,0462M45101,0462M45102,0462M45102 0462M45103,0462M45104,0462M45105,0462M45109,0462M45110,0462M45309,0462M45310,0462M45401 0462M45402,0462M45410,0462M45501,0462M45502,0462M45503,0462M45504,0462M45505,0462M45601 0462M45908,0462M45909,0462M45910,0462M46001,0462M46003,0462M46004,0462M46006,0462M46104 0462M46105,0462M46106,0462M46107,0462M46108,0462M46109,0462M46110,0462M46201,0462M46202 0462M46204,0462M46205,0462M46209,0462M46210,0462M46310,0462M46404,0462M46405,0462M46409 0462M46410,0462M46501,0462M46502,0462M46503,0462M46504,0462M46609,0462M46610,0462M46901 0462M47205,0462M47206,0462M47207,0462M47209,0462M47210,0462M47301,0462M47401,0462M47402 0462M47403,0462M47404,0462M47405,0462M47406,0462M47607,0462M47608,0462M47609,0462M47710 0462M47807,0462M47903,0462M47904,0462M47905,0462M47906,0462M48101,0462M48102,0462M48103 0462M48104,0462M48105,0462M48107,0462M48110,0462M48205,0462M48404,0462M48405,0462M48406 0462M48407,0462M48408,0462M48409,0462M48410,0462M48507,0462M48508,0462M48509,0462M48606 0462M48607,0462M48608,0462M48609,0462M48610,0462M48701,0462M48702,0462M48703,0462M48704 0462M48705,0462M48706,0462M48707,0462M48708,0462M48709,0462M48801,0462M48907,0462M48909 0462M48910,0462M49201,0462M49207,0462M49208,0462M49209,0462M49303,0462M49304,0462M49305 0462M49306,0462M49501,0462M49602,0462M49603,0462M49604,0462M49605,0462M49704,0462M49705 0462M49707,0462M50007,0562M45405,0562M45409,0562M45410,0562M45506,0562M45509,0562M45701 0562M45702,0562M45703,0562M45704,3ZC5C240B00A,3ZC5C240B00E,3ZC5C240C00C,3ZC5C2412003 3ZC5C2412008,3ZC5C2412009,3ZC5C2413001,3ZC5C2413003,3ZC5C2413005,3ZC5C241300B,3ZC5C2415003 3ZC5C2417004,3ZC5C2418002,3ZC5C2419001,3ZC5C241A007,3ZC5C241A008,3ZC5C241A009,3ZC5C241A010 3ZC5C241A011,3ZC5C241A013,3ZC5C241A014,3ZC5C241B008,3ZC5C241B010,3ZC5C2421003,3ZC5C2422001 3ZC5C2423010,3ZC5C2423022,3ZC5C2424004,3ZC5C2425006,3ZC5C2425012,3ZC5C2426001,3ZC5C2426002 3ZC5C2427004,3ZC5C2428001,3ZC5C2428002,3ZC5C2428005,3ZC5C2428006,3ZC5C2428010,3ZC5C2429008 3ZC5C2429014,3ZC5C242A008,3ZC5C242B003,3ZC5C242C001,3ZC5C242C009,3ZC5C242C010,3ZC5C242C011 3ZC5C242C014,3ZC5C242C017,3ZC5C2431008,3ZC5C2431018,3ZC5C2432008,3ZC5C2432013,3ZC5C2432014 3ZC5C2432018,3ZC5C2433007,3ZC5C2433008,3ZC5C2433010,3ZC5C2434001,3ZC5C2436001,3ZC5C2436002 3ZC5C2437002,3ZC5C2437003,3ZC5C2437005,3ZC5C2437008,3ZC5C2439006,3ZC5C243A004,3ZC5C243A005 3ZC5C243B004,3ZC5C243B006,3ZC5C243B007,3ZC5C243B008,3ZC5C243C002,3ZC5C243C005,3ZC5C243C007 3ZC5C243C010,3ZC5C243C012,3ZC5C243C013,3ZC5C2441010,3ZC5C2442007,3ZC5C2442021,3ZC5C2443002 3ZC5C2443010,3ZC5C2443017,3ZC5C2443018,3ZC5C2447020,3ZC5C2449006,3ZC5C244A002,3ZC5C244A007 3ZC5C244C003,3ZC5C244C004,3ZC5C244C005,3ZC5C244C006,3ZC5C244C007,3ZC5C244C008,3ZC5C244C009 3ZC5C244C011,3ZC5C244C013,3ZC5C244C017,3ZC5C2451014,3ZC5C2451015,3ZC5C2451020,3ZC5C2452005 3ZC5C2452022,3ZC5C2452037,3ZC5C2452038,3ZC5C2452039,3ZC5C2453001,3ZC5C2453002,3ZC5C2453012 3ZC5C2454002,3ZC5C2454008,3ZC5C2454009,3ZC5C2456002,3ZC5C2456004,3ZC5C2456005,3ZC5C250C001 3ZC5C2514001,3ZC5C2532001,3ZC5C2533001,3ZC5C2539001,3ZC5C253C001,3ZC5C255B002,3ZC5C2571001 3ZC5C2581001,3ZC5C2592001,3ZC5C266C001,3ZC5C268A001,3ZC5C271A001,3ZC5C271A002,3ZC5C2A8B005 3ZC5C3014001,3ZC5C3014002,3ZC5C3014003,3ZC5C3021001,3ZC5C3021002,3ZC5C3031001,3ZC5C3031002 3ZC5C3031003,3ZC5C3032001,3ZC5C3037001,3ZC5C303C001,3ZC5C303C002,3ZC5C303C003,3ZC5C303C004 3ZC5C303C005,3ZC5C3041002,3ZC5C3049006,3ZC5C3B32001,3ZC5C425A001,3ZC5C425B008,3ZC5C425C007 3ZC5C4261001,3ZC5C4261010,3ZC5C4261012,3ZC5C4262010,3ZC5C4262017,3ZC5C4262019,3ZC5C4262020 3ZC5C4263005,3ZC5C4263018,3ZC5C4263029,3ZC5C4264003,3ZC5C4266001,3ZC5C4268011,3ZC5C426B009 3ZC5C4271004,3ZC5C4272008,3ZC5C4272009,3ZC5C4273011,3ZC5C4273012,3ZC5C4274010,3ZC5C4278013 3ZC5C4279006,3ZC5C4279010,3ZC5C4279011,3ZC5C427A001,3ZC5C427A005,3ZC5C427A007,3ZC5C427C004 3ZC5C4282001,3ZC5C4283014,3ZC5C4283015,3ZC5C4283020,3ZC5C4287010,3ZC5C4289020,3ZC5C4289021 3ZC5C428B016,3ZC5C4292009,3ZC5C4292011,3ZC5C4292017,3ZC5C4293006,3ZC5C4293009,3ZC5C4293017 3ZC5C4295001,3ZC5C4295002,3ZC5C561B003,LZC5C5196004,LZC5C5196012,LZC5C5196033 LZC5C5196036(*),LZC5C5197005(*) 出荷時期:平成16年9月 〜 令和元年8月(*) 出荷台数:812台(*) (7)天井走行式X線管懸垂器 CH-200M 製造番号:0162M40101,0162M40102,0162M40103,0162M40104,0162M40108,0162M40109,0162M40110,0162M40111 0162M40112,0162M40113,0162M40114,0162M40115,0162M40116,0162M40117,0162M40118,0162M40119 0162M40120,0162M40201,0162M40202,0162M40203,0162M40204,0162M40205,0162M40209,0162M40210 0162M40211,0162M40212,0162M40213,0162M40214,0162M40215,0162M40217,0162M40218,0162M40219 0162M40220,0162M40302,0162M40303,0162M40304,0162M40305,0162M40306,0162M40307,0162M40308 0162M40309,0162M40310,0162M40312,0162M40313,0162M40314,0162M40315,0162M40316,0162M40317 0162M40318,0162M40319,0162M40320,0162M40401,0162M40402,0162M40403,0162M40404,0162M40405 0162M40406,0162M40407,0162M40408,0162M40409,0162M40410,0162M40501,0162M40502,0162M40504 0162M40505,0162M40507,0162M40508,0162M40509,0162M40510,0162M40514,0162M40515,0162M40516 0162M40517,0162M40518,0162M40519,0162M40520,0162M40601,0162M40602,0162M40603,0162M40604 0162M40605,0162M40606,0162M40607,0162M40611,0162M40612,0162M40613,0162M40616,0162M40617 0162M40618,0162M40619,0162M40620,0162M40701,0162M40702,0162M40703,0162M40704,0162M40705 0162M40706,0162M40707,0162M40708,0162M40709,0162M40710,0162M40711,0162M40712,0162M40713 0162M40714,0162M40716,0162M40717,0162M40718,0162M40719,0162M40720,0162M40801,0162M40802 0162M40803,0162M40804,0162M40805,0162M40806,0162M40807,0162M40808,0162M40809,0162M40810 0162M40815,0162M40816,0162M40817,0162M40818,0162M40820,0162M40901,0162M40902,0162M40903 0162M40904,0162M40905,0162M40906,0162M40907,0162M40910,0162M40911,0162M40912,0162M40913 0162M40914,0162M40915,0162M40916,0162M40917,0162M41001,0162M41002,0162M41003,0162M41004 0162M41005,0162M41009,0162M41011,0162M41012,0162M41013,0162M41015,0162M41017,0162M41018 0162M41019,0162M41020,0162M41103,0162M41104,0162M41105,0162M41106,0162M41108,0162M41109 0162M41111,0162M41112,0162M41113,0162M41114,0162M41115,0162M41116,0162M41117,0162M41118 0162M41119,0162M41120,0162M41202,0162M41203,0162M41204,0162M41205,0162M41207,0162M41208 0162M41209,0162M41210,0162M41211,0162M41212,0162M41213,0162M41214,0162M41215,0162M41216 0162M41217,0162M41219,0162M41220,0162M41301,0162M41302,0162M41305,0162M41306,0162M41307 0162M41311,0162M41312,0162M41313,0162M41314,0162M41316,0162M41317,0162M41319,0162M41320 0162M41401,0162M41402,0162M41404,0162M41405,0162M41406,0162M41407,0162M41408,0162M41409 0162M41410,0162M41411,0162M41412,0162M41413,0162M41414,0162M41416,0162M41417,0162M41418 0162M41419,0162M41420,0162M41501,0162M41502,0162M41503,0162M41506,0162M41507,0162M41511 0162M41512,0162M41513,0162M41514,0162M41515,0162M41516,0162M41517,0162M41518,0162M41519 0162M41520,0162M41601,0162M41602,0162M41604,0162M41605,0162M41606,0162M41607,0162M41608 0162M41609,0162M41611,0162M41612,0162M41613,0162M41614,0162M41615,0162M41616,0162M41617 0162M41618,0162M41619,0162M41620,0162M41711,0162M41712,0162M41714,0162M41715,0162M41716 0162M41717,0162M41718,0162M41719,0162M41720,0162M41801,0162M41802,0162M41805,0162M41806 0162M41807,0162M41808,0162M41809,0162M41810,0162M41811,0162M41812,0162M41813,0162M41814 0162M41815,0162M41816,0162M41817,0162M41818,0162M41819,0162M41820,0162M41901,0162M41905 0162M41907,0162M41909,0162M41910,0162M41914,0162M41915,0162M41916,0162M41917,0162M41918 0162M41919,0162M41920,0162M42003,0162M42004,0162M42005,0162M42006,0162M42009,0162M42012 0162M42013,0162M42016,0162M42017,0162M42018,0162M42019,0162M42020,0162M42101,0162M42201 0162M42202,0162M42210,0162M42211,0162M42212,0162M42213,0162M42214,0162M42215,0162M42216 0162M42217,0162M42218,0162M42219,0162M42303,0162M42306,0162M42308,0162M42309,0162M42312 0162M42313,0162M42314,0162M42315,0162M42401,0162M42403,0162M42407,0162M42408,0162M42409 0162M42410,0162M42411,0162M42412,0162M42501,0162M42503,0162M42504,0162M42505,0162M42506 0162M42507,0162M42508,0162M42509,0162M42510,0162M42511,0162M42512,0162M42515,0162M42516 0162M42517,0162M42519,0162M42520,0162M42601,0162M42602,0162M42604,0162M42605,0162M42607 0162M42608,0162M42609,0162M42610,0162M42611,0162M42612,0162M42613,0162M42614,0162M42615 0162M42616,0162M42617,0162M42619,0162M42620,0162M42702,0162M42704,0162M42706,0162M42707 0162M42708,0162M42709,0162M42710,0162M42711,0162M42712,0162M42713,0162M42714,0162M42715 0162M42717,0162M42719,0162M42720,0162M42901,0162M42902,0162M42904,0162M42905,0162M42908 0162M42909,0162M42910,0162M42912,0162M42913,0162M42914,0162M42915,0162M42916,0162M43003 0162M43005,0162M43008,0162M43010,0162M43012,0162M43013,0162M43014,0162M43104,0162M43105 0162M43106,0162M43107,0162M43109,0162M43110,0162M43111,0162M43112,0162M43113,0162M43114 0162M43118,0162M43201,0162M43202,0162M43203,0162M43204,0162M43206,0162M43207,0162M43209 0162M43210,0162M43211,0162M43212,0162M43213,0162M43214,0162M43215,0162M43216,0162M43217 0162M43219,0162M43220,0162M43501,0162M43503,0162M43504,0162M43505,0162M43507,0162M43508 0162M43509,0162M43510,0162M43512,0162M43513,0162M43514,0162M43515,0162M43516,0162M43517 0162M43519,0162M43520,0162M43602,0162M43603,0162M43604,0162M43605,0162M43608,0162M43609 0162M43703,0162M43705,0162M43706,0162M43707,0162M43708,0162M43711,0162M43712,0162M43713 0162M43714,0162M43718,0162M43719,0162M43903,0162M43904,0162M43905,0162M43906,0162M43907 0162M43908,0162M43910,0162M43911,0162M43912,0162M43913,0162M43915,0162M43917,0162M43918 0162M43920,0162M44001,0162M44002,0162M44003,0162M44004,0162M44005,0162M44009,0162M44010 0162M44011,0162M44012,0162M44013,0162M44014,0162M44015,0162M44016,0162M44017,0162M44020 0162M44103,0162M44104,0162M44107,0162M44108,0162M44109,0162M44110,0162M44202,0162M44203 0162M44204,0162M44205,0162M44206,0162M44207,0162M44208,0162M44209,0162M44210,0162M44301 0162M44303,0162M44304,0162M44305,0162M44307,0162M44308,0162M44309,0162M44310,0162M44401 0162M44403,0162M44404,0162M44405,0162M44501,0162M44502,0162M44503,0162M44504,0162M44505 0162M44506,0162M44507,0162M44601,0162M44602,0162M44603,0162M44606,0162M44607,0162M44609 0162M44610,0162M44701,0162M44703,0162M44704,0162M44705,0162M44709,0162M44710,0162M44801 0162M44802,0162M44803,0162M44804,0162M44805,0162M44806,0162M44807,0162M44809,0162M44810 0162M44901,0162M44902,0162M44903,0162M44905,0162M44909,0162M44910,0162M45001,0162M45002 0162M45003,0162M45004,0162M45005,0162M45006,0162M45007,0162M45008,0162M45009,0162M45010 0262M40102,0262M40103,0262M40104,0262M40105,0262M40109,0262M40110,0262M40206,0262M40207 0262M40208,0262M40209,0262M40210,0262M40309,0262M40401,0262M40402,0262M40403,0262M40404 0262M40405,0262M40406,0262M40407,0262M40408,0262M40409,0262M40410,0262M40501,0262M40502 0262M40503,0262M40504,0262M40505,0262M40601,0262M40602,0262M40603,0262M40604,0262M40605 0262M40702,0262M40703,0262M40704,0262M40705,0262M40706,0262M40707,0262M40708,0262M40709 0262M40801,0262M40802,0262M40803,0262M40804,0262M40805,0262M40806,0262M40807,0262M40808 0262M40809,0262M40810,0262M40901,0262M40902,0262M40903,0262M40904,0262M40905,0262M40906 0262M40907,0262M40908,0262M40909,0262M40910,0262M41002,0262M41003,0262M41004,0262M41005 0262M41008,0262M41009,0262M41010,0262M41101,0262M41102,0262M41103,0262M41104,0262M41105 0262M41106,0262M41107,0262M41108,0262M41109,0262M41110,0262M41201,0262M41301,0262M41302 0262M41303,0262M41305,0262M41306,0262M41307,0262M41308,0262M41309,0262M41310,0262M41401 0262M41402,0262M41403,0262M41404,0262M41405,0262M41406,0262M41407,0262M41408,0262M41409 0262M41410,0262M41501,0262M41503,0262M41504,0262M41505,0262M41507,0262M41508,0262M41509 0262M41510,0262M41605,0262M41606,0262M41607,0262M41608,0262M41610,0262M41701,0262M41703 0262M41704,0262M41706,0262M41707,0262M41710,0262M41801,0262M41803,0262M41804,0262M41805 0262M41806,0262M41807,0262M41809,0262M41810,0262M41903,0262M41909,0262M41910,0262M42001 0262M42005,0262M42006,0262M42007,0262M42008,0262M42201,0262M42202,0262M42207,0262M42208 0262M42209,0262M42210,0262M42306,0262M42307,0262M42309,0262M42310,0262M42401,0262M42402 0262M42403,0262M42407,0262M42408,0262M42409,0262M42410,0262M42504,0262M42505,0262M42506 0262M42507,0262M42508,0262M42601,0262M42602,0262M42605,0262M42606,0262M42607,0262M42608 0262M42609,0262M42610,0262M42701,0262M42702,0262M42703,0262M42704,0262M42705,0262M42706 0262M42707,0262M42708,0262M42805,0262M42806,0262M42807,0262M42809,0262M42903,0262M42904 0262M42906,0262M42907,0262M42908,0262M42909,0262M42910,0262M43001,0262M43002,0262M43003 0262M43004,0262M43005,0262M43007,0262M43008,0262M43009,0262M43010,0262M43102,0262M43103 0262M43104,0262M43105,0262M43106,0262M43108,0262M43109,0262M43110,0262M43201,0262M43203 0262M43205,0262M43206,0262M43207,0262M43208,0262M43209,0262M43401,0262M43402,0262M43403 0262M43405,0262M43407,0262M43408,0262M43410,0262M43501,0262M43502,0262M43503,0262M43504 0262M43505,0262M43506,0262M43507,0262M43508,0262M43509,0262M43604,0262M43605,0262M43606 0262M43607,0262M43608,0262M43609,0262M43610,0262M43705,0262M43706,0262M43707,0262M43708 0262M43710,0262M43901,0262M43903,0262M43906,0262M43908,0262M43909,0262M43910,0262M44001 0262M44002,0262M44003,0262M44004,0262M44006,0262M44008,0262M44010,0262M44102,0262M44105 0262M44106,0262M44107,0262M44110,0262M44201,0262M44206,0262M44207,0262M44208,0262M44209 0262M44210,0262M44305,0262M44306,0262M44307,0262M44308,0262M44309,0262M44310,0262M44401 0262M44402,0262M44501,0262M44502,0262M44503,0262M44504,0262M44505,0262M44506,0262M44507 0262M44508,0262M44509,0262M44510,0262M44602,0262M44603,0262M44604,0262M44606,0262M44607 0262M44608,0262M44609,0262M44701,0262M44702,0262M44703,0262M44704,0262M44705,0262M44707 0262M44708,0262M44709,0262M44710,0262M44810,0262M44902,0262M44903,0262M44904,0262M44905 0262M44907,0262M44908,0262M44909,0262M45001,0262M45003,0262M45004,0262M45005,0262M45006 0262M45007,0262M45008,0262M45009,0262M45010,0362M40102,0362M40104,0362M40105,0362M40106 0362M40107,0362M40108,0362M40109,0362M40110,0362M40201,0362M40202,0362M40203,0362M40204 0362M40205,0362M40206,0362M40207,0362M40208,0362M40210,0362M40303,0362M40305,0362M40308 0362M40309,0362M40310,0362M40401,0362M40403,0362M40406,0362M40407,0362M40408,0362M40409 0362M40410,0362M40507,0362M40508,0362M40509,0362M40602,0362M40603,0362M40604,0362M40605 0362M40606,0362M40607,0362M40608,0362M40609,0362M40610,0362M40805,0362M40806,0362M40808 0362M40809,0362M40810,0362M40910,0362M41002,0362M41003,0362M41010,0362M41103,0362M41105 0362M41106,0362M41107,0362M41108,0362M41109,0362M41110,0362M41201,0362M41202,0362M41203 0362M41204,0362M41205,0362M41206,0362M41207,0362M41208,0362M41210,0362M41301,0362M41302 0362M41303,0362M41304,0362M41305,0362M41307,0362M41308,0362M41309,0362M41403,0362M41404 0362M41405,0362M41501,0362M41502,0362M41503,0362M41504,0362M41505,0362M41506,0362M41507 0362M41508,0362M41509,0362M41510,0362M41601,0362M41602,0362M41605,0362M41608,0362M41609 0362M41610,0362M41701,0362M41702,0362M41703,0362M41704,0362M41705,0362M41706,0362M41707 0362M41708,0362M41710,0362M41803,0362M41804,0362M41805,0362M41809,0362M41810,0362M41905 0362M41906,0362M41907,0362M42003,0362M42004,0362M42101,0362M42102,0362M42103,0362M42104 0362M42108,0362M42109,0362M42110,0362M42206,0362M42207,0362M42208,0362M42209,0362M42210 0362M42301,0362M42302,0362M42304,0362M42305,0362M42402,0362M42505,0362M42509,0362M42510 0362M42601,0362M42605,0362M42608,0362M42609,0362M42610,0362M42804,0362M42902,0362M42903 0362M42904,0362M42905,0362M42906,0362M42908,0362M42909,0362M42910,0362M43003,0362M43004 0362M43005,0362M43006,0362M43007,0362M43008,0362M43009,0362M43010,0362M43101,0362M43102 0362M43103,0362M43104,0362M43105,0362M43106,0362M43107,0362M43109,0362M43110,0362M43201 0362M43202,0362M43203,0362M43204,0362M43205,0362M43206,0362M43207,0362M43208,0362M43209 0362M43210,0362M43401,0362M43402,0362M43403,0362M43404,0362M43405,0362M43406,0362M43407 0362M43409,0362M43410,0362M43501,0362M43502,0362M43503,0362M43504,0362M43505,0362M43507 0362M43508,0362M43509,0362M43510,0362M43605,0362M43608,0362M43609,0362M43610,0362M43701 0362M43702,0362M43703,0362M43704,0362M43705,0362M43706,0362M43707,0362M43708,0362M43709 0362M43801,0362M43809,0362M43908,0362M43909,0362M43910,0362M44007,0362M44008,0362M44009 0362M44010,0362M44105,0362M44106,0362M44107,0362M44109,0362M44110,0362M44202,0362M44203 0362M44204,0362M44205,0362M44206,0362M44208,0362M44209,0362M44210,0362M44301,0362M44302 0362M44303,0362M44304,0362M44305,0362M44307,0362M44308,0362M44309,0362M44310,0362M44404 0362M44405,0362M44504,0362M44601,0362M44604,0362M44605,0362M44609,0362M44610,0362M44706 0362M44707,0362M44708,0362M44709,0362M44710,0362M44801,0362M44802,0362M44803,0362M44804 0362M44805,0362M44809,0362M44810,0362M44907,0362M45006,0362M45007,0362M45008,0362M45009 0362M45010,0462M40104,0462M40105,0462M40106,0462M40107,0462M40108,0462M40109,0462M40110 0462M40203,0462M40204,0462M40205,0462M40206,0462M40207,0462M40208,0462M40209,0462M40210 0462M40301,0462M40302,0462M40303,0462M40304,0462M40305,0462M40306,0462M40307,0462M40308 0462M40309,0462M40310,0462M40401,0462M40404,0462M40405,0462M40406,0462M40407,0462M40408 0462M40409,0462M40410,0462M40501,0462M40502,0462M40503,0462M40504,0462M40505,0462M40506 0462M40507,0462M40508,0462M40509,0462M40510,0462M40601,0462M40602,0462M40603,0462M40604 0462M40605,0462M40606,0462M40607,0462M40608,0462M40609,0462M40610,0462M40701,0462M40702 0462M40706,0462M40708,0462M40709,0462M40710,0462M40801,0462M40901,0462M41001,0462M41002 0462M41003,0462M41004,0462M41005,0462M41007,0462M41008,0462M41009,0462M41010,0462M41102 0462M41103,0462M41104,0462M41105,0462M41106,0462M41107,0462M41108,0462M41110,0462M41208 0462M41209,0462M41210,0462M41301,0462M41302,0462M41303,0462M41305,0462M41306,0462M41307 0462M41308,0462M41309,0462M41403,0462M41404,0462M41405,0462M41406,0462M41407,0462M41409 0462M41410,0462M41501,0462M41508,0462M41509,0462M41510,0462M41603,0462M41604,0462M41605 0462M41607,0462M41609,0462M41610,0462M41702,0462M41706,0462M41707,0462M41708,0462M41709 0462M41710,0462M41801,0462M41802,0462M41803,0462M41804,0462M41805,0462M41807,0462M41808 0462M41809,0462M41810,0462M41903,0462M41904,0462M41910,0462M42010,0462M42106,0462M42107 0462M42108,0462M42109,0462M42110,0462M42204,0462M42209,0462M42210,0462M42405,0462M42407 0462M42408,0462M42409,0462M42410,0462M42501,0462M42503,0462M42504,0462M42507,0462M42508 0462M42509,0462M42510,0462M42601,0462M42602,0462M42603,0462M42604,0462M42605,0462M42608 0462M42609,0462M42610,0462M42701,0462M42705,0462M42706,0462M42707,0462M42708,0462M42710 0462M42801,0462M42802,0462M42803,0462M42804,0462M42805,0462M42901,0462M42902,0462M42903 0462M42904,0462M42905,0462M42907,0462M42908,0462M42909,0462M42910,0462M43004,0462M43005 0462M43006,0462M43007,0462M43008,0462M43009,0462M43010,0462M43101,0462M43104,0462M43105 0462M43106,0462M43107,0462M43108,0462M43109,0462M43110,0462M43205,0462M43207,0462M43208 0462M43209,0462M43210,0462M43301,0462M43302,0462M43303,0462M43304,0462M43305,0462M43407 0462M43408,0462M43409,0462M43410,0462M43506,0462M43507,0462M43508,0462M43510,0462M43602 0462M43603,0462M43604,0462M43605,0462M43607,0462M43608,0462M43609,0462M43610,0462M43701 0462M43702,0462M43703,0462M43704,0462M43705,0462M43706,0462M43709,0462M43801,0462M43802 0462M43901,0462M43902,0462M43903,0462M43904,0462M43905,0462M43908,0462M44003,0462M44005 0462M44006,0462M44007,0462M44008,0462M44009,0462M44010,0462M44101,0462M44102,0462M44103 0462M44104,0462M44105,0462M44106,0462M44107,0462M44108,0462M44109,0462M44110,0462M44206 0462M44207,0462M44208,0462M44209,0462M44210,0462M44301,0462M44302,0462M44304,0462M44308 0462M44310,0462M44402,0462M44403,0462M44404,0462M44405,0462M44406,0462M44407,0462M44408 0462M44409,0462M44410,0462M44501,0462M44502,0462M44503,0462M44504,0462M44505,0462M44506 0462M44507,0462M44508,0462M44601,0462M44604,0462M44606,0462M44607,0462M44608,0462M44609 0462M44610,0462M44701,0462M44702,0462M44703,0462M44704,0462M44705,0462M44706,0462M44707 0462M44708,0462M44709,0462M44710,0462M44802,0462M44803,0462M44804,0462M44805,0462M44806 0462M44807,0462M44808,0462M44809,0462M44810,0462M44906,0462M44907,0462M44908,0462M44909 0462M44910,0462M45006,0462M45007,0462M45008,0462M45010,0562M40104,0562M40105,0562M40106 0562M40109,0562M40201,0562M40202,0562M40203,0562M40205,0562M40208,0562M40209,0562M40210 0562M40301,0562M40302,0562M40303,0562M40304,0562M40305,0562M40306,0562M40307,0562M40308 0562M40309,0562M40310,0562M40406,0562M40407,0562M40408,0562M40409,0562M40410,0562M40501 0562M40502,0562M40503,0562M40504,0562M40505,0562M40506,0562M40510,0562M40602,0562M40603 0562M40605,0562M40606,0562M40607,0562M40608,0562M40609,0562M40610,0562M40705,0562M40706 0562M40708,0562M40709,0562M40710,0562M40802,0562M40803,0562M40804,0562M40806,0562M40807 0562M40808,0562M40809,0562M40810,0562M40902,0562M40903,0562M40904,0562M40905,0562M40910 0562M41002,0562M41004,0562M41005,0562M41006,0562M41007,0562M41008,0562M41009,0562M41104 0562M41107,0562M41108,0562M41109,0562M41110,0562M41202,0562M41203,0562M41204,0562M41205 0562M41207,0562M41208,0562M41210,0562M41301,0562M41302,0562M41303,0562M41304,0562M41305 0562M41306,0562M41307,0562M41308,0562M41402,0562M41403,0562M41404,0562M41407,0562M41408 0562M41409,0562M41410,0562M41501,0562M41502,0562M41504,0562M41505,0562M41506,0562M41507 0562M41508,0562M41509,0562M41510,0562M41601,0562M41602,0562M41603,0562M41604,0562M41606 0562M41607,0562M41608,0562M41609,0562M41610,0562M41706,0562M41707,0562M41708,0562M41801 0562M41802,0562M41803,0562M41804,0562M41808,0562M41907,0562M41908,0562M42002,0562M42003 0562M42004,0562M42005,0562M42006,0562M42007,0562M42008,0562M42105,0562M42106,0562M42107 0562M42108,0562M42109,0562M42202,0562M42203,0562M42204,0562M42205,0562M42206,0562M42209 0562M42301,0562M42303,0562M42304,0562M42305,0562M42306,0562M42307,0562M42308,0562M42309 0562M42401,0562M42402,0562M42403,0562M42404,0562M42405,0562M42406,0562M42407,0562M42408 0562M42409,0562M42410,0562M42502,0562M42506,0562M42507,0562M42508,0562M42509,0562M42510 0562M42601,0562M42604,0562M42607,0562M42609,0562M42610,0562M42704,0562M42705,0562M42706 0562M42707,0562M42708,0562M42709,0562M42710,0562M42809,0562M42810,0562M42908,0562M42909 0562M42910,0562M43003,0562M43004,0562M43006,0562M43008,0562M43009,0562M43010,0562M43101 0562M43103,0562M43106,0562M43107,0562M43201,0562M43202,0562M43203,0562M43204,0562M43205 0562M43206,0562M43207,0562M43208,0562M43209,0562M43210,0562M43301,0562M43302,0562M43303 0562M43304,0562M43307,0562M43308,0562M43309,0562M43310,0562M43404,0562M43405,0562M43406 0562M43407,0562M43408,0562M43409,0562M43410,0562M43501,0562M43502,0562M43503,0562M43504 0562M43505,0562M43510,0562M43601,0562M43603,0562M43604,0562M43605,0562M43606,0562M43607 0562M43608,0562M43609,0562M43610,0562M43701,0562M43702,0562M43703,0562M43704,0562M43707 0562M43709,0562M43710,0562M43801,0562M43802,0562M43803,0562M43805,0562M43806,0562M43807 0562M43808,0562M43809,0562M43810,0562M43904,0562M43905,0562M43908,0562M43909,0562M43910 0562M44001,0562M44002,0562M44003,0562M44004,0562M44005,0562M44006,0562M44007,0562M44008 0562M44009,0562M44010,0562M44102,0562M44103,0562M44104,0562M44108,0562M44109,0562M44110 0562M44201,0562M44202,0562M44203,0562M44204,0562M44205,0562M44206,0562M44207,0562M44301 0562M44303,0562M44304,0562M44305,0562M44306,0562M44308,0562M44309,0562M44310,0562M44402 0562M44403,0562M44404,0562M44406,0562M44506,0562M44702,0562M44703,0562M44704,0562M44708 0562M44710,0562M44801,0562M44802,0562M44804,0562M44805,0562M44806,0562M44807,0562M44808 0562M44810,0562M44903,0562M44905,0562M44906,0562M44907,0562M45006,0562M45010,0662M40110 0662M40201,0662M40205,0662M40503,0662M40609,0662M40610,0662M40706,0662M40909,0662M41106 0662M41110,0662M41201,0662M41205,0662M41208,0662M41309,0662M41310,0662M41401,0662M41404 0662M41506,3ZC5C100A00B,3ZC5C100A00D,3ZC5C100A016,3ZC5C100A01D,3ZC5C100A021,3ZC5C100B012 3ZC5C100B016,3ZC5C100B01B,3ZC5C100B020,3ZC5C100B021,3ZC5C100B02A,3ZC5C1011002,3ZC5C1011004 3ZC5C1011005,3ZC5C101100E,3ZC5C1011013,3ZC5C1011018,3ZC5C1012003,3ZC5C1012006,3ZC5C1012009 3ZC5C1013002,3ZC5C1013005,3ZC5C1013008,3ZC5C101300D,3ZC5C101300F,3ZC5C1013011,3ZC5C1015018 3ZC5C1015024,3ZC5C1015025,3ZC5C1015026,3ZC5C1016001,3ZC5C1016002,3ZC5C1016003,3ZC5C1016006 3ZC5C1017003,3ZC5C1017008,3ZC5C1017011,3ZC5C1017022,3ZC5C1017023,3ZC5C1017030,3ZC5C1018008 3ZC5C1018014,3ZC5C1018018,3ZC5C1018024,3ZC5C1018025,3ZC5C1019008,3ZC5C101A003,3ZC5C101A005 3ZC5C101A006,3ZC5C101A009,3ZC5C101B002,3ZC5C101B003,3ZC5C101B010,3ZC5C101B011,3ZC5C101C004 3ZC5C101C011,3ZC5C101C012,3ZC5C101C018,3ZC5C101C027,3ZC5C101C028,3ZC5C101C029,3ZC5C1021007 3ZC5C1021014,3ZC5C1021025,3ZC5C1021031,3ZC5C1021032,3ZC5C1022004,3ZC5C1022009,3ZC5C1022013 3ZC5C1023006,3ZC5C1023017,3ZC5C1023019,3ZC5C1024002,3ZC5C1024012,3ZC5C1024013,3ZC5C1024017 3ZC5C1024018,3ZC5C1024019,3ZC5C1024023,3ZC5C1024024,3ZC5C1024025,3ZC5C1024029,3ZC5C1024035 3ZC5C1025009,3ZC5C1025014,3ZC5C1026004,3ZC5C1026015,3ZC5C1026020,3ZC5C1026022,3ZC5C1026023 3ZC5C1027001,3ZC5C1027005,3ZC5C1027006,3ZC5C1027015,3ZC5C1028001,3ZC5C1028011,3ZC5C1028016 3ZC5C1029018,3ZC5C102A008,3ZC5C102A011,3ZC5C102A012,3ZC5C102A013,3ZC5C102A017,3ZC5C102A019 3ZC5C102B001,3ZC5C102B004,3ZC5C102B005,3ZC5C102B006,3ZC5C102B011,3ZC5C102B019,3ZC5C102B024 3ZC5C102B026,3ZC5C102C004,3ZC5C102C005,3ZC5C102C012,3ZC5C102C025,3ZC5C1031004,3ZC5C1031008 3ZC5C1031010,3ZC5C1031011,3ZC5C1031012,3ZC5C1031013,3ZC5C1031020,3ZC5C1031022,3ZC5C1032007 3ZC5C1032013,3ZC5C1032014,3ZC5C1032015,3ZC5C1033004,3ZC5C1033009,3ZC5C1033013,3ZC5C1033015 3ZC5C1033022,3ZC5C1033025,3ZC5C1034025,3ZC5C1035009,3ZC5C1036001,3ZC5C1036003,3ZC5C1036004 3ZC5C1036008,3ZC5C1036016,3ZC5C1036025,3ZC5C1036026,3ZC5C1037001,3ZC5C1037020,3ZC5C1037024 3ZC5C1038012,3ZC5C1038020,3ZC5C1038021,3ZC5C1038025,3ZC5C1039001,3ZC5C1039005,3ZC5C1039006 3ZC5C1039011,3ZC5C1039021,3ZC5C1039024,3ZC5C103A006,3ZC5C103A009,3ZC5C103B010,3ZC5C103C007 3ZC5C103C008,3ZC5C103C012,3ZC5C103C013,3ZC5C103C019,3ZC5C1041007,3ZC5C1041024,3ZC5C1042006 3ZC5C1042013,3ZC5C1042014,3ZC5C1042020,3ZC5C1043009,3ZC5C1043010,3ZC5C1043032,3ZC5C1043037 3ZC5C1043043,3ZC5C1044002,3ZC5C1044003,3ZC5C1044004,3ZC5C1044005,3ZC5C1044007,3ZC5C1044013 3ZC5C1044020,3ZC5C1044021,3ZC5C1044023,3ZC5C1044027,3ZC5C1044031,3ZC5C1044032,3ZC5C1045010 3ZC5C1046005,3ZC5C1046014,3ZC5C1046017,3ZC5C1046022,3ZC5C1046025,3ZC5C1049009,3ZC5C104A015 3ZC5C104B001,3ZC5C104B003,3ZC5C104B005,3ZC5C104B008,3ZC5C104B011,3ZC5C104B015,3ZC5C104B016 3ZC5C104C001,3ZC5C104C007,3ZC5C104C008,3ZC5C1051003,3ZC5C1051017,3ZC5C1052001,3ZC5C1052018 3ZC5C1053009,3ZC5C1053027,3ZC5C1054017,3ZC5C1055003,3ZC5C1055004,3ZC5C1055008,3ZC5C1055011 3ZC5C1055015,3ZC5C1056006,3ZC5C1056008,3ZC5C1056014,3ZC5C1057002,3ZC5C1057005,3ZC5C1057006 3ZC5C1057015,3ZC5C1057017,3ZC5C1057023,3ZC5C1058010,3ZC5C1059003,3ZC5C110A002,3ZC5C1113002 3ZC5C1113003,3ZC5C1114002,3ZC5C1117001,3ZC5C1119001,3ZC5C1119003,3ZC5C111B002,3ZC5C111C001 3ZC5C1122001,3ZC5C1122003,3ZC5C1124001,3ZC5C1124002,3ZC5C1124003,3ZC5C1124004,3ZC5C1125002 3ZC5C1125003,3ZC5C1127002,3ZC5C1127003,3ZC5C1128001,3ZC5C112C001,3ZC5C1131001,3ZC5C1135002 3ZC5C1135003,3ZC5C1135004,3ZC5C1136002,3ZC5C1136005,3ZC5C1137003,3ZC5C113A001,3ZC5C113A003 3ZC5C113B002,3ZC5C1142001,3ZC5C1144001,3ZC5C1144002,3ZC5C1147003,3ZC5C1157001,3ZC5C1157002 3ZC5C1158001,3ZC5C1159005,3ZC5C115B003,3ZC5C115C001,3ZC5C1163001,3ZC5C1163004,3ZC5C116B001 3ZC5C1172001,3ZC5C1172002,3ZC5C1172003,3ZC5C1173002,3ZC5C1174001,3ZC5C1179001,3ZC5C1179002 3ZC5C117A001,3ZC5C1182001,3ZC5C1184002,3ZC5C1213001,3ZC5C1219003,3ZC5C1223001,3ZC5C1224003 3ZC5C1224005,3ZC5C1225001,3ZC5C1228001,3ZC5C122A001,3ZC5C1232005,3ZC5C1233001,3ZC5C1235001 3ZC5C1237002,3ZC5C1239001,3ZC5C1239002,3ZC5C123C001,3ZC5C1241001,3ZC5C1242001,3ZC5C1245001 3ZC5C1249001,3ZC5C125A001,3ZC5C1261001,3ZC5C126B002,3ZC5C1271002,3ZC5C1272001,3ZC5C1279004 3ZC5C127A001,3ZC5C1285001,3ZC5C1289001,3ZC5C128C005,3ZC5C1292003,3ZC5C1293002,3ZC5C1C23001 3ZC5C1C29001,3ZC5C1C37001,3ZC5C1C4C001,3ZC5C1C5A001,3ZC5C1C65001,3ZC5C1C67003,3ZC5C1C71001 3ZC5C1C75001,3ZC5C1C75002,3ZC5C1C79001,3ZC5C1C81001,3ZC5C1C81002,3ZC5C1C82001,3ZC5C1C91001 3ZC5C205A006,3ZC5C205A010,3ZC5C205A014,3ZC5C205A015,3ZC5C205A018,3ZC5C205B001,3ZC5C205B004 3ZC5C205B009,3ZC5C205B012,3ZC5C205C006,3ZC5C205C008,3ZC5C205C014,3ZC5C2061003,3ZC5C2061009 3ZC5C2061016,3ZC5C2062001,3ZC5C2062002,3ZC5C2062003,3ZC5C2062004,3ZC5C2062007,3ZC5C2062013 3ZC5C2063003,3ZC5C2063006,3ZC5C2063009,3ZC5C2063015,3ZC5C2063019,3ZC5C2063024,3ZC5C2063032 3ZC5C2064003,3ZC5C2064005,3ZC5C2064007,3ZC5C2064012,3ZC5C2064017,3ZC5C2064021,3ZC5C2064022 3ZC5C2064023,3ZC5C2065001,3ZC5C2065006,3ZC5C2065007,3ZC5C2065009,3ZC5C2065019,3ZC5C2065025 3ZC5C2065026,3ZC5C2066007,3ZC5C2066011,3ZC5C2067002,3ZC5C2067003,3ZC5C2067007,3ZC5C2067013 3ZC5C2068002,3ZC5C2069013,3ZC5C206A001,3ZC5C206A002,3ZC5C206A005,3ZC5C206B001,3ZC5C206B002 3ZC5C206B003,3ZC5C206C001,3ZC5C206C009,3ZC5C2071003,3ZC5C2071004,3ZC5C2071010,3ZC5C2071015 3ZC5C2072001,3ZC5C2072002,3ZC5C2072005,3ZC5C2073001,3ZC5C2073002,3ZC5C2073003,3ZC5C2073004 3ZC5C2073013,3ZC5C2073016,3ZC5C2073018,3ZC5C2073029,3ZC5C2073033,3ZC5C2073036,3ZC5C2074006 3ZC5C2074011,3ZC5C2074015,3ZC5C2074016,3ZC5C2075002,3ZC5C2075003,3ZC5C2076003,3ZC5C2076008 3ZC5C2078015,3ZC5C2078020,3ZC5C2079003,3ZC5C2079005,3ZC5C2079006,3ZC5C2079012,3ZC5C2079013 3ZC5C2079026,3ZC5C207A011,3ZC5C207B008,3ZC5C207B010,3ZC5C207B011,3ZC5C207C010,3ZC5C207C024 3ZC5C2081009,3ZC5C2083004,3ZC5C2084004,3ZC5C2084018,3ZC5C2085018,3ZC5C2086005,3ZC5C2086007 3ZC5C2086009,3ZC5C2086019,3ZC5C2086020,3ZC5C2086021,3ZC5C2087014,3ZC5C2088003,3ZC5C2088004 3ZC5C2088008,3ZC5C2088012,3ZC5C2089015,3ZC5C2089023,3ZC5C208A011,3ZC5C208A012,3ZC5C208A017 3ZC5C208B005,3ZC5C208B006,3ZC5C208B012,3ZC5C208C003,3ZC5C2091002,3ZC5C2091003,3ZC5C2091006 3ZC5C2091009,3ZC5C2091019,3ZC5C2092008,3ZC5C2092020,3ZC5C2092021,3ZC5C2092028,3ZC5C2093003 3ZC5C2093007,3ZC5C2093008,LZC5C4C93009,LZC5C4C95030,LZC5C4C95031,LZC5C4C96010,LZC5C4C96011 LZC5C4C96013,LZC5C4C96019, LZC5C4C96032(*),LZC5C4C96042(*),LZC5C4C97002(*),LZC5C4C97011(*),LZC5C4C97017(*) LZC5C4C97027(*),LZC5C4C97039(*),LZC5C4C97041(*),LZC5C4D97002(*),LZC5C4E96002(*) LZC5C4E98001(*),LZC5C4F96002(*) 出荷時期:平成15年9月 〜 令和2年1月(*) 出荷台数:2,243台(*) 合計台数:6,574台(*)
製造販売業者等名称
製造販売業者の名称 : 株式会社島津製作所 製造販売業者の所在地: 京都府京都市中京区西ノ京桑原町1番地 許可の種類 : 第一種医療機器製造販売業 許可番号 : 26B1X00003
改修理由
当該装置は、X線管懸垂部とX線管装置を取り付けるX線管保持部から構成されており、X線管保持部はX線管懸垂 部の支柱に8本のボルトにて固定されていますが、使用中にこのボルトが緩む可能性があることが判明しまし た。使用にあたり、鉛直を軸としてX線管保持部の回転動作を行う際に繰り返しメカニカルストッパ等に当たり 強く衝撃が加わることがない限り、ボルトの緩みが発生することはないと考えていますが、ボルトの緩んだ状態 では回転動作時にボルトに力が加わり疲労によりボルトの破損に至る可能性があります。8本すべてのボルトが 破損に至った場合には、最終的にX線管保持部が支柱より外れます。 今回自主改修として、万一、すべてのボルトが破損に至った場合でもX線管保持部が外れないよう保護部品を追 加する対策を実施します。
危惧される具体的な健康被害
X線管保持部が支柱より外れた場合には被検者や操作者に接触し打撲等の健康被害が起きる可能性が否定できま せん。ただし、本機器は医療従事者の監視のもと使用されるため、本機器の使用により重篤な健康被害が発生 する恐れはありません。X線管保持部はX線管装置の位置合わせのため、ブレーキの解除し、回転させることが できますが、ボルトが全て破損した場合、ブレーキを解除しない状態でも、回転動作を行える異常な状態とな りす。ブレーキ動作の確認は日常点検の項目に含まれており、日常点検を適切に実施していれば異常に気が付 くことができます。
改修開始年月日
令和5年4月10日
効能・効果又は用途等
(1)〜(5)人体を透過したX線の蛍光作用、写真作用及び電離作用を利用して人体画像情報を診療のために提供し ます。 (6)、(7)診断用X線装置に付属するX線管装置の取付け、保持、位置調整のために使用する構造を形成してい る天井取付け式金具組立品、及び関連する機械式、電子式、又はソフトウェア式制御機能を有し、X線高電圧 装置及び撮影スタンドや撮影テーブルと組合わせて被検者の立位や臥位のX線撮影を行ないます。
その他
納入いたしました医療機関はすべて弊社が把握しておりますので、連絡の上、改修を実施します。 なお、令和2年1月(*)以降の出荷品は、ボルトが全て破断した状態でもX線管保持部が外れない構造となってい るため、問題はありません。
担当者及び連絡先
担当者 : 谷川 勝哉・山本 武史 連絡先 : 株式会社島津製作所 メディカル規制統括部 安全管理グループ 電話番号: 075-823-1928 FAX番号 : 075-823-2530