閉じる |
令和 4年 6月23日作成
一般的名称及び販売名
一般的名称: (1)自動植込み型除細動器 (2)除細動機能付植込み型両心室ペーシングパルスジェネレータ 販売名 : (1)Cobalt MRI ICDシリーズ (2)Cobalt MRI CRT-Dシリーズ
対象ロット、数量及び出荷時期
(1)Cobalt MRI ICDシリーズ モデル番号: DDPA2D1, DDPA2D4, DVPA2D1, DVPA2D4 回収対象製造番号: RSC601291S, RSD603425S, RSD603492S, RSD603971S, RSL601920S, RSM603972S, RSM605206S, RSM606563S, RSM606878S, RSM606934S, RSM607754S, RSM607776S, RSM607835S, RSM608473S, RSM608599S, RSM609036S, RSM609094S, RSM609098S, RSM609142S, RSM609180S, RSM609219S, RSM609238S 患者モニタリング対象製造番号: RSC600057S, RSC600067S, RSC600068S, RSC600070S, RSC600072S, RSC600073S, RSC600076S, RSC600078S, RSC600083S, RSC600088S, RSC600089S, RSC600090S, RSC600092S, RSC600094S, RSC600095S, RSC600097S, RSC600101S, RSC600102S, RSC600104S, RSC600105S, RSC600106S, RSC600107S, RSC600108S, RSC600109S, RSC600114S, RSC600117S, RSC600119S, RSC600122S, RSC600124S, RSC600266S, RSC600439S, RSC600454S, RSC600456S, RSC600457S, RSC600458S, RSC600522S, RSC600527S, RSC600528S, RSC600531S, RSC600532S, RSC600533S, RSC600535S, RSC600537S, RSC600541S, RSC600546S, RSC600556S, RSC600557S, RSC600558S, RSC600561S, RSC600563S, RSC600565S, RSC600566S, RSC600567S, RSC600570S, RSC600574S, RSC600576S, RSC600582S, RSC600584S, RSC600587S, RSC600590S, RSC600591S, RSC600592S, RSC600594S, RSC600595S, RSC600596S, RSC600600S, RSC600601S, RSC600602S, RSC600603S, RSC600607S, RSC600608S, RSC600614S, RSC600829S, RSC600853S, RSC600854S, RSC600856S, RSC600859S, RSC600867S, RSC600868S, RSC600871S, RSC600872S, RSC600873S, RSC600875S, RSC600877S, RSC600879S, RSC600882S, RSC600887S, RSC600896S, RSC600897S, RSC600898S, RSC600900S, RSC600901S, RSC600903S, RSC600906S, RSC600909S, RSC600918S, RSC600928S, RSC600934S, RSC600935S, RSC600936S, RSC600937S, RSC600942S, RSC600943S, RSC600944S, RSC600945S, RSC600950S, RSC600953S, RSC600954S, RSC600955S, RSC600957S, RSC600972S, RSC600973S, RSC600974S, RSC600975S, RSC600976S, RSC600977S, RSC600979S, RSC600980S, RSC600984S, RSC600985S, RSC600991S, RSC600994S, RSC600995S, RSC600996S, RSC600997S, RSC600998S, RSC601000S, RSC601013S, RSC601279S, RSC601283S, RSC601284S, RSC601293S, RSC601562S, RSC601591S, RSC601596S, RSC601728S, RSC601754S, RSC601768S, RSC601789S, RSC601792S, RSC601793S, RSC601794S, RSC601799S, RSC601801S, RSC601805S, RSC601815S, RSC601820S, RSC601828S, RSC601927S, RSC601959S, RSC602000S, RSC602033S, RSC602067S, RSD600050S, RSD600051S, RSD600052S, RSD600053S, RSD600054S, RSD600055S, RSD600056S, RSD600058S, RSD600066S, RSD600068S, RSD600069S, RSD600070S, RSD600072S, RSD600073S, RSD600076S, RSD600077S, RSD600078S, RSD600079S, RSD600082S, RSD600083S, RSD600084S, RSD600085S, RSD600086S, RSD600088S, RSD600101S, RSD600102S, RSD600104S, RSD600105S, RSD600107S, RSD600109S, RSD600110S, RSD600111S, RSD600113S, RSD600116S, RSD600117S, RSD600118S, RSD600119S, RSD600122S, RSD600125S, RSD600126S, RSD600128S, RSD600131S, RSD600133S, RSD600135S, RSD600140S, RSD600141S, RSD600142S, RSD600144S, RSD600146S, RSD600147S, RSD600149S, RSD600151S, RSD600152S, RSD600153S, RSD600155S, RSD600157S, RSD600160S, RSD600162S, RSD600166S, RSD600167S, RSD600173S, RSD600613S, RSD600626S, RSD600627S, RSD600631S, RSD600632S, RSD600637S, RSD600638S, RSD600640S, RSD600642S, RSD600643S, RSD600648S, RSD600649S, RSD600650S, RSD600651S, RSD600653S, RSD600655S, RSD600657S, RSD600658S, RSD600659S, RSD600660S, RSD600661S, RSD600662S, RSD600663S, RSD600665S, RSD600668S, RSD600670S, RSD600671S, RSD600673S, RSD600674S, RSD600675S, RSD600676S, RSD600677S, RSD600678S, RSD600680S, RSD600681S, RSD600682S, RSD600683S, RSD600685S, RSD600687S, RSD600688S, RSD600689S, RSD600690S, RSD600692S, RSD600693S, RSD600694S, RSD600696S, RSD600698S, RSD600699S, RSD600702S, RSD600703S, RSD600704S, RSD600706S, RSD600708S, RSD600709S, RSD600710S, RSD600714S, RSD600718S, RSD600719S, RSD600723S, RSD600724S, RSD600725S, RSD600726S, RSD600729S, RSD600730S, RSD600731S, RSD600734S, RSD600735S, RSD600738S, RSD600740S, RSD600742S, RSD600747S, RSD600749S, RSD600750S, RSD600751S, RSD600752S, RSD600756S, RSD600757S, RSD600762S, RSD600774S, RSD600786S, RSD600791S, RSD600799S, RSD600804S, RSD600816S, RSD601023S, RSD601024S, RSD601027S, RSD601029S, RSD601030S, RSD601035S, RSD601037S, RSD601038S, RSD601040S, RSD601041S, RSD601044S, RSD601046S, RSD601048S, RSD601049S, RSD601051S, RSD601055S, RSD601057S, RSD601058S, RSD601061S, RSD601068S, RSD601071S, RSD601084S, RSD601085S, RSD601090S, RSD601093S, RSD601105S, RSD601109S, RSD601110S, RSD601113S, RSD601118S, RSD601120S, RSD601137S, RSD601187S, RSD601194S, RSD601195S, RSD601196S, RSD601207S, RSD601212S, RSD601221S, RSD601223S, RSD601231S, RSD601234S, RSD601238S, RSD601253S, RSD601256S, RSD601267S, RSD601275S, RSD601281S, RSD601297S, RSD601343S, RSD601383S, RSD601393S, RSD601464S, RSD601465S, RSD601468S, RSD601469S, RSD601471S, RSD601473S, RSD601474S, RSD601479S, RSD601488S, RSD601492S, RSD601503S, RSD601510S, RSD601515S, RSD601526S, RSD601540S, RSD601623S, RSD601636S, RSD601638S, RSD601647S, RSD601648S, RSD601656S, RSD601671S, RSD601675S, RSD601688S, RSD601692S, RSD601693S, RSD601697S, RSD601698S, RSD601699S, RSD601702S, RSD601703S, RSD601704S, RSD601707S, RSD601708S, RSD601710S, RSD601755S, RSD601764S, RSD601766S, RSD601767S, RSD601768S, RSD601776S, RSD601777S, RSD601779S, RSD601780S, RSD601781S, RSD601783S, RSD601787S, RSD601790S, RSD602125S, RSD602126S, RSD602131S, RSD602133S, RSD602148S, RSD602159S, RSD602436S, RSD602437S, RSD602442S, RSD602445S, RSD602449S, RSD602506S, RSD602507S, RSD602508S, RSD602577S, RSD602950S, RSD602955S, RSD602956S, RSD602962S, RSD602964S, RSD603076S, RSD603092S, RSD603113S, RSD603143S, RSD603206S, RSD603264S, RSD603268S, RSD603275S, RSD603282S, RSD603284S, RSD603435S, RSD603450S, RSD603454S, RSD603468S, RSD603472S, RSD603475S, RSD603476S, RSD603495S, RSD603919S, RSD603921S, RSD603931S, RSD604034S, RSD604109S, RSD604115S, RSD604146S, RSD604153S, RSD604157S, RSD604167S, RSD604168S, RSD604174S, RSD604177S, RSD604198S, RSD604204S, RSD604232S, RSD604234S, RSD604240S, RSD604255S, RSD604256S, RSD604262S, RSD604289S, RSD604292S, RSD604361S, RSD604391S, RSD604392S, RSD604398S, RSD604401S, RSD604402S, RSD604407S, RSD604410S, RSD604417S, RSD604419S, RSD604420S, RSD604425S, RSD604426S, RSD604428S, RSD604561S, RSD604641S, RSD604642S, RSD604644S, RSD604646S, RSD604648S, RSD604649S, RSD604651S, RSD604654S, RSD604655S, RSD604657S, RSD604658S, RSD604660S, RSD604821S, RSD604824S, RSD604829S, RSD604836S, RSD604838S, RSD604842S, RSD604844S, RSD604845S, RSD604851S, RSD604858S, RSD604863S, RSD604867S, RSD605024S, RSD605048S, RSD605071S, RSD605074S, RSD605075S, RSD605087S, RSD605098S, RSD605101S, RSD605111S, RSD605117S, RSD605129S, RSD605462S, RSD605487S, RSD605496S, RSD605510S, RSD605517S, RSD605527S, RSD605558S, RSD605597S, RSL600127S, RSL600128S, RSL600131S, RSL600132S, RSL600134S, RSL600135S, RSL600136S, RSL600137S, RSL600138S, RSL600139S, RSL600140S, RSL600143S, RSL600144S, RSL600145S, RSL600146S, RSL600325S, RSL600476S, RSL600731S, RSL600732S, RSL600734S, RSL600735S, RSL600737S, RSL600738S, RSL600739S, RSL600740S, RSL600742S, RSL600744S, RSL600745S, RSL600747S, RSL600748S, RSL600749S, RSL600750S, RSL600752S, RSL600756S, RSL600760S, RSL600763S, RSL600766S, RSL600767S, RSL600768S, RSL600773S, RSL600774S, RSL600775S, RSL600776S, RSL600780S, RSL600783S, RSL600784S, RSL600786S, RSL600791S, RSL600792S, RSL600796S, RSL600799S, RSL600800S, RSL600829S, RSL600830S, RSL600831S, RSL600832S, RSL600834S, RSL600835S, RSL600836S, RSL600837S, RSL600838S, RSL600839S, RSL600841S, RSL600843S, RSL600844S, RSL600847S, RSL600849S, RSL600850S, RSL600851S, RSL600852S, RSL600853S, RSL600854S, RSL600856S, RSL600859S, RSL600861S, RSL600862S, RSL600863S, RSL600864S, RSL600865S, RSL600866S, RSL600867S, RSL600872S, RSL600877S, RSL600880S, RSL600881S, RSL600883S, RSL600885S, RSL600886S, RSL600887S, RSL600888S, RSL600889S, RSL600890S, RSL600892S, RSL600893S, RSL600895S, RSL600897S, RSL600898S, RSL600899S, RSL600900S, RSL600902S, RSL600903S, RSL600905S, RSL600907S, RSL600908S, RSL600909S, RSL600911S, RSL600912S, RSL600913S, RSL600914S, RSL600916S, RSL600917S, RSL600918S, RSL600920S, RSL600921S, RSL600923S, RSL600924S, RSL600925S, RSL600926S, RSL600927S, RSL600928S, RSL600929S, RSL600931S, RSL600932S, RSL600933S, RSL600936S, RSL600937S, RSL600938S, RSL600941S, RSL600942S, RSL600943S, RSL600944S, RSL600946S, RSL600947S, RSL600948S, RSL600950S, RSL600952S, RSL600953S, RSL600956S, RSL600957S, RSL600959S, RSL600960S, RSL600961S, RSL600962S, RSL600965S, RSL600966S, RSL600967S, RSL600968S, RSL600969S, RSL600971S, RSL600973S, RSL600974S, RSL600975S, RSL600977S, RSL600978S, RSL600980S, RSL600982S, RSL600983S, RSL600984S, RSL600986S, RSL600987S, RSL600988S, RSL600989S, RSL600991S, RSL600992S, RSL600994S, RSL600995S, RSL600996S, RSL600997S, RSL601073S, RSL601188S, RSL601189S, RSL601191S, RSL601194S, RSL601200S, RSL601201S, RSL601205S, RSL601215S, RSL601217S, RSL601221S, RSL601222S, RSL601223S, RSL601227S, RSL601228S, RSL601230S, RSL601237S, RSL601238S, RSL601243S, RSL601244S, RSL601247S, RSL601250S, RSL601251S, RSL601252S, RSL601254S, RSL601255S, RSL601285S, RSL601286S, RSL601288S, RSL601289S, RSL601291S, RSL601292S, RSL601315S, RSL601316S, RSL601318S, RSL601319S, RSL601324S, RSL601326S, RSL601328S, RSL601390S, RSL601391S, RSL601392S, RSL601394S, RSL601395S, RSL601396S, RSL601411S, RSL601414S, RSL601415S, RSL601417S, RSL601418S, RSL601421S, RSL601422S, RSL601424S, RSL601425S, RSL601426S, RSL601428S, RSL601547S, RSL601548S, RSL601550S, RSL601551S, RSL601553S, RSL601556S, RSL601557S, RSL601558S, RSL601879S, RSL601880S, RSL601883S, RSL601884S, RSL601886S, RSL601888S, RSL601890S, RSL601891S, RSL601892S, RSL601893S, RSL601894S, RSL601897S, RSL601903S, RSL601909S, RSL601918S, RSL601923S, RSL601924S, RSL601926S, RSL602008S, RSL602092S, RSL602093S, RSL602094S, RSL602095S, RSL602100S, RSL602106S, RSL602121S, RSL602123S, RSL602124S, RSL602125S, RSL602126S, RSL602127S, RSL602132S, RSL602133S, RSL602135S, RSL602136S, RSL602137S, RSL602138S, RSL602140S, RSL602142S, RSL602143S, RSL602144S, RSL602145S, RSL602148S, RSL602149S, RSL602150S, RSL602151S, RSL602153S, RSL602154S, RSL602155S, RSL602156S, RSL602160S, RSL602172S, RSL602201S, RSL602203S, RSL602205S, RSL602208S, RSL602270S, RSL602273S, RSL602279S, RSL602280S, RSL602281S, RSL602282S, RSL602283S, RSL602284S, RSL602287S, RSL602298S, RSL602300S, RSL602301S, RSL602307S, RSL602313S, RSL602314S, RSL602315S, RSL602320S, RSL602330S, RSL602331S, RSL602334S, RSL602339S, RSL602341S, RSL602342S, RSL602343S, RSL602350S, RSL602362S, RSL602365S, RSL602366S, RSL602373S, RSL602375S, RSL602379S, RSL602384S, RSL602385S, RSL602387S, RSL602391S, RSL602392S, RSL602396S, RSL602398S, RSL602401S, RSL602402S, RSL602639S, RSL602653S, RSL602658S, RSM600038S, RSM600039S, RSM600065S, RSM600066S, RSM600067S, RSM600068S, RSM600069S, RSM600070S, RSM600071S, RSM600072S, RSM600073S, RSM600075S, RSM600077S, RSM600078S, RSM600079S, RSM600084S, RSM600085S, RSM600086S, RSM600087S, RSM600089S, RSM600090S, RSM600091S, RSM600092S, RSM600093S, RSM600094S, RSM600095S, RSM600096S, RSM600097S, RSM600098S, RSM600099S, RSM600100S, RSM600101S, RSM600102S, RSM600103S, RSM600104S, RSM600105S, RSM600106S, RSM600107S, RSM600108S, RSM600109S, RSM600376S, RSM600385S, RSM600502S, RSM600518S, RSM600519S, RSM600520S, RSM600523S, RSM600524S, RSM600525S, RSM600526S, RSM600527S, RSM600528S, RSM600529S, RSM600530S, RSM600531S, RSM600532S, RSM600533S, RSM600534S, RSM600536S, RSM600538S, RSM600539S, RSM600540S, RSM600541S, RSM600543S, RSM600544S, RSM600545S, RSM600548S, RSM600549S, RSM600551S, RSM600552S, RSM600553S, RSM600554S, RSM600555S, RSM600557S, RSM600559S, RSM600560S, RSM600561S, RSM600562S, RSM600563S, RSM600564S, RSM600565S, RSM600568S, RSM600569S, RSM600570S, RSM600571S, RSM600572S, RSM600573S, RSM600574S, RSM600575S, RSM600576S, RSM600577S, RSM600578S, RSM600579S, RSM600581S, RSM600582S, RSM600583S, RSM600585S, RSM600586S, RSM600587S, RSM600588S, RSM600589S, RSM600590S, RSM600591S, RSM600592S, RSM600593S, RSM600594S, RSM600599S, RSM600600S, RSM600602S, RSM600604S, RSM600605S, RSM600606S, RSM600607S, RSM600609S, RSM600610S, RSM600611S, RSM600612S, RSM600613S, RSM600614S, RSM600615S, RSM600617S, RSM600618S, RSM600619S, RSM600620S, RSM600621S, RSM600622S, RSM600623S, RSM600624S, RSM600626S, RSM600627S, RSM600628S, RSM600629S, RSM600632S, RSM600634S, RSM600635S, RSM600642S, RSM600645S, RSM600648S, RSM600941S, RSM601180S, RSM601181S, RSM601182S, RSM601183S, RSM601184S, RSM601185S, RSM601186S, RSM601188S, RSM601189S, RSM601190S, RSM601191S, RSM601192S, RSM601193S, RSM601195S, RSM601196S, RSM601198S, RSM601199S, RSM601200S, RSM601201S, RSM601202S, RSM601203S, RSM601204S, RSM601205S, RSM601206S, RSM601207S, RSM601208S, RSM601209S, RSM601210S, RSM601211S, RSM601212S, RSM601213S, RSM601214S, RSM601215S, RSM601216S, RSM601217S, RSM601218S, RSM601219S, RSM601220S, RSM601221S, RSM601222S, RSM601223S, RSM601224S, RSM601225S, RSM601226S, RSM601227S, RSM601228S, RSM601229S, RSM601230S, RSM601231S, RSM601232S, RSM601233S, RSM601234S, RSM601235S, RSM601236S, RSM601237S, RSM601238S, RSM601239S, RSM601240S, RSM601241S, RSM601242S, RSM601243S, RSM601244S, RSM601245S, RSM601246S, RSM601247S, RSM601248S, RSM601249S, RSM601250S, RSM601251S, RSM601252S, RSM601253S, RSM601254S, RSM601255S, RSM601256S, RSM601258S, RSM601259S, RSM601260S, RSM601261S, RSM601262S, RSM601263S, RSM601264S, RSM601265S, RSM601266S, RSM601267S, RSM601268S, RSM601269S, RSM601270S, RSM601271S, RSM601272S, RSM601273S, RSM601276S, RSM601277S, RSM601278S, RSM601279S, RSM601280S, RSM601281S, RSM601282S, RSM601283S, RSM601284S, RSM601285S, RSM601286S, RSM601287S, RSM601288S, RSM601289S, RSM601290S, RSM601291S, RSM601293S, RSM601294S, RSM601295S, RSM601297S, RSM601298S, RSM601299S, RSM601300S, RSM601301S, RSM601302S, RSM601303S, RSM601305S, RSM601308S, RSM601309S, RSM601310S, RSM601311S, RSM601312S, RSM601313S, RSM601314S, RSM601315S, RSM601316S, RSM601317S, RSM601318S, RSM601319S, RSM601320S, RSM601321S, RSM601322S, RSM601323S, RSM601326S, RSM601327S, RSM601330S, RSM601331S, RSM601332S, RSM601333S, RSM601334S, RSM601335S, RSM601336S, RSM601337S, RSM601338S, RSM601339S, RSM601340S, RSM601341S, RSM601342S, RSM601343S, RSM601344S, RSM601345S, RSM601346S, RSM601347S, RSM601348S, RSM601349S, RSM601350S, RSM601351S, RSM601352S, RSM601353S, RSM601354S, RSM601355S, RSM601356S, RSM601357S, RSM601358S, RSM601359S, RSM601360S, RSM601361S, RSM601362S, RSM601363S, RSM601364S, RSM601365S, RSM601366S, RSM601369S, RSM601370S, RSM601371S, RSM601372S, RSM601373S, RSM601374S, RSM601375S, RSM601376S, RSM601377S, RSM601378S, RSM601379S, RSM601381S, RSM601382S, RSM601383S, RSM601384S, RSM601386S, RSM601387S, RSM601388S, RSM601390S, RSM601391S, RSM601392S, RSM601394S, RSM601395S, RSM601396S, RSM601398S, RSM601399S, RSM601400S, RSM601401S, RSM601402S, RSM601403S, RSM601404S, RSM601405S, RSM601406S, RSM601407S, RSM601408S, RSM601409S, RSM601410S, RSM601411S, RSM601412S, RSM601414S, RSM601415S, RSM601416S, RSM601417S, RSM601418S, RSM601419S, RSM601420S, RSM601421S, RSM601422S, RSM601424S, RSM601425S, RSM601427S, RSM601428S, RSM601429S, RSM601430S, RSM601431S, RSM601432S, RSM601433S, RSM601434S, RSM601435S, RSM601436S, RSM601437S, RSM601438S, RSM601439S, RSM601440S, RSM601442S, RSM601443S, RSM601444S, RSM601445S, RSM601446S, RSM601448S, RSM601449S, RSM601452S, RSM601457S, RSM601462S, RSM601463S, RSM601469S, RSM601471S, RSM601480S, RSM601481S, RSM601483S, RSM601485S, RSM601486S, RSM601487S, RSM601489S, RSM601491S, RSM601493S, RSM601494S, RSM601495S, RSM601496S, RSM601497S, RSM601498S, RSM601500S, RSM601501S, RSM601502S, RSM601503S, RSM601505S, RSM601506S, RSM601507S, RSM601508S, RSM601511S, RSM601512S, RSM601514S, RSM601515S, RSM601516S, RSM601517S, RSM601518S, RSM601520S, RSM601521S, RSM601522S, RSM601525S, RSM601526S, RSM601527S, RSM601532S, RSM601533S, RSM601534S, RSM601535S, RSM601536S, RSM601538S, RSM601540S, RSM601541S, RSM601542S, RSM601715S, RSM601772S, RSM601782S, RSM601796S, RSM601800S, RSM601816S, RSM601820S, RSM601822S, RSM601829S, RSM601830S, RSM601831S, RSM601835S, RSM601837S, RSM601841S, RSM601864S, RSM601865S, RSM601866S, RSM601867S, RSM601868S, RSM601869S, RSM601871S, RSM601872S, RSM601873S, RSM601874S, RSM601877S, RSM601878S, RSM601879S, RSM601880S, RSM601881S, RSM601882S, RSM601885S, RSM601886S, RSM601887S, RSM601888S, RSM601889S, RSM601890S, RSM601891S, RSM601892S, RSM601893S, RSM601895S, RSM601896S, RSM601897S, RSM601900S, RSM601902S, RSM601903S, RSM601904S, RSM601917S, RSM601920S, RSM601928S, RSM601934S, RSM601935S, RSM601937S, RSM601939S, RSM601941S, RSM601942S, RSM601946S, RSM601947S, RSM601948S, RSM601951S, RSM601954S, RSM601956S, RSM601957S, RSM601959S, RSM601960S, RSM601963S, RSM601964S, RSM601967S, RSM601970S, RSM601971S, RSM601973S, RSM601974S, RSM601975S, RSM601976S, RSM601978S, RSM601981S, RSM601983S, RSM601987S, RSM601989S, RSM601990S, RSM601993S, RSM601997S, RSM601998S, RSM602003S, RSM602004S, RSM602005S, RSM602008S, RSM602012S, RSM602016S, RSM602017S, RSM602022S, RSM602023S, RSM602026S, RSM602028S, RSM602030S, RSM602042S, RSM602046S, RSM602060S, RSM602062S, RSM602063S, RSM602064S, RSM602066S, RSM602067S, RSM602068S, RSM602070S, RSM602071S, RSM602073S, RSM602074S, RSM602075S, RSM602076S, RSM602077S, RSM602078S, RSM602080S, RSM602086S, RSM602089S, RSM602090S, RSM602091S, RSM602092S, RSM602095S, RSM602097S, RSM602099S, RSM602100S, RSM602102S, RSM602105S, RSM602106S, RSM602109S, RSM602111S, RSM602113S, RSM602114S, RSM602115S, RSM602117S, RSM602118S, RSM602119S, RSM602120S, RSM602125S, RSM602126S, RSM602127S, RSM602133S, RSM602136S, RSM602139S, RSM602140S, RSM602146S, RSM602147S, RSM602149S, RSM602152S, RSM602153S, RSM602154S, RSM602155S, RSM602156S, RSM602158S, RSM602159S, RSM602161S, RSM602165S, RSM602168S, RSM602170S, RSM602171S, RSM602177S, RSM602180S, RSM602182S, RSM602183S, RSM602184S, RSM602187S, RSM602188S, RSM602189S, RSM602194S, RSM602196S, RSM602197S, RSM602204S, RSM602205S, RSM602208S, RSM602209S, RSM602211S, RSM602212S, RSM602214S, RSM602215S, RSM602219S, RSM602222S, RSM602223S, RSM602224S, RSM602227S, RSM602230S, RSM602232S, RSM602235S, RSM602238S, RSM602239S, RSM602240S, RSM602241S, RSM602242S, RSM602243S, RSM602245S, RSM602246S, RSM602249S, RSM602250S, RSM602261S, RSM602265S, RSM602271S, RSM602272S, RSM602273S, RSM602274S, RSM602275S, RSM602276S, RSM602277S, RSM602281S, RSM602282S, RSM602284S, RSM602285S, RSM602286S, RSM602288S, RSM602289S, RSM602294S, RSM602296S, RSM602303S, RSM602304S, RSM602305S, RSM602307S, RSM602313S, RSM602314S, RSM602315S, RSM602316S, RSM602321S, RSM602323S, RSM602324S, RSM602327S, RSM602330S, RSM602332S, RSM602333S, RSM602334S, RSM602335S, RSM602337S, RSM602338S, RSM602344S, RSM602346S, RSM602350S, RSM602351S, RSM602353S, RSM602355S, RSM602360S, RSM602361S, RSM602362S, RSM602364S, RSM602367S, RSM602370S, RSM602371S, RSM602372S, RSM602375S, RSM602378S, RSM602383S, RSM602384S, RSM602392S, RSM602393S, RSM602394S, RSM602395S, RSM602397S, RSM602398S, RSM602399S, RSM602400S, RSM602401S, RSM602402S, RSM602403S, RSM602404S, RSM602405S, RSM602407S, RSM602408S, RSM602409S, RSM602410S, RSM602411S, RSM602413S, RSM602414S, RSM602415S, RSM602416S, RSM602417S, RSM602418S, RSM602419S, RSM602420S, RSM602422S, RSM602424S, RSM602426S, RSM602427S, RSM602429S, RSM602430S, RSM602431S, RSM602432S, RSM602435S, RSM602436S, RSM602437S, RSM602438S, RSM602439S, RSM602440S, RSM602442S, RSM602443S, RSM602444S, RSM602445S, RSM602446S, RSM602448S, RSM602449S, RSM602450S, RSM602451S, RSM602452S, RSM602454S, RSM602455S, RSM602456S, RSM602458S, RSM602460S, RSM602461S, RSM602462S, RSM602467S, RSM602468S, RSM602469S, RSM602470S, RSM602471S, RSM602476S, RSM602477S, RSM602480S, RSM602481S, RSM602487S, RSM602490S, RSM602493S, RSM602494S, RSM602496S, RSM602499S, RSM602502S, RSM602503S, RSM602504S, RSM602507S, RSM602508S, RSM602510S, RSM602513S, RSM602514S, RSM602522S, RSM602528S, RSM602529S, RSM602531S, RSM602534S, RSM602539S, RSM602540S, RSM602547S, RSM602550S, RSM602555S, RSM602558S, RSM602560S, RSM602564S, RSM602566S, RSM602567S, RSM602571S, RSM602573S, RSM602574S, RSM602575S, RSM602576S, RSM602578S, RSM602579S, RSM602582S, RSM602583S, RSM602584S, RSM602586S, RSM602587S, RSM602588S, RSM602592S, RSM602593S, RSM602596S, RSM602597S, RSM602598S, RSM602600S, RSM602603S, RSM602604S, RSM602606S, RSM602611S, RSM602612S, RSM602614S, RSM602615S, RSM602616S, RSM602618S, RSM602620S, RSM602621S, RSM602622S, RSM602625S, RSM602628S, RSM602629S, RSM602630S, RSM602631S, RSM602650S, RSM602666S, RSM602686S, RSM602690S, RSM602701S, RSM602705S, RSM602718S, RSM602727S, RSM602729S, RSM602731S, RSM602733S, RSM602750S, RSM602753S, RSM602756S, RSM602757S, RSM602762S, RSM602763S, RSM602764S, RSM602770S, RSM602772S, RSM602778S, RSM602779S, RSM602780S, RSM602781S, RSM602787S, RSM602789S, RSM602790S, RSM602796S, RSM602797S, RSM602798S, RSM602799S, RSM602800S, RSM602801S, RSM602802S, RSM602803S, RSM602804S, RSM602805S, RSM602806S, RSM602807S, RSM602808S, RSM602809S, RSM602812S, RSM602813S, RSM602814S, RSM602815S, RSM602818S, RSM602819S, RSM602820S, RSM602823S, RSM602825S, RSM602826S, RSM602827S, RSM602832S, RSM602834S, RSM602835S, RSM602836S, RSM602840S, RSM602843S, RSM602844S, RSM602845S, RSM602847S, RSM602849S, RSM602857S, RSM602860S, RSM602863S, RSM602869S, RSM602875S, RSM602876S, RSM602879S, RSM602881S, RSM602884S, RSM602885S, RSM602886S, RSM602888S, RSM602890S, RSM602891S, RSM602892S, RSM602893S, RSM602895S, RSM602897S, RSM602900S, RSM602902S, RSM602904S, RSM602906S, RSM602908S, RSM602909S, RSM602913S, RSM602914S, RSM602915S, RSM602918S, RSM602919S, RSM602922S, RSM602923S, RSM602925S, RSM602926S, RSM602928S, RSM602936S, RSM602937S, RSM602938S, RSM602939S, RSM602941S, RSM602942S, RSM602943S, RSM602944S, RSM602946S, RSM602947S, RSM602948S, RSM602949S, RSM602950S, RSM602951S, RSM602952S, RSM602953S, RSM602954S, RSM602955S, RSM602956S, RSM602957S, RSM602958S, RSM602959S, RSM602960S, RSM602961S, RSM602962S, RSM602963S, RSM602964S, RSM602965S, RSM602966S, RSM602967S, RSM602968S, RSM602969S, RSM602970S, RSM602971S, RSM602972S, RSM602973S, RSM602974S, RSM602975S, RSM602976S, RSM602977S, RSM602978S, RSM602979S, RSM602980S, RSM602981S, RSM602982S, RSM602983S, RSM602984S, RSM602985S, RSM602986S, RSM602987S, RSM602988S, RSM602989S, RSM602990S, RSM602991S, RSM602992S, RSM602993S, RSM602994S, RSM602995S, RSM602996S, RSM602997S, RSM602998S, RSM602999S, RSM603000S, RSM603001S, RSM603002S, RSM603003S, RSM603004S, RSM603005S, RSM603006S, RSM603007S, RSM603008S, RSM603009S, RSM603010S, RSM603011S, RSM603012S, RSM603013S, RSM603014S, RSM603015S, RSM603016S, RSM603017S, RSM603018S, RSM603019S, RSM603020S, RSM603021S, RSM603022S, RSM603023S, RSM603024S, RSM603025S, RSM603026S, RSM603027S, RSM603028S, RSM603029S, RSM603030S, RSM603031S, RSM603032S, RSM603033S, RSM603034S, RSM603044S, RSM603059S, RSM603082S, RSM603135S, RSM603136S, RSM603138S, RSM603140S, RSM603141S, RSM603142S, RSM603143S, RSM603148S, RSM603152S, RSM603158S, RSM603159S, RSM603160S, RSM603163S, RSM603165S, RSM603166S, RSM603173S, RSM603174S, RSM603175S, RSM603176S, RSM603178S, RSM603181S, RSM603184S, RSM603185S, RSM603193S, RSM603196S, RSM603197S, RSM603199S, RSM603201S, RSM603202S, RSM603204S, RSM603207S, RSM603208S, RSM603209S, RSM603210S, RSM603212S, RSM603214S, RSM603215S, RSM603216S, RSM603217S, RSM603218S, RSM603226S, RSM603227S, RSM603228S, RSM603230S, RSM603232S, RSM603234S, RSM603235S, RSM603237S, RSM603240S, RSM603242S, RSM603243S, RSM603244S, RSM603245S, RSM603246S, RSM603247S, RSM603249S, RSM603252S, RSM603253S, RSM603254S, RSM603255S, RSM603256S, RSM603257S, RSM603259S, RSM603260S, RSM603261S, RSM603262S, RSM603263S, RSM603264S, RSM603266S, RSM603268S, RSM603270S, RSM603271S, RSM603272S, RSM603273S, RSM603274S, RSM603277S, RSM603280S, RSM603282S, RSM603283S, RSM603287S, RSM603290S, RSM603292S, RSM603293S, RSM603294S, RSM603295S, RSM603296S, RSM603297S, RSM603298S, RSM603299S, RSM603300S, RSM603301S, RSM603303S, RSM603304S, RSM603305S, RSM603306S, RSM603308S, RSM603309S, RSM603310S, RSM603313S, RSM603314S, RSM603315S, RSM603316S, RSM603317S, RSM603318S, RSM603319S, RSM603321S, RSM603322S, RSM603324S, RSM603325S, RSM603326S, RSM603327S, RSM603328S, RSM603329S, RSM603330S, RSM603331S, RSM603332S, RSM603333S, RSM603335S, RSM603358S, RSM603359S, RSM603360S, RSM603361S, RSM603369S, RSM603372S, RSM603373S, RSM603375S, RSM603381S, RSM603386S, RSM603388S, RSM603391S, RSM603396S, RSM603412S, RSM603422S, RSM603427S, RSM603472S, RSM603476S, RSM603480S, RSM603492S, RSM603494S, RSM603601S, RSM603628S, RSM603629S, RSM603630S, RSM603631S, RSM603632S, RSM603633S, RSM603634S, RSM603635S, RSM603636S, RSM603637S, RSM603638S, RSM603639S, RSM603640S, RSM603641S, RSM603642S, RSM603643S, RSM603644S, RSM603645S, RSM603646S, RSM603647S, RSM603648S, RSM603649S, RSM603650S, RSM603651S, RSM603653S, RSM603654S, RSM603655S, RSM603656S, RSM603657S, RSM603658S, RSM603659S, RSM603660S, RSM603661S, RSM603662S, RSM603663S, RSM603664S, RSM603665S, RSM603666S, RSM603667S, RSM603668S, RSM603669S, RSM603670S, RSM603671S, RSM603672S, RSM603673S, RSM603674S, RSM603675S, RSM603676S, RSM603677S, RSM603678S, RSM603679S, RSM603680S, RSM603681S, RSM603682S, RSM603683S, RSM603684S, RSM603685S, RSM603686S, RSM603687S, RSM603688S, RSM603689S, RSM603690S, RSM603691S, RSM603692S, RSM603693S, RSM603694S, RSM603695S, RSM603696S, RSM603697S, RSM603698S, RSM603699S, RSM603700S, RSM603701S, RSM603702S, RSM603703S, RSM603704S, RSM603705S, RSM603707S, RSM603708S, RSM603709S, RSM603710S, RSM603711S, RSM603712S, RSM603713S, RSM603714S, RSM603715S, RSM603716S, RSM603717S, RSM603718S, RSM603719S, RSM603720S, RSM603721S, RSM603722S, RSM603723S, RSM603725S, RSM603726S, RSM603727S, RSM603728S, RSM603729S, RSM603730S, RSM603731S, RSM603732S, RSM603733S, RSM603734S, RSM603735S, RSM603736S, RSM603737S, RSM603738S, RSM603739S, RSM603740S, RSM603741S, RSM603742S, RSM603743S, RSM603744S, RSM603745S, RSM603746S, RSM603747S, RSM603748S, RSM603749S, RSM603750S, RSM603751S, RSM603753S, RSM603754S, RSM603755S, RSM603756S, RSM603757S, RSM603758S, RSM603759S, RSM603761S, RSM603762S, RSM603763S, RSM603764S, RSM603765S, RSM603766S, RSM603767S, RSM603768S, RSM603769S, RSM603770S, RSM603771S, RSM603772S, RSM603773S, RSM603774S, RSM603775S, RSM603776S, RSM603777S, RSM603778S, RSM603779S, RSM603780S, RSM603782S, RSM603783S, RSM603784S, RSM603785S, RSM603786S, RSM603787S, RSM603788S, RSM603789S, RSM603790S, RSM603791S, RSM603792S, RSM603793S, RSM603794S, RSM603795S, RSM603796S, RSM603797S, RSM603798S, RSM603799S, RSM603800S, RSM603801S, RSM603802S, RSM603803S, RSM603804S, RSM603805S, RSM603806S, RSM603807S, RSM603808S, RSM603809S, RSM603811S, RSM603812S, RSM603813S, RSM603814S, RSM603816S, RSM603817S, RSM603818S, RSM603819S, RSM603821S, RSM603822S, RSM603823S, RSM603826S, RSM603827S, RSM603828S, RSM603829S, RSM603830S, RSM603831S, RSM603832S, RSM603833S, RSM603834S, RSM603835S, RSM603836S, RSM603837S, RSM603838S, RSM603839S, RSM603840S, RSM603841S, RSM603842S, RSM603844S, RSM603845S, RSM603846S, RSM603847S, RSM603849S, RSM603850S, RSM603852S, RSM603853S, RSM603854S, RSM603855S, RSM603857S, RSM603858S, RSM603859S, RSM603860S, RSM603861S, RSM603862S, RSM603863S, RSM603864S, RSM603865S, RSM603866S, RSM603867S, RSM603868S, RSM603870S, RSM603871S, RSM603872S, RSM603873S, RSM603874S, RSM603875S, RSM603876S, RSM603877S, RSM603878S, RSM603879S, RSM603880S, RSM603881S, RSM603883S, RSM603884S, RSM603885S, RSM603886S, RSM603887S, RSM603888S, RSM603889S, RSM603890S, RSM603891S, RSM603892S, RSM603894S, RSM603911S, RSM603926S, RSM603946S, RSM603963S, RSM603966S, RSM603967S, RSM603968S, RSM603970S, RSM603974S, RSM603976S, RSM603982S, RSM603984S, RSM603985S, RSM603988S, RSM603989S, RSM603990S, RSM603991S, RSM603994S, RSM604001S, RSM604002S, RSM604004S, RSM604008S, RSM604011S, RSM604012S, RSM604014S, RSM604015S, RSM604016S, RSM604017S, RSM604018S, RSM604023S, RSM604024S, RSM604025S, RSM604030S, RSM604031S, RSM604032S, RSM604033S, RSM604034S, RSM604035S, RSM604036S, RSM604038S, RSM604043S, RSM604046S, RSM604048S, RSM604050S, RSM604055S, RSM604057S, RSM604058S, RSM604060S, RSM604061S, RSM604065S, RSM604066S, RSM604070S, RSM604071S, RSM604073S, RSM604074S, RSM604077S, RSM604080S, RSM604081S, RSM604087S, RSM604090S, RSM604091S, RSM604093S, RSM604095S, RSM604097S, RSM604098S, RSM604099S, RSM604100S, RSM604103S, RSM604106S, RSM604107S, RSM604108S, RSM604110S, RSM604113S, RSM604114S, RSM604115S, RSM604118S, RSM604119S, RSM604121S, RSM604122S, RSM604125S, RSM604126S, RSM604127S, RSM604128S, RSM604131S, RSM604132S, RSM604134S, RSM604135S, RSM604136S, RSM604137S, RSM604138S, RSM604139S, RSM604141S, RSM604142S, RSM604170S, RSM604177S, RSM604191S, RSM604192S, RSM604193S, RSM604194S, RSM604195S, RSM604196S, RSM604198S, RSM604199S, RSM604200S, RSM604201S, RSM604234S, RSM604237S, RSM604241S, RSM604250S, RSM604251S, RSM604254S, RSM604255S, RSM604276S, RSM604281S, RSM604283S, RSM604284S, RSM604285S, RSM604286S, RSM604287S, RSM604289S, RSM604295S, RSM604299S, RSM604300S, RSM604320S, RSM604354S, RSM604380S, RSM604386S, RSM604440S, RSM604451S, RSM604452S, RSM604454S, RSM604460S, RSM604471S, RSM604573S, RSM604638S, RSM604642S, RSM604644S, RSM604653S, RSM604658S, RSM604663S, RSM604665S, RSM604669S, RSM604670S, RSM604673S, RSM604674S, RSM604675S, RSM604676S, RSM604677S, RSM604725S, RSM604752S, RSM605208S, RSM605214S, RSM605217S, RSM605220S, RSM605222S, RSM605223S, RSM605225S, RSM605229S, RSM605230S, RSM605231S, RSM605238S, RSM605242S, RSM605243S, RSM605244S, RSM605245S, RSM605277S, RSM605278S, RSM605279S, RSM605280S, RSM605281S, RSM605282S, RSM605283S, RSM605284S, RSM605285S, RSM605286S, RSM605287S, RSM605288S, RSM605289S, RSM605290S, RSM605292S, RSM605293S, RSM605295S, RSM605297S, RSM605298S, RSM605299S, RSM605300S, RSM605301S, RSM605302S, RSM605303S, RSM605304S, RSM605305S, RSM605306S, RSM605307S, RSM605308S, RSM605309S, RSM605310S, RSM605311S, RSM605312S, RSM605313S, RSM605314S, RSM605315S, RSM605316S, RSM605317S, RSM605318S, RSM605320S, RSM605321S, RSM605322S, RSM605323S, RSM605324S, RSM605325S, RSM605326S, RSM605327S, RSM605328S, RSM605329S, RSM605330S, RSM605467S, RSM605468S, RSM605469S, RSM605476S, RSM605478S, RSM605482S, RSM605485S, RSM605491S, RSM605492S, RSM605504S, RSM605658S, RSM605664S, RSM605668S, RSM605686S, RSM605694S, RSM605696S, RSM605707S, RSM605708S, RSM606126S, RSM606130S, RSM606151S, RSM606153S, RSM606154S, RSM606158S, RSM606161S, RSM606166S, RSM606170S, RSM606173S, RSM606175S, RSM606182S, RSM606184S, RSM606188S, RSM606192S, RSM606195S, RSM606198S, RSM606199S, RSM606243S, RSM606245S, RSM606269S, RSM606270S, RSM606288S, RSM606322S, RSM606395S, RSM606411S, RSM606415S, RSM606425S, RSM606447S, RSM606457S, RSM606481S, RSM606486S, RSM606501S, RSM606509S, RSM606512S, RSM606513S, RSM606515S, RSM606516S, RSM606518S, RSM606519S, RSM606520S, RSM606521S, RSM606522S, RSM606523S, RSM606524S, RSM606526S, RSM606528S, RSM606529S, RSM606531S, RSM606532S, RSM606533S, RSM606534S, RSM606535S, RSM606536S, RSM606537S, RSM606538S, RSM606539S, RSM606540S, RSM606541S, RSM606542S, RSM606543S, RSM606544S, RSM606545S, RSM606546S, RSM606547S, RSM606550S, RSM606551S, RSM606552S, RSM606553S, RSM606554S, RSM606555S, RSM606556S, RSM606557S, RSM606558S, RSM606560S, RSM606571S, RSM606573S, RSM606574S, RSM606575S, RSM606576S, RSM606577S, RSM606579S, RSM606581S, RSM606582S, RSM606584S, RSM606585S, RSM606590S, RSM606591S, RSM606592S, RSM606593S, RSM606594S, RSM606595S, RSM606597S, RSM606599S, RSM606653S, RSM606701S, RSM606710S, RSM606713S, RSM606714S, RSM606716S, RSM606720S, RSM606731S, RSM606739S, RSM606748S, RSM606749S, RSM606750S, RSM606752S, RSM606761S, RSM606763S, RSM606764S, RSM606779S, RSM606793S, RSM606807S, RSM606810S, RSM606811S, RSM606812S, RSM606813S, RSM606818S, RSM606819S, RSM606820S, RSM606823S, RSM606826S, RSM606827S, RSM606837S, RSM606845S, RSM606855S, RSM606859S, RSM606865S, RSM606867S, RSM606869S, RSM606870S, RSM606871S, RSM606872S, RSM606874S, RSM606875S, RSM606876S, RSM606877S, RSM606880S, RSM606881S, RSM606883S, RSM606884S, RSM606885S, RSM606886S, RSM606887S, RSM606889S, RSM606891S, RSM606892S, RSM606895S, RSM606898S, RSM606900S, RSM606902S, RSM606905S, RSM606908S, RSM606910S, RSM606911S, RSM606912S, RSM606914S, RSM606915S, RSM606916S, RSM606918S, RSM606919S, RSM606920S, RSM606921S, RSM606922S, RSM606925S, RSM606927S, RSM606929S, RSM606931S, RSM606932S, RSM606933S, RSM606935S, RSM606936S, RSM606938S, RSM606939S, RSM606943S, RSM606945S, RSM606946S, RSM606951S, RSM606953S, RSM606954S, RSM606956S, RSM606957S, RSM606960S, RSM606963S, RSM606964S, RSM606965S, RSM606966S, RSM606969S, RSM606973S, RSM606975S, RSM606979S, RSM606980S, RSM606983S, RSM606985S, RSM606987S, RSM606991S, RSM606993S, RSM606994S, RSM606996S, RSM606997S, RSM606999S, RSM607000S, RSM607002S, RSM607003S, RSM607005S, RSM607013S, RSM607015S, RSM607017S, RSM607018S, RSM607020S, RSM607021S, RSM607026S, RSM607032S, RSM607033S, RSM607039S, RSM607040S, RSM607041S, RSM607042S, RSM607043S, RSM607048S, RSM607050S, RSM607051S, RSM607053S, RSM607056S, RSM607059S, RSM607060S, RSM607061S, RSM607062S, RSM607065S, RSM607066S, RSM607067S, RSM607069S, RSM607075S, RSM607076S, RSM607077S, RSM607079S, RSM607084S, RSM607086S, RSM607088S, RSM607093S, RSM607100S, RSM607101S, RSM607121S, RSM607123S, RSM607124S, RSM607125S, RSM607128S, RSM607129S, RSM607131S, RSM607135S, RSM607136S, RSM607138S, RSM607144S, RSM607146S, RSM607147S, RSM607148S, RSM607150S, RSM607151S, RSM607153S, RSM607158S, RSM607159S, RSM607164S, RSM607169S, RSM607170S, RSM607171S, RSM607172S, RSM607173S, RSM607174S, RSM607175S, RSM607176S, RSM607177S, RSM607178S, RSM607179S, RSM607182S, RSM607189S, RSM607316S, RSM607385S, RSM607451S, RSM607481S, RSM607608S, RSM607610S, RSM607611S, RSM607612S, RSM607614S, RSM607615S, RSM607616S, RSM607618S, RSM607619S, RSM607621S, RSM607622S, RSM607628S, RSM607631S, RSM607632S, RSM607634S, RSM607636S, RSM607637S, RSM607638S, RSM607639S, RSM607640S, RSM607641S, RSM607643S, RSM607645S, RSM607647S, RSM607648S, RSM607649S, RSM607651S, RSM607654S, RSM607655S, RSM607656S, RSM607657S, RSM607658S, RSM607660S, RSM607661S, RSM607663S, RSM607664S, RSM607665S, RSM607666S, RSM607667S, RSM607668S, RSM607672S, RSM607675S, RSM607678S, RSM607679S, RSM607680S, RSM607682S, RSM607727S, RSM607730S, RSM607731S, RSM607735S, RSM607736S, RSM607738S, RSM607740S, RSM607742S, RSM607744S, RSM607746S, RSM607748S, RSM607750S, RSM607751S, RSM607752S, RSM607753S, RSM607755S, RSM607756S, RSM607758S, RSM607761S, RSM607762S, RSM607763S, RSM607764S, RSM607770S, RSM607771S, RSM607772S, RSM607773S, RSM607774S, RSM607775S, RSM607777S, RSM607778S, RSM607780S, RSM607781S, RSM607782S, RSM607783S, RSM607784S, RSM607786S, RSM607787S, RSM607788S, RSM607789S, RSM607792S, RSM607794S, RSM607795S, RSM607798S, RSM607800S, RSM607801S, RSM607802S, RSM607803S, RSM607804S, RSM607805S, RSM607816S, RSM607818S, RSM607832S, RSM607837S, RSM607839S, RSM607869S, RSM607871S, RSM607879S, RSM607882S, RSM607883S, RSM607890S, RSM607899S, RSM608005S, RSM608024S, RSM608071S, RSM608258S, RSM608316S, RSM608336S, RSM608340S, RSM608347S, RSM608355S, RSM608363S, RSM608365S, RSM608377S, RSM608387S, RSM608423S, RSM608425S, RSM608426S, RSM608428S, RSM608429S, RSM608430S, RSM608433S, RSM608434S, RSM608436S, RSM608437S, RSM608438S, RSM608439S, RSM608440S, RSM608441S, RSM608445S, RSM608446S, RSM608447S, RSM608448S, RSM608449S, RSM608451S, RSM608454S, RSM608455S, RSM608457S, RSM608458S, RSM608459S, RSM608460S, RSM608461S, RSM608462S, RSM608463S, RSM608464S, RSM608465S, RSM608467S, RSM608468S, RSM608471S, RSM608472S, RSM608474S, RSM608476S, RSM608477S, RSM608479S, RSM608480S, RSM608481S, RSM608482S, RSM608483S, RSM608484S, RSM608489S, RSM608490S, RSM608543S, RSM608544S, RSM608545S, RSM608552S, RSM608567S, RSM608569S, RSM608572S, RSM608573S, RSM608574S, RSM608575S, RSM608578S, RSM608580S, RSM608581S, RSM608583S, RSM608585S, RSM608588S, RSM608589S, RSM608590S, RSM608592S, RSM608597S, RSM608602S, RSM608603S, RSM608605S, RSM608606S, RSM608610S, RSM608614S, RSM608615S, RSM608616S, RSM608618S, RSM608624S, RSM608630S, RSM608631S, RSM608633S, RSM608634S, RSM608635S, RSM608636S, RSM608638S, RSM608640S, RSM608641S, RSM608643S, RSM608644S, RSM608645S, RSM608646S, RSM608647S, RSM608650S, RSM608652S, RSM608654S, RSM608655S, RSM608657S, RSM608658S, RSM608659S, RSM608660S, RSM608661S, RSM608662S, RSM608663S, RSM608664S, RSM608666S, RSM608668S, RSM608670S, RSM608671S, RSM608672S, RSM608673S, RSM608676S, RSM608677S, RSM608678S, RSM608681S, RSM608683S, RSM608722S, RSM608790S, RSM608938S, RSM608940S, RSM609017S, RSM609018S, RSM609019S, RSM609037S, RSM609040S, RSM609041S, RSM609044S, RSM609045S, RSM609046S, RSM609047S, RSM609048S, RSM609049S, RSM609054S, RSM609056S, RSM609059S, RSM609063S, RSM609066S, RSM609067S, RSM609068S, RSM609071S, RSM609072S, RSM609073S, RSM609074S, RSM609076S, RSM609077S, RSM609078S, RSM609079S, RSM609080S, RSM609081S, RSM609082S, RSM609083S, RSM609086S, RSM609089S, RSM609091S, RSM609092S, RSM609096S, RSM609101S, RSM609108S, RSM609112S, RSM609113S, RSM609114S, RSM609116S, RSM609120S, RSM609121S, RSM609124S, RSM609125S, RSM609128S, RSM609130S, RSM609131S, RSM609132S, RSM609133S, RSM609134S, RSM609136S, RSM609137S, RSM609138S, RSM609139S, RSM609140S, RSM609141S, RSM609143S, RSM609144S, RSM609145S, RSM609146S, RSM609147S, RSM609150S, RSM609151S, RSM609153S, RSM609155S, RSM609157S, RSM609158S, RSM609160S, RSM609162S, RSM609163S, RSM609164S, RSM609165S, RSM609166S, RSM609167S, RSM609168S, RSM609169S, RSM609171S, RSM609172S, RSM609173S, RSM609174S, RSM609175S, RSM609176S, RSM609177S, RSM609178S, RSM609179S, RSM609181S, RSM609183S, RSM609184S, RSM609185S, RSM609187S, RSM609188S, RSM609189S, RSM609190S, RSM609191S, RSM609192S, RSM609193S, RSM609194S, RSM609195S, RSM609200S, RSM609201S, RSM609204S, RSM609206S, RSM609208S, RSM609209S, RSM609210S, RSM609211S, RSM609212S, RSM609213S, RSM609214S, RSM609215S, RSM609216S, RSM609217S, RSM609220S, RSM609221S, RSM609223S, RSM609224S, RSM609225S, RSM609227S, RSM609228S, RSM609229S, RSM609232S, RSM609234S, RSM609236S, RSM609237S, RSM609239S, RSM609240S, RSM609241S, RSM609242S, RSM609243S, RSM609244S, RSM609245S, RSM609247S, RSM609248S, RSM609249S, RSM609255S, RSM609257S, RSM609259S, RSM609260S, RSM609262S, RSM609285S, RSM609341S, RSM609361S, RSM609380S, RSM609388S, RSM609411S, RSM609413S, RSM609414S, RSM609415S, RSM609416S, RSM609417S, RSM609418S, RSM609422S, RSM609423S, RSM609425S, RSM609426S, RSM609427S, RSM609432S, RSM609433S, RSM609434S, RSM609435S, RSM609437S, RSM609438S, RSM609439S, RSM609444S, RSM609446S, RSM609447S, RSM609448S, RSM609449S, RSM609467S, RSM609477S, RSM609504S, RSM609535S, RSM609538S, RSM609539S, RSM609541S, RSM609542S, RSM609543S, RSM609544S, RSM609545S, RSM609546S, RSM609547S, RSM609548S, RSM609550S, RSM609551S, RSM609552S, RSM609553S, RSM609555S, RSM609556S, RSM609557S, RSM609558S, RSM609559S, RSM609560S, RSM609561S, RSM609562S, RSM609563S, RSM609564S, RSM609566S, RSM609567S, RSM609568S, RSM609569S, RSM609570S, RSM609571S, RSM609572S, RSM609573S, RSM609574S, RSM609575S, RSM609576S, RSM609577S, RSM609578S, RSM609579S, RSM609580S, RSM609581S, RSM609582S, RSM609583S, RSM609585S, RSM609586S, RSM609587S, RSM609588S, RSM609589S, RSM609590S, RSM609592S, RSM609593S, RSM609594S, RSM609596S, RSM609597S, RSM609598S, RSM609599S, RSM609600S, RSM609602S, RSM609603S, RSM609604S, RSM609605S, RSM609606S, RSM609607S, RSM609608S, RSM609609S, RSM609610S, RSM609611S, RSM609612S, RSM609613S, RSM609614S, RSM609615S, RSM609616S, RSM609617S, RSM609618S, RSM609619S, RSM609620S, RSM609622S, RSM609623S, RSM609624S, RSM609625S, RSM609626S, RSM609627S, RSM609629S, RSM609630S, RSM609631S, RSM609632S, RSM609633S, RSM609634S, RSM609635S, RSM609636S, RSM609637S, RSM609638S, RSM609639S, RSM609640S, RSM609641S, RSM609643S, RSM609644S, RSM609645S, RSM609647S, RSM609648S, RSM609649S, RSM609650S, RSM609651S, RSM609652S, RSM609654S, RSM609655S, RSM609656S, RSM609659S, RSM609660S, RSM609661S, RSM609662S, RSM609663S, RSM609666S, RSM609667S, RSM609668S, RSM609671S, RSM609674S, RSM609675S, RSM609676S, RSM609677S, RSM609678S, RSM609680S, RSM609681S, RSM609682S, RSM609683S, RSM609684S, RSM609686S, RSM609687S, RSM609689S, RSM609692S, RSM609694S, RSM609698S, RSM609700S, RSM609703S, RSM609705S, RSM609708S, RSM609710S, RSM609711S, RSM609712S, RSM609715S, RSM609716S, RSM609720S, RSM609722S, RSM609723S, RSM609726S, RSM609728S, RSM609729S, RSM609730S, RSM609732S, RSM609734S, RSM609736S, RSM609740S, RSM609743S, RSM609747S, RSM609748S, RSM609750S, RSM609751S, RSM609755S, RSM609759S, RSM610089S, RSM610090S, RSM610120S, RSM610125S, RSM610128S, RSM610129S, RSM610134S, RSM610136S, RSM610138S, RSM610142S, RSM610143S, RSM610146S, RSM610147S, RSM610149S, RSM610152S, RSM610153S, RSM610154S, RSM610155S, RSM610156S, RSM610158S, RSM610159S, RSM610161S, RSM610162S, RSM610163S, RSM610164S, RSM610165S, RSM610166S, RSM610167S, RSM610170S, RSM610171S, RSM610172S, RSM610173S, RSM610174S, RSM610175S, RSM610177S, RSM610178S, RSM610179S, RSM610181S, RSM610182S, RSM610183S, RSM610184S, RSM610185S, RSM610187S, RSM610188S, RSM610189S, RSM610190S, RSM610192S, RSM610194S, RSM610195S, RSM610197S, RSM610198S, RSM610199S, RSM610200S, RSM610201S, RSM610202S, RSM610203S, RSM610204S, RSM610207S, RSM610208S, RSM610209S, RSM610210S, RSM610211S, RSM610212S, RSM610213S, RSM610215S, RSM610217S, RSM610218S, RSM610219S, RSM610220S, RSM610221S, RSM610222S, RSM610229S, RSM610231S, RSM610237S, RSM610238S, RSM610240S, RSM610241S, RSM610244S, RSM610246S, RSM610249S, RSM610250S, RSM610254S, RSM610255S, RSM610257S, RSM610258S, RSM610265S, RSM610273S, RSM610301S, RSM610356S, RSM610358S, RSM610363S, RSM610366S, RSM610367S, RSM610369S, RSM610370S, RSM610373S, RSM610375S, RSM610381S, RSM610464S, RSM610465S, RSM610468S, RSM610477S, RSM610483S, RSM610488S, RSM610489S, RSM610495S, RSM610497S, RSM610504S, RSM610516S, RSM610518S, RSM610520S, RSM610521S, RSM610524S, RSM610529S, RSM610533S, RSM610534S, RSM610539S, RSM610543S, RSM610547S, RSM610556S, RSM610581S, RSM610583S, RSM611127S, RSM611128S, RSM611215S, RSM611276S, RSM611318S, RSM611321S, RSM611322S, RSM611323S, RSM611324S, RSM611325S, RSM611326S, RSM611327S, RSM611328S, RSM611329S, RSM611330S, RSM611331S, RSM611332S, RSM611333S, RSM611334S, RSM611335S, RSM611340S, RSM611341S, RSM611342S, RSM611344S, RSM611351S, RSM611352S, RSM611354S, RSM611355S, RSM611356S, RSM611357S, RSM611358S, RSM611360S, RSM611361S, RSM611362S, RSM611368S, RSM611370S, RSM611371S, RSM611372S, RSM611390S, RSM611392S, RSM611396S, RSM611398S, RSM611400S, RSM611401S, RSM611402S, RSM611410S, RSM611411S, RSM611412S, RSM611423S, RSM611429S, RSM611439S, RSM611440S, RSM611450S, RSM611452S, RSM611453S, RSM611461S, RSM611478S, RSM611479S, RSM611482S, RSM611484S, RSM611504S, RSM611505S, RSM611507S, RSM611509S, RSM611510S, RSM611511S, RSM611518S, RSM611519S, RSM611550S, RSM611554S, RSM611555S, RSM611557S, RSM611574S, RSM611577S, RSM611584S, RSM611597S, RSM611601S, RSM611796S, RSM611827S, RSM611855S, RSM611860S, RSM611873S, RSM611931S, RSM611945S, RSM611946S, RSM611959S, RSM611962S, RSM611969S, RSM611970S, RSM611971S, RSM611972S, RSM611976S, RSM611980S, RSM611987S, RSM611990S, RSM611992S, RSM611993S, RSM611995S, RSM611996S, RSM611999S, RSM612000S, RSM612001S, RSM612003S, RSM612004S, RSM612005S, RSM612006S, RSM612007S, RSM612009S, RSM612010S, RSM612013S, RSM612025S, RSM612027S, RSM612028S, RSM612032S, RSM612037S, RSM612053S, RSM612054S, RSM612057S, RSM612059S, RSM612064S, RSM612066S, RSM612069S, RSM612072S, RSM612075S, RSM612076S, RSM612077S, RSM612079S, RSM612080S, RSM612088S, RSM612089S, RSM612095S, RSM612096S, RSM612097S, RSM612098S, RSM612104S, RSM612105S, RSM612107S, RSM612111S, RSM612112S, RSM612115S, RSM612116S, RSM612119S, RSM612120S, RSM612121S, RSM612122S, RSM612124S, RSM612125S, RSM612126S, RSM612128S, RSM612129S, RSM612131S, RSM612132S, RSM612138S, RSM612140S, RSM612148S, RSM612150S, RSM612157S, RSM612164S, RSM612172S, RSM612177S, RSM612178S, RSM612187S, RSM612197S, RSM612198S, RSM612206S, RSM612208S, RSM612316S, RSM612329S, RSM612330S, RSM612370S, RSM612371S, RSM612374S, RSM612414S, RSM612416S, RSM612420S, RSM612421S, RSM612428S, RSM612430S, RSM612433S, RSM612439S, RSM612466S, RSM612473S, RSM612480S, RSM612481S, RSM612483S, RSM612486S, RSM613666S, RSM613670S, RSM613778S 回収対象数量 :22台 患者モニタリング対象数量:3,635台 出荷時期 :令和2年7月8日から令和4年6月6日 (2)Cobalt MRI CRT-Dシリーズ モデル番号: DTPA2D1, DTPA2D4, DTPA2Q1, DTPA2QQ 回収対象製造番号: RTC604035S, RTC604255S, RTC606873S, RTC608344S, RTC608633S, RTD601868S, RTD601877S, RTG601241S, RTG601829S, RTG602061S, RTG602258S, RTG602669S, RTG602697S, RTG602724S, RTH602129S, RTH602327S, RTH602409S 患者モニタリング対象製造番号: RTC601082S, RTC601083S, RTC601085S, RTC601086S, RTC601087S, RTC601088S, RTC601090S, RTC601093S, RTC601094S, RTC601097S, RTC601099S, RTC601100S, RTC601101S, RTC601102S, RTC601103S, RTC601104S, RTC601106S, RTC601107S, RTC601108S, RTC601109S, RTC601110S, RTC601111S, RTC601112S, RTC601113S, RTC601114S, RTC601115S, RTC601116S, RTC601117S, RTC601118S, RTC601119S, RTC601121S, RTC601123S, RTC601124S, RTC601125S, RTC601126S, RTC601127S, RTC601128S, RTC601129S, RTC601131S, RTC601132S, RTC601134S, RTC601137S, RTC601138S, RTC601141S, RTC601142S, RTC601144S, RTC601145S, RTC601146S, RTC601147S, RTC601149S, RTC601150S, RTC601151S, RTC601152S, RTC601153S, RTC601154S, RTC601157S, RTC601158S, RTC601160S, RTC601161S, RTC601162S, RTC601163S, RTC601164S, RTC601167S, RTC601168S, RTC601170S, RTC601172S, RTC601173S, RTC601175S, RTC601176S, RTC601177S, RTC601178S, RTC601179S, RTC601180S, RTC601181S, RTC601182S, RTC601183S, RTC601184S, RTC601185S, RTC601186S, RTC601187S, RTC601188S, RTC601189S, RTC601190S, RTC601191S, RTC601192S, RTC601193S, RTC601194S, RTC601195S, RTC601196S, RTC601197S, RTC601198S, RTC601199S, RTC601200S, RTC601201S, RTC601202S, RTC601203S, RTC601204S, RTC601205S, RTC601206S, RTC601207S, RTC601208S, RTC601209S, RTC601210S, RTC601211S, RTC601212S, RTC601213S, RTC601214S, RTC601215S, RTC601216S, RTC601217S, RTC601218S, RTC601219S, RTC601221S, RTC601222S, RTC601223S, RTC601224S, RTC601225S, RTC601226S, RTC601227S, RTC601228S, RTC601229S, RTC601231S, RTC601232S, RTC601234S, RTC601235S, RTC601236S, RTC601237S, RTC601239S, RTC601240S, RTC601241S, RTC601242S, RTC601243S, RTC601244S, RTC601245S, RTC601246S, RTC601248S, RTC601249S, RTC601250S, RTC601251S, RTC601252S, RTC601253S, RTC601254S, RTC601255S, RTC601256S, RTC601259S, RTC601261S, RTC601262S, RTC601263S, RTC601264S, RTC601265S, RTC601266S, RTC601267S, RTC601268S, RTC601269S, RTC601270S, RTC601271S, RTC601273S, RTC601274S, RTC601275S, RTC601276S, RTC601277S, RTC601279S, RTC601280S, RTC601281S, RTC601282S, RTC601283S, RTC601284S, RTC601285S, RTC601287S, RTC601288S, RTC601290S, RTC601291S, RTC601292S, RTC601293S, RTC601297S, RTC601298S, RTC601299S, RTC601301S, RTC601303S, RTC601304S, RTC601305S, RTC601308S, RTC601309S, RTC601310S, RTC601311S, RTC601312S, RTC601313S, RTC601314S, RTC601316S, RTC601317S, RTC601318S, RTC601319S, RTC601323S, RTC601324S, RTC601326S, RTC601327S, RTC601328S, RTC601331S, RTC601332S, RTC601333S, RTC601334S, RTC601336S, RTC601337S, RTC601338S, RTC601339S, RTC601340S, RTC601341S, RTC601342S, RTC601343S, RTC601344S, RTC601345S, RTC601346S, RTC601348S, RTC601349S, RTC601351S, RTC601352S, RTC601353S, RTC601354S, RTC601355S, RTC601356S, RTC601357S, RTC601358S, RTC601359S, RTC601360S, RTC601361S, RTC601362S, RTC601363S, RTC601366S, RTC601367S, RTC601368S, RTC601371S, RTC601372S, RTC601373S, RTC601375S, RTC601376S, RTC601379S, RTC601381S, RTC601383S, RTC601384S, RTC601386S, RTC601387S, RTC601389S, RTC601390S, RTC601391S, RTC601392S, RTC601393S, RTC601394S, RTC601395S, RTC601396S, RTC601397S, RTC601399S, RTC601400S, RTC601401S, RTC601402S, RTC601403S, RTC601404S, RTC601405S, RTC601407S, RTC601408S, RTC601409S, RTC601410S, RTC601412S, RTC601413S, RTC601414S, RTC601416S, RTC601417S, RTC601418S, RTC601419S, RTC601420S, RTC601421S, RTC601423S, RTC601424S, RTC601425S, RTC601427S, RTC601428S, RTC601429S, RTC601430S, RTC601431S, RTC601432S, RTC601433S, RTC601434S, RTC601435S, RTC601436S, RTC601438S, RTC601439S, RTC601442S, RTC601443S, RTC601444S, RTC601445S, RTC601446S, RTC601447S, RTC601448S, RTC601449S, RTC601450S, RTC601451S, RTC601452S, RTC601453S, RTC601454S, RTC601455S, RTC601456S, RTC601457S, RTC601458S, RTC601459S, RTC601460S, RTC601461S, RTC601463S, RTC601464S, RTC601467S, RTC601469S, RTC601470S, RTC601471S, RTC601472S, RTC601473S, RTC601474S, RTC601475S, RTC601476S, RTC601478S, RTC601482S, RTC601483S, RTC601484S, RTC601485S, RTC601486S, RTC601487S, RTC601488S, RTC601489S, RTC601490S, RTC601493S, RTC601494S, RTC601495S, RTC601496S, RTC601806S, RTC601889S, RTC601892S, RTC601895S, RTC601897S, RTC601899S, RTC601901S, RTC601902S, RTC601942S, RTC601979S, RTC601987S, RTC601991S, RTC601992S, RTC602003S, RTC602004S, RTC602005S, RTC602008S, RTC602011S, RTC602012S, RTC602016S, RTC602017S, RTC602018S, RTC602019S, RTC602022S, RTC602025S, RTC602026S, RTC602027S, RTC602028S, RTC602029S, RTC602030S, RTC602031S, RTC602033S, RTC602034S, RTC602035S, RTC602037S, RTC602038S, RTC602039S, RTC602040S, RTC602041S, RTC602042S, RTC602043S, RTC602044S, RTC602046S, RTC602047S, RTC602048S, RTC602049S, RTC602050S, RTC602051S, RTC602053S, RTC602054S, RTC602055S, RTC602056S, RTC602057S, RTC602058S, RTC602059S, RTC602060S, RTC602061S, RTC602062S, RTC602063S, RTC602064S, RTC602065S, RTC602067S, RTC602068S, RTC602069S, RTC602070S, RTC602071S, RTC602072S, RTC602073S, RTC602075S, RTC602076S, RTC602078S, RTC602079S, RTC602080S, RTC602082S, RTC602083S, RTC602084S, RTC602085S, RTC602086S, RTC602087S, RTC602088S, RTC602090S, RTC602091S, RTC602092S, RTC602093S, RTC602095S, RTC602096S, RTC602097S, RTC602098S, RTC602099S, RTC602100S, RTC602101S, RTC602103S, RTC602104S, RTC602105S, RTC602107S, RTC602108S, RTC602109S, RTC602110S, RTC602111S, RTC602112S, RTC602113S, RTC602114S, RTC602115S, RTC602116S, RTC602118S, RTC602119S, RTC602120S, RTC602121S, RTC602122S, RTC602123S, RTC602124S, RTC602126S, RTC602127S, RTC602128S, RTC602130S, RTC602131S, RTC602132S, RTC602133S, RTC602134S, RTC602135S, RTC602136S, RTC602137S, RTC602139S, RTC602140S, RTC602141S, RTC602143S, RTC602145S, RTC602146S, RTC602147S, RTC602148S, RTC602150S, RTC602151S, RTC602152S, RTC602153S, RTC602154S, RTC602155S, RTC602156S, RTC602157S, RTC602158S, RTC602159S, RTC602160S, RTC602161S, RTC602162S, RTC602163S, RTC602164S, RTC602165S, RTC602167S, RTC602168S, RTC602169S, RTC602170S, RTC602171S, RTC602172S, RTC602173S, RTC602174S, RTC602175S, RTC602176S, RTC602177S, RTC602178S, RTC602179S, RTC602180S, RTC602181S, RTC602182S, RTC602183S, RTC602184S, RTC602186S, RTC602188S, RTC602190S, RTC602191S, RTC602192S, RTC602193S, RTC602194S, RTC602195S, RTC602196S, RTC602197S, RTC602198S, RTC602199S, RTC602200S, RTC602201S, RTC602202S, RTC602203S, RTC602204S, RTC602211S, RTC602212S, RTC602213S, RTC602214S, RTC602215S, RTC602216S, RTC602217S, RTC602218S, RTC602219S, RTC602220S, RTC602225S, RTC602230S, RTC602231S, RTC602233S, RTC602235S, RTC602247S, RTC602251S, RTC602254S, RTC602257S, RTC602294S, RTC602298S, RTC602303S, RTC602306S, RTC602309S, RTC602319S, RTC602327S, RTC602333S, RTC602347S, RTC602348S, RTC602453S, RTC602455S, RTC602456S, RTC602457S, RTC602458S, RTC602462S, RTC602479S, RTC602481S, RTC602486S, RTC602492S, RTC602494S, RTC602497S, RTC602498S, RTC602501S, RTC602503S, RTC602506S, RTC602511S, RTC602514S, RTC602516S, RTC602517S, RTC602523S, RTC602525S, RTC602528S, RTC602535S, RTC602537S, RTC602546S, RTC602548S, RTC602550S, RTC602551S, RTC602558S, RTC602559S, RTC602567S, RTC602568S, RTC602576S, RTC602577S, RTC602579S, RTC602580S, RTC602582S, RTC602588S, RTC602591S, RTC602592S, RTC602593S, RTC602594S, RTC602595S, RTC602598S, RTC602599S, RTC602600S, RTC602603S, RTC602605S, RTC602615S, RTC602620S, RTC602626S, RTC602632S, RTC602635S, RTC602642S, RTC602645S, RTC602650S, RTC602652S, RTC602654S, RTC602659S, RTC602660S, RTC602663S, RTC602665S, RTC602667S, RTC602682S, RTC602684S, RTC602701S, RTC602704S, RTC602705S, RTC602713S, RTC602714S, RTC602716S, RTC602718S, RTC602727S, RTC602746S, RTC602758S, RTC602820S, RTC602863S, RTC602955S, RTC602957S, RTC602960S, RTC602969S, RTC603009S, RTC603012S, RTC603014S, RTC603016S, RTC603018S, RTC603023S, RTC603034S, RTC603044S, RTC603051S, RTC603054S, RTC603055S, RTC603056S, RTC603064S, RTC603065S, RTC603069S, RTC603072S, RTC603078S, RTC603079S, RTC603081S, RTC603082S, RTC603089S, RTC603102S, RTC603104S, RTC603108S, RTC603112S, RTC603119S, RTC603121S, RTC603133S, RTC603134S, RTC603138S, RTC603142S, RTC603143S, RTC603148S, RTC603149S, RTC603166S, RTC603170S, RTC603174S, RTC603179S, RTC603205S, RTC603206S, RTC603209S, RTC603210S, RTC603211S, RTC603218S, RTC603231S, RTC603242S, RTC603248S, RTC603249S, RTC603253S, RTC603254S, RTC603255S, RTC603256S, RTC603257S, RTC603259S, RTC603260S, RTC603261S, RTC603262S, RTC603264S, RTC603265S, RTC603266S, RTC603267S, RTC603268S, RTC603269S, RTC603271S, RTC603272S, RTC603274S, RTC603275S, RTC603276S, RTC603277S, RTC603278S, RTC603279S, RTC603280S, RTC603281S, RTC603282S, RTC603283S, RTC603284S, RTC603285S, RTC603286S, RTC603287S, RTC603288S, RTC603290S, RTC603291S, RTC603292S, RTC603293S, RTC603294S, RTC603295S, RTC603296S, RTC603298S, RTC603299S, RTC603300S, RTC603301S, RTC603302S, RTC603303S, RTC603312S, RTC603316S, RTC603317S, RTC603319S, RTC603322S, RTC603323S, RTC603324S, RTC603325S, RTC603326S, RTC603327S, RTC603328S, RTC603329S, RTC603330S, RTC603331S, RTC603332S, RTC603333S, RTC603334S, RTC603337S, RTC603338S, RTC603339S, RTC603340S, RTC603341S, RTC603342S, RTC603344S, RTC603345S, RTC603347S, RTC603349S, RTC603350S, RTC603351S, RTC603352S, RTC603353S, RTC603354S, RTC603356S, RTC603358S, RTC603359S, RTC603360S, RTC603361S, RTC603363S, RTC603364S, RTC603365S, RTC603366S, RTC603367S, RTC603369S, RTC603371S, RTC603372S, RTC603373S, RTC603378S, RTC603379S, RTC603382S, RTC603400S, RTC603438S, RTC603460S, RTC603471S, RTC603481S, RTC603482S, RTC603483S, RTC603485S, RTC603488S, RTC603493S, RTC603496S, RTC603502S, RTC603503S, RTC603504S, RTC603506S, RTC603508S, RTC603510S, RTC603513S, RTC603516S, RTC603524S, RTC603525S, RTC603528S, RTC603529S, RTC603532S, RTC603535S, RTC603539S, RTC603545S, RTC603547S, RTC603565S, RTC603568S, RTC603575S, RTC603579S, RTC603586S, RTC603587S, RTC603598S, RTC603600S, RTC603605S, RTC603627S, RTC603631S, RTC603632S, RTC603635S, RTC603637S, RTC603638S, RTC603639S, RTC603640S, RTC603641S, RTC603642S, RTC603645S, RTC603646S, RTC603647S, RTC603648S, RTC603649S, RTC603650S, RTC603651S, RTC603652S, RTC603653S, RTC603655S, RTC603657S, RTC603658S, RTC603660S, RTC603661S, RTC603664S, RTC603666S, RTC603667S, RTC603668S, RTC603670S, RTC603675S, RTC603676S, RTC603677S, RTC603678S, RTC603679S, RTC603680S, RTC603681S, RTC603684S, RTC603686S, RTC603690S, RTC603691S, RTC603692S, RTC603694S, RTC603695S, RTC603698S, RTC603702S, RTC603703S, RTC603705S, RTC603706S, RTC603707S, RTC603709S, RTC603710S, RTC603712S, RTC603717S, RTC603720S, RTC603721S, RTC603722S, RTC603724S, RTC603725S, RTC603727S, RTC603728S, RTC603729S, RTC603730S, RTC603732S, RTC603733S, RTC603734S, RTC603735S, RTC603736S, RTC603738S, RTC603739S, RTC603741S, RTC603743S, RTC603748S, RTC603750S, RTC603754S, RTC603755S, RTC603756S, RTC603760S, RTC603763S, RTC603764S, RTC603765S, RTC603848S, RTC603852S, RTC603853S, RTC603859S, RTC603880S, RTC603881S, RTC603886S, RTC603887S, RTC603906S, RTC603927S, RTC603955S, RTC603956S, RTC603957S, RTC603958S, RTC603959S, RTC603961S, RTC603962S, RTC603965S, RTC603966S, RTC603967S, RTC603968S, RTC603969S, RTC603970S, RTC603971S, RTC603972S, RTC603973S, RTC603974S, RTC603975S, RTC603976S, RTC603977S, RTC603978S, RTC603979S, RTC603980S, RTC603981S, RTC603982S, RTC603986S, RTC603988S, RTC603989S, RTC603991S, RTC603997S, RTC603998S, RTC604000S, RTC604001S, RTC604002S, RTC604003S, RTC604006S, RTC604007S, RTC604008S, RTC604009S, RTC604011S, RTC604013S, RTC604014S, RTC604015S, RTC604017S, RTC604018S, RTC604019S, RTC604020S, RTC604021S, RTC604022S, RTC604023S, RTC604024S, RTC604025S, RTC604026S, RTC604027S, RTC604028S, RTC604030S, RTC604031S, RTC604032S, RTC604033S, RTC604034S, RTC604036S, RTC604037S, RTC604038S, RTC604039S, RTC604041S, RTC604042S, RTC604043S, RTC604044S, RTC604045S, RTC604046S, RTC604047S, RTC604048S, RTC604049S, RTC604050S, RTC604051S, RTC604052S, RTC604053S, RTC604054S, RTC604055S, RTC604056S, RTC604057S, RTC604058S, RTC604059S, RTC604060S, RTC604061S, RTC604062S, RTC604063S, RTC604064S, RTC604065S, RTC604066S, RTC604068S, RTC604069S, RTC604070S, RTC604071S, RTC604072S, RTC604073S, RTC604074S, RTC604075S, RTC604076S, RTC604077S, RTC604078S, RTC604079S, RTC604080S, RTC604081S, RTC604082S, RTC604083S, RTC604084S, RTC604085S, RTC604086S, RTC604087S, RTC604088S, RTC604089S, RTC604090S, RTC604091S, RTC604092S, RTC604093S, RTC604094S, RTC604095S, RTC604096S, RTC604098S, RTC604099S, RTC604100S, RTC604101S, RTC604103S, RTC604104S, RTC604105S, RTC604106S, RTC604107S, RTC604108S, RTC604109S, RTC604110S, RTC604111S, RTC604112S, RTC604113S, RTC604114S, RTC604115S, RTC604116S, RTC604117S, RTC604118S, RTC604119S, RTC604120S, RTC604121S, RTC604122S, RTC604123S, RTC604124S, RTC604125S, RTC604127S, RTC604128S, RTC604129S, RTC604130S, RTC604257S, RTC604259S, RTC604270S, RTC604274S, RTC604286S, RTC604307S, RTC604309S, RTC604311S, RTC604320S, RTC604322S, RTC604328S, RTC604329S, RTC604345S, RTC604348S, RTC604355S, RTC604366S, RTC604368S, RTC604374S, RTC604376S, RTC604394S, RTC604400S, RTC604401S, RTC604404S, RTC604407S, RTC604409S, RTC604412S, RTC604413S, RTC604414S, RTC604417S, RTC604421S, RTC604428S, RTC604429S, RTC604433S, RTC604436S, RTC604441S, RTC604442S, RTC604450S, RTC604463S, RTC604464S, RTC604473S, RTC604480S, RTC604509S, RTC604581S, RTC604596S, RTC604598S, RTC604599S, RTC604600S, RTC604601S, RTC604602S, RTC604603S, RTC604604S, RTC604605S, RTC604606S, RTC604607S, RTC604608S, RTC604609S, RTC604610S, RTC604611S, RTC604612S, RTC604613S, RTC604614S, RTC604615S, RTC604616S, RTC604617S, RTC604618S, RTC604619S, RTC604620S, RTC604621S, RTC604622S, RTC604623S, RTC604624S, RTC604625S, RTC604626S, RTC604627S, RTC604628S, RTC604629S, RTC604630S, RTC604631S, RTC604633S, RTC604634S, RTC604635S, RTC604636S, RTC604637S, RTC604638S, RTC604640S, RTC604641S, RTC604642S, RTC604644S, RTC604645S, RTC604646S, RTC604647S, RTC604648S, RTC604649S, RTC604650S, RTC604653S, RTC604654S, RTC604655S, RTC604656S, RTC604657S, RTC604658S, RTC604659S, RTC604660S, RTC604663S, RTC604664S, RTC604665S, RTC604666S, RTC604667S, RTC604668S, RTC604669S, RTC604670S, RTC604672S, RTC604673S, RTC604674S, RTC604675S, RTC604677S, RTC604678S, RTC604679S, RTC604682S, RTC604683S, RTC604684S, RTC604686S, RTC604688S, RTC604690S, RTC604691S, RTC604692S, RTC604693S, RTC604695S, RTC604696S, RTC604698S, RTC604699S, RTC604700S, RTC604701S, RTC604702S, RTC604705S, RTC604706S, RTC604707S, RTC604708S, RTC604710S, RTC604711S, RTC604712S, RTC604713S, RTC604714S, RTC604715S, RTC604717S, RTC604721S, RTC604725S, RTC604726S, RTC604727S, RTC604729S, RTC604730S, RTC604731S, RTC604732S, RTC604733S, RTC604734S, RTC604735S, RTC604736S, RTC604737S, RTC604740S, RTC604741S, RTC604742S, RTC604744S, RTC604747S, RTC604748S, RTC604749S, RTC604750S, RTC604751S, RTC604754S, RTC604755S, RTC604759S, RTC604760S, RTC604761S, RTC604765S, RTC604766S, RTC604767S, RTC604768S, RTC604771S, RTC604774S, RTC604778S, RTC604781S, RTC604782S, RTC604786S, RTC604787S, RTC604788S, RTC604790S, RTC604791S, RTC604792S, RTC604793S, RTC604794S, RTC604795S, RTC604798S, RTC604799S, RTC604800S, RTC604801S, RTC605621S, RTC605623S, RTC605788S, RTC605816S, RTC605946S, RTC605967S, RTC605973S, RTC605978S, RTC605998S, RTC606007S, RTC606019S, RTC606020S, RTC606025S, RTC606033S, RTC606042S, RTC606181S, RTC606220S, RTC606221S, RTC606222S, RTC606224S, RTC606225S, RTC606226S, RTC606227S, RTC606228S, RTC606230S, RTC606231S, RTC606232S, RTC606233S, RTC606234S, RTC606235S, RTC606236S, RTC606237S, RTC606238S, RTC606240S, RTC606241S, RTC606243S, RTC606244S, RTC606245S, RTC606246S, RTC606247S, RTC606249S, RTC606251S, RTC606252S, RTC606253S, RTC606254S, RTC606255S, RTC606256S, RTC606257S, RTC606258S, RTC606259S, RTC606260S, RTC606261S, RTC606262S, RTC606263S, RTC606264S, RTC606265S, RTC606266S, RTC606310S, RTC606353S, RTC606458S, RTC606575S, RTC606624S, RTC606666S, RTC606669S, RTC606674S, RTC606694S, RTC606698S, RTC606703S, RTC606710S, RTC606711S, RTC606712S, RTC606720S, RTC606733S, RTC606739S, RTC606747S, RTC606753S, RTC606757S, RTC606761S, RTC606762S, RTC606764S, RTC606800S, RTC606807S, RTC606830S, RTC606836S, RTC606839S, RTC606844S, RTC606847S, RTC606848S, RTC606849S, RTC606851S, RTC606858S, RTC606863S, RTC606875S, RTC606889S, RTC606894S, RTC606902S, RTC606949S, RTC606958S, RTC606960S, RTC607024S, RTC607026S, RTC607030S, RTC607034S, RTC607038S, RTC607040S, RTC607049S, RTC607059S, RTC607060S, RTC607064S, RTC607065S, RTC607066S, RTC607233S, RTC607234S, RTC607239S, RTC607241S, RTC607243S, RTC607244S, RTC607245S, RTC607246S, RTC607247S, RTC607249S, RTC607252S, RTC607255S, RTC607259S, RTC607273S, RTC607279S, RTC607281S, RTC607286S, RTC607289S, RTC607293S, RTC607296S, RTC607298S, RTC607301S, RTC607304S, RTC607306S, RTC607308S, RTC607314S, RTC607324S, RTC607330S, RTC607334S, RTC607339S, RTC607345S, RTC607356S, RTC607357S, RTC607360S, RTC607365S, RTC607397S, RTC607398S, RTC607399S, RTC607403S, RTC607404S, RTC607416S, RTC607417S, RTC607418S, RTC607419S, RTC607420S, RTC607421S, RTC607425S, RTC607429S, RTC607430S, RTC607432S, RTC607439S, RTC607442S, RTC607443S, RTC607444S, RTC607445S, RTC607446S, RTC607447S, RTC607449S, RTC607450S, RTC607452S, RTC607453S, RTC607454S, RTC607455S, RTC607459S, RTC607460S, RTC607465S, RTC607468S, RTC607471S, RTC607477S, RTC607478S, RTC607480S, RTC607483S, RTC607496S, RTC607498S, RTC607500S, RTC607501S, RTC607502S, RTC607504S, RTC607506S, RTC607507S, RTC607508S, RTC607509S, RTC607511S, RTC607513S, RTC607515S, RTC607517S, RTC607529S, RTC607531S, RTC607532S, RTC607533S, RTC607536S, RTC607537S, RTC607540S, RTC607546S, RTC607562S, RTC607564S, RTC607572S, RTC607573S, RTC607576S, RTC607580S, RTC607583S, RTC607588S, RTC607589S, RTC607592S, RTC607600S, RTC607603S, RTC607612S, RTC607615S, RTC607661S, RTC607693S, RTC607717S, RTC607730S, RTC607772S, RTC607780S, RTC607812S, RTC607889S, RTC607891S, RTC607892S, RTC607893S, RTC607897S, RTC607902S, RTC607903S, RTC607907S, RTC607908S, RTC607911S, RTC607914S, RTC607917S, RTC607919S, RTC607925S, RTC607926S, RTC607928S, RTC607929S, RTC607938S, RTC607940S, RTC607948S, RTC607950S, RTC607951S, RTC607953S, RTC607956S, RTC607957S, RTC607958S, RTC607960S, RTC607961S, RTC607962S, RTC607963S, RTC607964S, RTC607967S, RTC607969S, RTC607977S, RTC607978S, RTC607979S, RTC607980S, RTC607981S, RTC607982S, RTC607984S, RTC607985S, RTC607986S, RTC607987S, RTC607988S, RTC607989S, RTC607990S, RTC607991S, RTC607993S, RTC607995S, RTC607996S, RTC608006S, RTC608012S, RTC608013S, RTC608014S, RTC608015S, RTC608017S, RTC608018S, RTC608024S, RTC608026S, RTC608031S, RTC608034S, RTC608056S, RTC608071S, RTC608074S, RTC608075S, RTC608076S, RTC608077S, RTC608078S, RTC608082S, RTC608091S, RTC608099S, RTC608105S, RTC608109S, RTC608110S, RTC608111S, RTC608112S, RTC608113S, RTC608114S, RTC608115S, RTC608116S, RTC608117S, RTC608118S, RTC608120S, RTC608121S, RTC608122S, RTC608123S, RTC608124S, RTC608127S, RTC608128S, RTC608129S, RTC608130S, RTC608131S, RTC608132S, RTC608133S, RTC608134S, RTC608135S, RTC608136S, RTC608137S, RTC608138S, RTC608140S, RTC608141S, RTC608142S, RTC608143S, RTC608144S, RTC608145S, RTC608146S, RTC608147S, RTC608148S, RTC608149S, RTC608150S, RTC608152S, RTC608153S, RTC608154S, RTC608156S, RTC608157S, RTC608158S, RTC608159S, RTC608162S, RTC608165S, RTC608166S, RTC608167S, RTC608168S, RTC608174S, RTC608175S, RTC608176S, RTC608178S, RTC608180S, RTC608181S, RTC608183S, RTC608184S, RTC608185S, RTC608226S, RTC608271S, RTC608274S, RTC608276S, RTC608277S, RTC608290S, RTC608292S, RTC608294S, RTC608295S, RTC608296S, RTC608299S, RTC608304S, RTC608305S, RTC608306S, RTC608307S, RTC608308S, RTC608321S, RTC608339S, RTC608346S, RTC608354S, RTC608357S, RTC608358S, RTC608359S, RTC608392S, RTC608393S, RTC608398S, RTC608401S, RTC608402S, RTC608404S, RTC608415S, RTC608416S, RTC608417S, RTC608420S, RTC608422S, RTC608423S, RTC608428S, RTC608429S, RTC608430S, RTC608432S, RTC608433S, RTC608435S, RTC608437S, RTC608439S, RTC608440S, RTC608442S, RTC608443S, RTC608444S, RTC608445S, RTC608448S, RTC608449S, RTC608450S, RTC608451S, RTC608452S, RTC608455S, RTC608457S, RTC608460S, RTC608463S, RTC608464S, RTC608465S, RTC608467S, RTC608468S, RTC608469S, RTC608470S, RTC608471S, RTC608472S, RTC608474S, RTC608475S, RTC608477S, RTC608478S, RTC608479S, RTC608482S, RTC608483S, RTC608485S, RTC608486S, RTC608487S, RTC608488S, RTC608489S, RTC608490S, RTC608491S, RTC608493S, RTC608497S, RTC608498S, RTC608499S, RTC608500S, RTC608502S, RTC608504S, RTC608505S, RTC608511S, RTC608512S, RTC608513S, RTC608515S, RTC608516S, RTC608517S, RTC608522S, RTC608525S, RTC608527S, RTC608529S, RTC608531S, RTC608533S, RTC608534S, RTC608535S, RTC608537S, RTC608538S, RTC608539S, RTC608542S, RTC608543S, RTC608544S, RTC608546S, RTC608547S, RTC608548S, RTC608550S, RTC608551S, RTC608552S, RTC608554S, RTC608559S, RTC608560S, RTC608561S, RTC608564S, RTC608565S, RTC608566S, RTC608572S, RTC608573S, RTC608574S, RTC608576S, RTC608577S, RTC608578S, RTC608579S, RTC608580S, RTC608582S, RTC608583S, RTC608584S, RTC608585S, RTC608586S, RTC608599S, RTC608600S, RTC608602S, RTC608604S, RTC608608S, RTC608611S, RTC608613S, RTC608618S, RTC608619S, RTC608628S, RTC608630S, RTC608631S, RTC608632S, RTC608634S, RTC608640S, RTC608676S, RTC608677S, RTC608679S, RTC608682S, RTC608683S, RTC608697S, RTC608700S, RTC608701S, RTC608702S, RTC608705S, RTC608708S, RTC608713S, RTC608716S, RTC608717S, RTC608729S, RTC608731S, RTC608733S, RTC608739S, RTC608741S, RTC608752S, RTC608753S, RTC608754S, RTC608756S, RTC608759S, RTC608760S, RTC608761S, RTC608766S, RTC608767S, RTC608769S, RTC608770S, RTC608771S, RTC608773S, RTC608774S, RTC608775S, RTC608777S, RTC608781S, RTC608783S, RTC608784S, RTC608785S, RTC608786S, RTC608789S, RTC608790S, RTC608791S, RTC608792S, RTC608794S, RTC608795S, RTC608796S, RTC608797S, RTC608811S, RTC608815S, RTC608816S, RTC608841S, RTC609215S, RTC609232S, RTC609256S, RTC609307S, RTC609308S, RTC609312S, RTC609313S, RTC609315S, RTC609316S, RTC609317S, RTC609319S, RTC609322S, RTC609323S, RTC609324S, RTC609325S, RTC609327S, RTC609328S, RTC609329S, RTC609332S, RTC609334S, RTC609336S, RTC609339S, RTC609341S, RTC609347S, RTC609348S, RTC609353S, RTC609354S, RTC609355S, RTC609357S, RTC609363S, RTC609364S, RTC609366S, RTC609367S, RTC609368S, RTC609380S, RTC609383S, RTC609384S, RTC609385S, RTC609387S, RTC609388S, RTC609389S, RTC609391S, RTC609392S, RTC609393S, RTC609394S, RTC609395S, RTC609397S, RTC609398S, RTC609400S, RTC609401S, RTC609402S, RTC609403S, RTC609404S, RTC609405S, RTC609409S, RTC609412S, RTC609413S, RTC609415S, RTC609416S, RTC609417S, RTC609422S, RTC609423S, RTC609426S, RTC609428S, RTC609429S, RTC609430S, RTC609434S, RTC609437S, RTC609441S, RTC609443S, RTC609447S, RTC609449S, RTC609451S, RTC609453S, RTC609454S, RTC609455S, RTC609457S, RTC609486S, RTC609664S, RTC609665S, RTC609673S, RTC609675S, RTC609677S, RTC609679S, RTC609680S, RTC609681S, RTC609737S, RTC609776S, RTC609795S, RTC609796S, RTC609798S, RTC609799S, RTC609800S, RTC609801S, RTC609802S, RTC609804S, RTC609805S, RTC609806S, RTC609808S, RTC609809S, RTC609810S, RTC609812S, RTC609815S, RTC609816S, RTC609826S, RTC609827S, RTC609828S, RTC609831S, RTC609832S, RTC609834S, RTC609835S, RTC609836S, RTC609841S, RTC609843S, RTC609845S, RTC609859S, RTC609860S, RTC609861S, RTC609862S, RTC609863S, RTC609864S, RTC609865S, RTC609875S, RTC609876S, RTC609877S, RTC609878S, RTC609880S, RTC609881S, RTC609882S, RTC609883S, RTC609884S, RTC609885S, RTC609887S, RTC609888S, RTC609889S, RTC609890S, RTC609891S, RTC609892S, RTC609894S, RTC609896S, RTC609897S, RTC609898S, RTC609899S, RTC609900S, RTC609902S, RTC609903S, RTC609904S, RTC609905S, RTC609906S, RTC609907S, RTC609909S, RTC609910S, RTC609917S, RTC609936S, RTC610052S, RTC610487S, RTC610499S, RTC610500S, RTC610502S, RTC610503S, RTC610505S, RTC610506S, RTC611094S, RTC611148S, RTC611149S, RTC611150S, RTC611154S, RTC611155S, RTC611156S, RTC611158S, RTC611161S, RTC611162S, RTC611163S, RTC611165S, RTC611166S, RTC611168S, RTC611169S, RTC611171S, RTC611174S, RTC611176S, RTC611179S, RTC611181S, RTC611182S, RTC611189S, RTC611192S, RTC611194S, RTC611196S, RTC611197S, RTC611200S, RTC611201S, RTC611202S, RTC611203S, RTC611204S, RTC611206S, RTC611217S, RTC611225S, RTC611226S, RTC611227S, RTC611232S, RTC611233S, RTC611236S, RTC611237S, RTC611244S, RTC611245S, RTC611246S, RTC611248S, RTC611249S, RTC611255S, RTC611256S, RTC611257S, RTC611259S, RTC611261S, RTC611262S, RTC611263S, RTC611264S, RTC611265S, RTC611266S, RTC611267S, RTC611269S, RTC611273S, RTC611275S, RTC611277S, RTC611278S, RTC611280S, RTC611281S, RTC611292S, RTC611294S, RTC611300S, RTC611302S, RTC611306S, RTC611307S, RTC611311S, RTC611318S, RTC611324S, RTC611326S, RTC611329S, RTC611338S, RTC611408S, RTC611410S, RTC611411S, RTC611412S, RTC611415S, RTC611416S, RTC611417S, RTC611420S, RTC611422S, RTC611427S, RTC611430S, RTC611434S, RTC611440S, RTC611444S, RTC611445S, RTC611447S, RTC611448S, RTC611451S, RTC611452S, RTC611453S, RTC611454S, RTC611456S, RTC611458S, RTC611461S, RTC611464S, RTC611465S, RTC611466S, RTC611467S, RTC611468S, RTC611469S, RTC611475S, RTC611476S, RTC611481S, RTC611482S, RTC611485S, RTC611487S, RTC611491S, RTC611493S, RTC611494S, RTC611496S, RTC611504S, RTC611506S, RTC611510S, RTC611512S, RTC611513S, RTC611514S, RTC611520S, RTC611521S, RTC611522S, RTC611523S, RTC611525S, RTC611526S, RTC611527S, RTC611530S, RTC611531S, RTC611532S, RTC611534S, RTC611535S, RTC611536S, RTC611537S, RTC611539S, RTC611540S, RTC611541S, RTC611542S, RTC611546S, RTC611547S, RTC611548S, RTC611550S, RTC611551S, RTC611552S, RTC611553S, RTC611555S, RTC611558S, RTC611559S, RTC611561S, RTC611562S, RTC611563S, RTC611565S, RTC611566S, RTC611572S, RTC611577S, RTC611578S, RTC611580S, RTC611582S, RTC611583S, RTC611587S, RTC611588S, RTC611590S, RTC611593S, RTC611594S, RTC611595S, RTC611596S, RTC611600S, RTC611601S, RTC611607S, RTC611612S, RTC611624S, RTC611626S, RTC611634S, RTC611636S, RTC611637S, RTC611639S, RTC611642S, RTC611644S, RTC611647S, RTC611648S, RTC611649S, RTC611650S, RTC611651S, RTC611653S, RTC611654S, RTC611655S, RTC611698S, RTC611700S, RTC611703S, RTC611725S, RTC611752S, RTC611758S, RTC611769S, RTC611775S, RTC611796S, RTC611803S, RTC611805S, RTC611806S, RTC611808S, RTC611811S, RTC611836S, RTC611845S, RTC611853S, RTC611855S, RTC611858S, RTC611859S, RTC611860S, RTC611861S, RTC611862S, RTC611864S, RTC611867S, RTC611914S, RTC611918S, RTC611926S, RTC611931S, RTC611932S, RTC611933S, RTC611935S, RTC611937S, RTC611940S, RTC611941S, RTC611944S, RTC611946S, RTC611953S, RTC611958S, RTC611964S, RTC611969S, RTC611974S, RTC611978S, RTC611983S, RTC611984S, RTC611988S, RTC611989S, RTC611992S, RTC611993S, RTC611994S, RTC611996S, RTC611997S, RTC611998S, RTC611999S, RTC612005S, RTC612006S, RTC612007S, RTC612009S, RTC612011S, RTC612013S, RTC612014S, RTC612015S, RTC612016S, RTC612017S, RTC612020S, RTC612022S, RTC612024S, RTC612028S, RTC612029S, RTC612031S, RTC612032S, RTC612033S, RTC612036S, RTC612042S, RTC612043S, RTC612048S, RTC612049S, RTC612050S, RTC612052S, RTC612053S, RTC612055S, RTC612056S, RTC612058S, RTC612067S, RTC612068S, RTC612072S, RTC612075S, RTC612076S, RTC612077S, RTC612085S, RTC612087S, RTC612088S, RTC612089S, RTC612097S, RTC612098S, RTC612101S, RTC612113S, RTC612116S, RTC612117S, RTC612119S, RTC612125S, RTC612129S, RTC612139S, RTC612140S, RTC612141S, RTC612143S, RTC612147S, RTC612152S, RTC612153S, RTC612154S, RTC612155S, RTC612156S, RTC612157S, RTC612158S, RTC612159S, RTC612160S, RTC612161S, RTC612162S, RTC612163S, RTC612164S, RTC612166S, RTC612167S, RTC612168S, RTC612169S, RTC612170S, RTC612171S, RTC612172S, RTC612173S, RTC612174S, RTC612175S, RTC612176S, RTC612177S, RTC612178S, RTC612179S, RTC612181S, RTC612182S, RTC612183S, RTC612184S, RTC612185S, RTC612186S, RTC612187S, RTC612188S, RTC612189S, RTC612190S, RTC612191S, RTC612192S, RTC612193S, RTC612194S, RTC612195S, RTC612196S, RTC612197S, RTC612198S, RTC612199S, RTC612200S, RTC612201S, RTC612202S, RTC612203S, RTC612204S, RTC612205S, RTC612206S, RTC612207S, RTC612208S, RTC612210S, RTC612211S, RTC612212S, RTC612213S, RTC612214S, RTC612215S, RTC612217S, RTC612220S, RTC612221S, RTC612222S, RTC612223S, RTC612225S, RTC612226S, RTC612227S, RTC612228S, RTC612229S, RTC612230S, RTC612231S, RTC612232S, RTC612233S, RTC612234S, RTC612236S, RTC612237S, RTC612238S, RTC612239S, RTC612241S, RTC612242S, RTC612243S, RTC612246S, RTC612248S, RTC612249S, RTC612250S, RTC612251S, RTC612252S, RTC612253S, RTC612255S, RTC612256S, RTC612257S, RTC612259S, RTC612260S, RTC612261S, RTC612262S, RTC612263S, RTC612264S, RTC612266S, RTC612267S, RTC612268S, RTC612269S, RTC612271S, RTC612272S, RTC612274S, RTC612275S, RTC612276S, RTC612277S, RTC612281S, RTC612284S, RTC612287S, RTC612289S, RTC612292S, RTC612294S, RTC612295S, RTC612296S, RTC612301S, RTC612302S, RTC612304S, RTC612305S, RTC612306S, RTC612307S, RTC612310S, RTC612313S, RTC612314S, RTC612315S, RTC612440S, RTC612455S, RTC612632S, RTC612637S, RTC612638S, RTC612642S, RTC612643S, RTC612644S, RTC612656S, RTC612660S, RTC612685S, RTC612686S, RTC612689S, RTC612692S, RTC612701S, RTC612702S, RTC612703S, RTC612705S, RTC612706S, RTC612724S, RTC612726S, RTC612728S, RTC612732S, RTC612733S, RTC612737S, RTC612738S, RTC612740S, RTC612742S, RTC612743S, RTC612744S, RTC612745S, RTC612747S, RTC612748S, RTC612749S, RTC612750S, RTC612751S, RTC612752S, RTC612753S, RTC612755S, RTC612756S, RTC612757S, RTC612758S, RTC612759S, RTC612761S, RTC612762S, RTC612763S, RTC612766S, RTC612768S, RTC612832S, RTC613010S, RTC613051S, RTC613059S, RTC613068S, RTC613105S, RTC613106S, RTC613149S, RTC613163S, RTC613212S, RTC613223S, RTC613228S, RTC613253S, RTC613261S, RTC613278S, RTC613324S, RTC613330S, RTC613403S, RTC613408S, RTC613638S, RTC613785S, RTC614015S, RTC614016S, RTC614021S, RTC614210S, RTC614523S, RTC614528S, RTC614639S, RTC614681S, RTC614795S, RTC614798S, RTC614799S, RTC614800S, RTC614801S, RTC614803S, RTC614809S, RTC614812S, RTC614814S, RTC614817S, RTC614823S, RTC614824S, RTC614829S, RTC614830S, RTC614832S, RTC614836S, RTC614837S, RTC614839S, RTC614843S, RTC614845S, RTC614856S, RTC614857S, RTC614858S, RTC614864S, RTC614865S, RTC614867S, RTC614879S, RTC614880S, RTC614881S, RTC614882S, RTC614883S, RTC614884S, RTC614885S, RTC614886S, RTC614889S, RTC614892S, RTC614896S, RTC614897S, RTC614900S, RTC614901S, RTC614904S, RTC614905S, RTC614906S, RTC614908S, RTC614909S, RTC614910S, RTC614911S, RTC614912S, RTC614915S, RTC614917S, RTC614920S, RTC614921S, RTC614922S, RTC614923S, RTC614924S, RTC614926S, RTC614927S, RTC614928S, RTC614929S, RTC614930S, RTC614931S, RTC614932S, RTC614935S, RTC614936S, RTC614937S, RTC614939S, RTC614941S, RTC614942S, RTC614943S, RTC614944S, RTC614946S, RTC614948S, RTC614949S, RTC614950S, RTC614951S, RTC614952S, RTC614953S, RTC614954S, RTC614955S, RTC614956S, RTC614957S, RTC614958S, RTC614959S, RTC614960S, RTC614963S, RTC614964S, RTC614965S, RTC614966S, RTC614968S, RTC614969S, RTC614970S, RTC614971S, RTC614972S, RTC614975S, RTC614977S, RTC614982S, RTC614983S, RTC614984S, RTC614986S, RTC614988S, RTC614991S, RTC614992S, RTC614993S, RTC614996S, RTC614997S, RTC614998S, RTC614999S, RTC615000S, RTC615001S, RTC615011S, RTC615012S, RTC615014S, RTC615016S, RTC615018S, RTC615020S, RTC615023S, RTC615030S, RTC615041S, RTC615051S, RTC615074S, RTC615076S, RTC615109S, RTC615141S, RTC615149S, RTC615198S, RTC615275S, RTC615301S, RTC615309S, RTC615310S, RTC615311S, RTC615316S, RTC615320S, RTC615321S, RTC615322S, RTC615324S, RTC615325S, RTC615352S, RTC615363S, RTC615368S, RTC615369S, RTC615370S, RTC615388S, RTC615404S, RTC615476S, RTC615478S, RTC615944S, RTC615947S, RTC616013S, RTC616023S, RTC616035S, RTC616041S, RTC616062S, RTC616063S, RTC616077S, RTC616079S, RTC616086S, RTC616093S, RTC616100S, RTC616937S, RTD600559S, RTD600566S, RTD600574S, RTD600614S, RTD600615S, RTD600618S, RTD600622S, RTD600624S, RTD600626S, RTD600627S, RTD600779S, RTD600925S, RTD600926S, RTD600962S, RTD600987S, RTD600988S, RTD600989S, RTD600991S, RTD600993S, RTD600995S, RTD600996S, RTD600997S, RTD601065S, RTD601080S, RTD601100S, RTD601102S, RTD601104S, RTD601111S, RTD601112S, RTD601113S, RTD601114S, RTD601115S, RTD601117S, RTD601118S, RTD601119S, RTD601121S, RTD601122S, RTD601124S, RTD601125S, RTD601128S, RTD601129S, RTD601130S, RTD601131S, RTD601132S, RTD601135S, RTD601138S, RTD601139S, RTD601140S, RTD601141S, RTD601144S, RTD601145S, RTD601146S, RTD601148S, RTD601150S, RTD601153S, RTD601154S, RTD601155S, RTD601157S, RTD601158S, RTD601160S, RTD601161S, RTD601162S, RTD601164S, RTD601165S, RTD601166S, RTD601167S, RTD601169S, RTD601171S, RTD601172S, RTD601173S, RTD601174S, RTD601181S, RTD601195S, RTD601196S, RTD601389S, RTD601513S, RTD601527S, RTD601528S, RTD601537S, RTD601560S, RTD601606S, RTD601607S, RTD601618S, RTD601620S, RTD601623S, RTD601625S, RTD601866S, RTD601869S, RTD601871S, RTD601883S, RTD601887S, RTD601892S, RTD601905S, RTD602125S, RTD602154S, RTD602155S, RTD602159S, RTD602164S, RTD602172S, RTD602179S, RTD602228S, RTD602235S, RTD602238S, RTD602243S, RTD602244S, RTD602256S, RTD602257S, RTD602259S, RTD602260S, RTD602262S, RTD602264S, RTD602265S, RTD602270S, RTD602271S, RTD602275S, RTD602308S, RTD602314S, RTD602315S, RTD602408S, RTD602422S, RTD602428S, RTD602481S, RTD602486S, RTD602493S, RTD602494S, RTD602496S, RTG600302S, RTG600337S, RTG600344S, RTG600353S, RTG600355S, RTG600357S, RTG600362S, RTG600365S, RTG600371S, RTG600376S, RTG600379S, RTG600382S, RTG600387S, RTG600389S, RTG600391S, RTG600392S, RTG600396S, RTG600397S, RTG600407S, RTG600409S, RTG600410S, RTG600413S, RTG600414S, RTG600416S, RTG600417S, RTG600419S, RTG600420S, RTG600421S, RTG600422S, RTG600425S, RTG600427S, RTG600429S, RTG600431S, RTG600432S, RTG600433S, RTG600439S, RTG600442S, RTG600775S, RTG600776S, RTG600777S, RTG600778S, RTG600785S, RTG600789S, RTG600790S, RTG600797S, RTG600802S, RTG600804S, RTG600807S, RTG600814S, RTG600815S, RTG600818S, RTG600822S, RTG600823S, RTG600824S, RTG600825S, RTG600826S, RTG600827S, RTG600828S, RTG600830S, RTG600832S, RTG600833S, RTG600835S, RTG600836S, RTG600847S, RTG600861S, RTG600868S, RTG600870S, RTG600875S, RTG600992S, RTG600994S, RTG600995S, RTG601007S, RTG601009S, RTG601012S, RTG601025S, RTG601029S, RTG601031S, RTG601032S, RTG601040S, RTG601091S, RTG601093S, RTG601094S, RTG601095S, RTG601097S, RTG601099S, RTG601100S, RTG601101S, RTG601104S, RTG601105S, RTG601107S, RTG601110S, RTG601113S, RTG601114S, RTG601115S, RTG601116S, RTG601117S, RTG601118S, RTG601119S, RTG601120S, RTG601121S, RTG601122S, RTG601123S, RTG601124S, RTG601125S, RTG601126S, RTG601128S, RTG601131S, RTG601132S, RTG601133S, RTG601188S, RTG601208S, RTG601214S, RTG601220S, RTG601233S, RTG601250S, RTG601251S, RTG601254S, RTG601259S, RTG601261S, RTG601266S, RTG601267S, RTG601278S, RTG601279S, RTG601280S, RTG601282S, RTG601290S, RTG601291S, RTG601297S, RTG601305S, RTG601311S, RTG601569S, RTG601823S, RTG601825S, RTG601835S, RTG601838S, RTG601946S, RTG601947S, RTG602003S, RTG602009S, RTG602010S, RTG602044S, RTG602050S, RTG602056S, RTG602062S, RTG602063S, RTG602068S, RTG602076S, RTG602084S, RTG602087S, RTG602091S, RTG602097S, RTG602098S, RTG602100S, RTG602101S, RTG602110S, RTG602252S, RTG602255S, RTG602256S, RTG602262S, RTG602276S, RTG602277S, RTG602280S, RTG602285S, RTG602293S, RTG602298S, RTG602323S, RTG602337S, RTG602343S, RTG602344S, RTG602402S, RTG602541S, RTG602570S, RTG602578S, RTG602597S, RTG602606S, RTG602612S, RTG602618S, RTG602629S, RTG602667S, RTG602715S, RTG602717S, RTG602727S, RTG602729S, RTG602746S, RTG602748S, RTG602749S, RTG602766S, RTG602786S, RTG602797S, RTG602799S, RTG602800S, RTG602805S, RTG602807S, RTG602808S, RTG602809S, RTG602810S, RTG602814S, RTG602815S, RTG602819S, RTG602823S, RTG602831S, RTG602833S, RTG602834S, RTG602838S, RTG602839S, RTG602840S, RTG602844S, RTG602845S, RTG602847S, RTG602848S, RTG602850S, RTG602852S, RTG602856S, RTG602858S, RTG602865S, RTG602984S, RTG603016S, RTG603026S, RTG603032S, RTG603047S, RTG603053S, RTG603056S, RTG603058S, RTG603060S, RTG603111S, RTG603112S, RTG603123S, RTG603148S, RTG603154S, RTG603158S, RTG603160S, RTH600724S, RTH600725S, RTH600726S, RTH600729S, RTH600731S, RTH600732S, RTH600735S, RTH600736S, RTH600737S, RTH600738S, RTH600742S, RTH600743S, RTH600744S, RTH600746S, RTH600747S, RTH600749S, RTH600751S, RTH600752S, RTH600755S, RTH600756S, RTH600757S, RTH600758S, RTH600766S, RTH600767S, RTH600769S, RTH600777S, RTH600783S, RTH600786S, RTH600794S, RTH600798S, RTH600799S, RTH600800S, RTH600801S, RTH600802S, RTH600803S, RTH600804S, RTH600805S, RTH600807S, RTH600808S, RTH600809S, RTH600810S, RTH600812S, RTH600814S, RTH600815S, RTH600816S, RTH600817S, RTH600818S, RTH600819S, RTH600820S, RTH600821S, RTH600822S, RTH600824S, RTH600825S, RTH600826S, RTH600827S, RTH600828S, RTH600829S, RTH600830S, RTH600831S, RTH600832S, RTH600833S, RTH600834S, RTH600837S, RTH600838S, RTH600839S, RTH600840S, RTH600841S, RTH600842S, RTH600843S, RTH600844S, RTH600846S, RTH600849S, RTH600850S, RTH600851S, RTH600853S, RTH600854S, RTH600855S, RTH600857S, RTH600859S, RTH600860S, RTH600862S, RTH600863S, RTH600868S, RTH600869S, RTH600870S, RTH600872S, RTH600873S, RTH600875S, RTH600876S, RTH600877S, RTH600878S, RTH600879S, RTH600881S, RTH600884S, RTH600887S, RTH600888S, RTH600889S, RTH600893S, RTH600897S, RTH600898S, RTH600905S, RTH600906S, RTH600918S, RTH600922S, RTH600923S, RTH600926S, RTH600937S, RTH600940S, RTH600944S, RTH600945S, RTH600947S, RTH600953S, RTH600956S, RTH600961S, RTH600965S, RTH600966S, RTH600969S, RTH600985S, RTH600990S, RTH600992S, RTH600995S, RTH601001S, RTH601027S, RTH601029S, RTH601030S, RTH601037S, RTH601053S, RTH601061S, RTH601095S, RTH601129S, RTH601132S, RTH601139S, RTH601142S, RTH601205S, RTH601206S, RTH601207S, RTH601208S, RTH601209S, RTH601210S, RTH601211S, RTH601212S, RTH601214S, RTH601216S, RTH601217S, RTH601218S, RTH601219S, RTH601220S, RTH601221S, RTH601222S, RTH601223S, RTH601224S, RTH601225S, RTH601226S, RTH601227S, RTH601228S, RTH601229S, RTH601230S, RTH601231S, RTH601232S, RTH601233S, RTH601234S, RTH601235S, RTH601236S, RTH601237S, RTH601238S, RTH601239S, RTH601240S, RTH601241S, RTH601242S, RTH601243S, RTH601244S, RTH601245S, RTH601246S, RTH601247S, RTH601248S, RTH601249S, RTH601250S, RTH601251S, RTH601252S, RTH601305S, RTH601308S, RTH601311S, RTH601312S, RTH601326S, RTH601327S, RTH601329S, RTH601332S, RTH601339S, RTH601346S, RTH601347S, RTH601348S, RTH601352S, RTH601355S, RTH601381S, RTH601385S, RTH601389S, RTH601958S, RTH601962S, RTH601964S, RTH601970S, RTH601974S, RTH601976S, RTH601979S, RTH601982S, RTH601986S, RTH601993S, RTH601996S, RTH602008S, RTH602009S, RTH602114S, RTH602115S, RTH602117S, RTH602119S, RTH602121S, RTH602123S, RTH602126S, RTH602127S, RTH602128S, RTH602130S, RTH602131S, RTH602289S, RTH602352S, RTH602360S, RTH602361S, RTH602363S, RTH602365S, RTH602366S, RTH602368S, RTH602369S, RTH602372S, RTH602374S, RTH602375S, RTH602377S, RTH602380S, RTH602381S, RTH602387S, RTH602391S, RTH602392S, RTH602394S, RTH602395S, RTH602399S, RTH602402S, RTH602408S, RTH602415S, RTH602417S, RTH602419S, RTH602467S, RTH602483S, RTH602487S, RTH602488S, RTH602491S, RTH602492S, RTH602493S, RTH602494S, RTH602496S, RTH602497S, RTH602498S, RTH602499S, RTH602500S, RTH602501S, RTH602502S, RTH602503S, RTH602504S, RTH602505S, RTH602506S, RTH602507S, RTH602508S, RTH602509S, RTH602511S, RTH602512S, RTH602513S, RTH602514S, RTH602516S, RTH602518S, RTH602519S, RTH602521S, RTH602544S, RTH602616S, RTH602710S, RTH602713S, RTH602717S, RTH602719S, RTH602720S, RTH602721S, RTH602728S, RTH602731S, RTH602733S, RTH602742S, RTH602750S, RTH602754S, RTH602763S, RTH602768S, RTH602770S, RTH602771S, RTH602773S, RTH602775S, RTH602780S, RTH602792S, RTH602801S, RTH602831S, RTH602845S, RTH602865S, RTH602866S, RTH602883S, RTH602884S, RTH602886S, RTH602887S, RTH602924S, RTH602927S, RTH602934S, RTH602944S 回収対象数量 :17台 患者モニタリング対象数量:3,278台 出荷時期 :令和2年10月29日から令和4年6月2日
製造販売業者等名称
製造販売業者の名称 : 日本メドトロニック株式会社 製造販売業者の所在地: 東京都港区港南一丁目2番70号 許可の種類 : 第一種医療機器製造販売業 許可番号 : 13B1X00261 製造業者: Medtronic, Inc.(メドトロニック社) 所在国 : 米国
回収理由
海外製造元より、Cobalt MRI ICDシリーズ及びCobalt MRI CRT-D シリーズにおいて、副次的に生じた微小電流 によって間欠的に短絡保護機能(SCP)が作動することで不十分なエネルギー(設定されたエネルギーの約79%) で高電圧治療が行われる可能性があるとの連絡を受領しました。 そのため、海外製造元でスクリーニング検査導入前に出荷された未使用の対象製品について自主回収することと しました。
危惧される具体的な健康被害
一連の高電圧治療の中で本事象が発生したとしても、対象製品は設計仕様通り、不整脈に対して1セット最大6回 までの高電圧治療を送出します。再度不整脈を検出した場合、2セット目の高電圧治療が送出されます。3セッ ト目以降も不整脈を再検出する限り、高電圧治療は継続します。本事象による高電圧治療の不成功は、実臨床上 においても確認されていません。また、高電圧治療と併用して、初期設定である抗頻拍ペーシング(ATP)機能に よる不整脈の治療が可能です。従って、本事象により不整脈が停止しない可能性は極めて低いと考えられるた め、重篤な健康被害に至る可能性はないと判断しています。
回収開始年月日
令和4年6月22日 情報提供の開始
効能・効果又は用途等
(1)Cobalt MRI ICDシリーズ 本品は、心室性頻拍等の治療を目的として、体内に植え込み、心室センシング、ペーシング、抗頻拍ペーシング 治療及び除細動を行う植込み型除細動器です。 なお、本品は撮像可能条件に適合する場合にのみ限定的にMRI検査を行うことが可能となる機器です。 (2)Cobalt MRI CRT-Dシリーズ 本品は、至適薬物療法が行われているにもかかわらず、症状が改善しない心不全患者の症状改善を目的としてい ます。 なお、本品は撮像可能条件に適合する場合にのみ限定的にMRI検査を行うことが可能となる機器です。
その他
対象製品の所在は全て把握しておりますので、該当する医療機関及び代理店に対して情報提供の上、回収を速や かに実施いたします。 本製品をお使いの患者への対応として、患者モニタリングを実施します。通常のフォローアップの中で患者ごと の評価により医師の判断において推奨設定値の適用をご検討いただきます。 医療機関での検討結果を弊社にて確認し、記録します。 なお、本事象を解消するためのソフトウェアの供給準備が整い次第、植込み能動型機器用プログラマを用いて対 象製品にソフトウェアアップデートを行い、改修を実施します。
担当者及び連絡先
担当者 : 品質・市販後安全管理 嶋田 稔之、志田 理絵 連絡先 : 日本メドトロニック株式会社 東京都港区港南一丁目2番70号 電話番号: 03-6776-0040 FAX番号 : 03-6774-4675