令和 3年 5月25日作成
一般的名称及び販売名
一般的名称: 電動式歯科用ファイル 販売名 : マニーNRTファイル
対象ロット、数量及び出荷時期
回収数量 834,642本 対象品番 GPR 1S, GPR 1S‐4N, GPR 2S, GPR 3N, GPR 4N 対象ロット H125142100, H125159600, H125159700, H125159800, H125163200, H125164400, H125164700, H125166700, H125 167000, H125167100, H125167200, H125167300, H125169100, H125169600, H125169700, H125169800, H1251710 00, H125171100, H125171200, H125171300, H125171400, H125172100, H125172700, H125175500, H125176100, H125176200, H125176300, H125176400, H125178600, H125178900, H125179000, H125179100, H125179600, H125 179700, H125179800, H125179900, H125180000, H125180100, H125180500, H125181300, H125185500, H1251856 00, H125187900, H125189000, H125189100, H125194700, H125194800, H125194900, H125195000, H125195100, H125195200, H125195300, H125195400, H125195500, H125195600, H125195700, H125195800, H135014700, H135 015100, H135015800, H135015900, H135016400, H135016500, H135019100, H135032400, H135032500, H1350326 00, H135032700, H135032900, H135033000, H135033100, H135038600, H135039400, H135040400, H135041100, H135042000, H135042100, H135042200, H135042300, H135043900, H135047900, H135048000, H135048100, H135 048200, H135049700, H135051000, H135051300, H135051400, H135051500, H135051600, H135051700, H1350518 00, H135051900, H135052000, H135052100, H135052200, H135052300, H135052400, H135052500, H135053300, H135053400, H135053500, H135054500, H135054600, H135054700, H135054800, H135054900, H135055000, H135 055200, H135055300, H135056400, H135056500, H135056600, H135056700, H135060100, H135062500, H1350626 00, H135062700, H135062800, H135062900, H135066200, H135066500, H135066600, H135066700, H135066800, H135066900, H135067100, H135069900, H135071400, H135088600, H135088700, H135088800, H135090300, H13K 000200, H13L055800, H14A020500, H14B005900, H14B027700, H14B027900, H14C009200, H14C021200, H14D0135 00, H14D018800, H14D038100, H14D038400, H14F001200, H14F008200, H14F008300, H14F008400, H14H019900, H14I010900, H14I011000, H14I051700, H14I051800, H14I053700, H14I055700, H14I056000, H14I056100, H14I 056300, H14I056400, H14K033400, H14L043400, H14L043500, H14L043600, H14L049100, H15A046400, H15A0465 00, H15A046700, H15A046800, H15A046900, H15A047000, H15A047100, H15B001000, H15C025100, H15C032700, H15C032800, H15C032900, H15C060900, H15D016400, H15D027100, H15D060200, H15E042200, H15E056800, H15E 056900, H15E057000, H15E057100, H15E057200, H15E057300, H15G017800, H15G068300, H15G068800, H15G0689 00, H15G069000, H15G069001, H15G075100, H15H042800, H15H042900, H15H043000, H15H043100, H15I050400, H15I052700, H15J061500, H15J061600, H15J093700, H15K006600, H15K011800, H15K011900, H15K038700, H15K 038800, H15K038900, H15K039000, H15K040900, H15K040901, H15K061100, H15L011300, H15L021600, H16A0440 00, H16B110500, H16B127200, H16C017200, H16C017300, H16C033000, H16C065400, H16C065500, H16D000900, H16D054100, H16D066500, H16D099900, H16E044100, H16E072300, H16F023500, H16F023600, H16F023700, H16F 023800, H16F023900, H16F060100, H16G010000, H16G070000, H16H068100, H16H068200, H16H068300, H16H0684 00, H16H068500, H16I008200, H16I081500, H16I081600, H16I081700, H16I081800, H16J013500, H16J013600, H16J016600, H16J088100, H16J088200, H16J088300, H16J088400, H16J088500, H16K065800, H16K065900, H16K 075400, H16K075500, H16K075600, H16K075700, H16K075800, H16K077900, H16K078000, H16K080000, H16K0801 00, H16K080200, H17A166900, H17A167000, H17A167200, H17A184200, H17A184300, H17A184600, H17A184700, H17A184800, H17A184900, H17B002400, H17B002500, H17C010000, H17C010100, H17C010200, H17C010300, H17C 148200, H17C148300, H17C148400, H17C148500, H17C148600, H17C154000, H17D110200, H17D110600, H17D1107 00, H17D110800, H17D120100, H17D125000, H17D125100, H17E000900, H17F001400, H17F001500, H17F001600, H17F001700, H17F001800, H17F008200, H17F036900, H17F067100, H17F102400, H17F102500, H17F127600, H17F 127700, H17G020400, H17G020500, H17G020600, H17G020800, H17G046700, H17G046800, H17G046900, H17G1033 00, H17G124100, H17G124300, H17G124400, H17G131500, H17H014000, H17H014100, H17H026500, H17H032100, H17H109300, H17J009900, H17J010000, H17J010100, H17J010200, H17J013000, H17J022000, H17J022300, H17J 022400, H17J054000, H17J054100, H17K001100, H17K001200, H17K001300, H17K001400, H17K001500, H17K0690 00, H17K069100, H17L069800, H17L069900, H17L070000, H17L070100, H17L070200, H17L089600, H17L089700, H17L089800, H17L089900, H17L118500, H17L119700, H17L119800, H17L131100, H17L131200, H17L131300, H17L 131400, H17L131500, H18A043800, H18A043900, H18A075200, H18A075300, H18A101200, H18A101300, H18A1014 00, H18B026200, H18B081400, H18C027200, H18C027300, H18C027400, H18C027500, H18C041100, H18C045400, H18C045500, H18C115900, H18C116000, H18C116100, H18C116200, H18C118100, H18C119800, H18D128700, H18D 128800, H18D128900, H18D129000, H18D129100, H18E028400, H18E028500, H18E061700, H18E061800, H18E0619 00, H18E062000, H18E062100, H18E095700, H18G010700, H18G010800, H18G010900, H18G011000, H18G011100, H18G061500, H18G140900, H18H000200, H18H000300, H18H000400, H18H000500, H18I024000, H18I024200, H18I 024300, H18I024400, H18I024500, H18I129600, H18I129700, H18I129800, H18I129900, H18J004500, H18J0444 00, H18J044900, H18J045000, H18J045100, H18J063300, H18J074200, H18J076000, H18J076100, H18J076200, H18J099200, H18K032300, H18K032400, H18K032500, H18K057200, H18K118000, H18K118400, H18L126500, H19A 031300, H19A031400, H19A031500, H19A058300, H19B002400, H19B002500, H19B002600, H19B003800, H19B0055 00, H19B013200, H19B013300, H19C014100, H19C014200, H19C014300, H19C014400, H19C014500, H19C115000, H19C115100, H19C115200, H19C115300, H19C115400, H19D039700, H19D057200, H19D066300, H19D079300, H19D 079400, H19D079500, H19D079600, H19E127400, H19E127500, H19E127600, H19E127700, H19E127800, H19G0511 00, H19G051200, H19G051300, H19G051400, H19G051500, H19H000100, H19H000200, H19H000300, H19H013000, H19H013100, H19H013200, H19H021300, H19H071800, H19H071900, H19H072000, H19H072300, H19H074000, H19I 004400, H19I091300, H19I091400, H19I091500, H19I091600, H19I091700, H19J073100, H19J073200, H19J0733 00, H19K018600, H19K018700, H19K095900, H19K096000, H19K096100, H19K096200, H19K096300, H19L011700, H19L032300, H19L032400, H19L032500, H19L032600, H19L097100, H20A010800, H20A036300, H20A072700, H20A 072800, H20A082800, H20A082900, H20A083000, H20B111300, H20C000600, H20C000700, H20C000800, H20C0009 00, H20C082700, H20C082800, H20C082900, H20C083000, H20C083100, H20D064400, H20D064500, H20D064600, H20D064700, H20D088200, H20E023500, H20E024600, H20E026600, H20E026700, H20E026800, H20E026900, H20E 027000, H20G001300, H20G001600, H20G001800, H20G001900, H20G002100, H20G002700, H20G113900, H20G1140 00, H20G114100, H20G114200, H20H009400, H20H009500, H20H020000, H20H020100, H20H020200, H20H020300, H20H020400, H20H058300, H20I039600, H20I041100, H20I041200, H20J213700, H20J213800, H20J213900, H20J 214000, H20J221000, H20J221100, H20K042000, H20K042100, H20K050600, H20L052600, H20L065200, H20L0653 00, H20L065400, H20L079700, H20L089200, H21A067500, H21A067600, H21A067700, H21A067800, H21A091500, H21A102900, H21C004100, H21C135900, H21C136000, H21C136100, H21C136200, H21C136300, H21C136400 出荷時期 平成24年9月〜令和3年4月
製造販売業者等名称
製造販売業者の名称 : マニー株式会社 製造販売業者の所在地: 栃木県宇都宮市清原工業団地8番3 許可の種類 : 第一種医療機器製造販売業 許可番号 : 09B1X00006
回収理由
本品の一般的名称は「電動式歯科用ファイル」であるが、本品の「NRT-GPR」という品番において、異なる一般 的名称(電動式歯科用螺旋状除去器)の使用目的が標榜されていることが社内の点検において判明したため、本 品番が含まれる製品の自主回収を実施いたします。
危惧される具体的な健康被害
本品番を含む本品を充填物の除去に使用した場合であっても、重篤な健康被害発生のおそれはありません。しか しながら、本品番を電動式歯科用ファイルとして使用した場合においては、有効性を十分に発揮できないおそれ があります。 なお、現在までに本品番の使用方法に起因する健康被害発生の報告はありません。
回収開始年月日
令和3年5月18日 (情報提供開始日)
効能・効果又は用途等
上下方向の往復運動又は引き上げ動作による切削・研磨で根管を拡大し、根管壁を平滑にするために能動型医療 機器に接続して用いる歯科用器具である。
その他
出荷先は全て特定されており、速やかに自主回収を行う旨を通知し、回収作業を行います。
担当者及び連絡先
担当者 : 品質保証グループ 金子典夫 連絡先 : マニー株式会社 栃木県宇都宮市清原工業団地8番3 電話番号: 028-667-1811 FAX番号 : 028-667-9267