令和 3年 4月 8日作成 令和 4年10月28日訂正(*)
一般的名称及び販売名
一般的名称: 注射筒輸液ポンプ 販売名 : (1)JMSシリンジポンプ SP‐520 (2)JMSシリンジポンプ SP‐120
対象ロット、数量及び出荷時期
(A)新規出荷時より対象バッテリが入っている装置 (1) 品番:JM-SP520 製造ロット:18I3180B,18I3181B,18I3182B,18I3183B,18I3184B,18I3185B,18I3186B, 18I3187B,18I3188B,18I3189B,18I3190B,18I3191B,18I3192B,18I3193B, 18I3194B,18I3195B,18I3196B,18I3197B,18I3198B,18I3199B,18I3200B, 18I3201B,18I3202B,18I3203B,18I3204B,18I3205B,18I3206B,18I3207B, 18I3208B,18I3209B,18I3210B,18I3211B,18I3212B,18I3213B,18I3214B, 18I3215B,18I3216B,18I3217B,18I3218B,18I3219B,18I3220B,18I3221B, 18I3222B,18I3223B,18I3224B,18I3225B,18I3226B,18I3227B,18I3228B, 18I3229B,18I3230B,18I3231B,18I3232B,18I3233B,18I3234B,18I3235B, 18I3236B,18I3237B,18I3238B,18I3239B,18I3240B,18I3241B,18I3242B, 18I3243B,18I3244B,18I3245B,18I3246B,18I3247B,18I3248B,18I3249B, 18I3250B,18I3251B,18I3252B,18I3253B,18I3254B,18I3255B,18I3256B, 18I3257B,18I3258B,18I3259B,18I3260B,18I3261B,18I3262B,18I3263B, 18I3264B,18I3265B,18I3266B,18I3267B,18I3268B,18I3269B,18I3270B, 18I3271B,18I3272B,18I3273B,18I3274B,18I3275B,18I3276B,18I3277B, 18I3278B,18I3279B,18L3280B,18L3281B,18L3282B,18L3283B,18L3284B, 18L3285B,18L3286B,18L3287B,18L3288B,18L3289B,18L3290B,18L3291B, 18L3292B,18L3293B,18L3294B,18L3295B,18L3296B,18L3297B,18L3298B, 18L3299B,18L3300B,18L3301B,18L3302B,18L3303B,18L3304B,18L3305B, 18L3306B,18L3307B,18L3308B,18L3309B,18L3310B,18L3311B,18L3312B, 18L3313B,18L3314B,18L3315B,18L3316B,18L3317B,18L3318B,18L3319B, 18L3320B,18L3321B,18L3322B,18L3323B,18L3324B,18L3325B,18L3326B, 18L3327B,18L3328B,18L3329B,18L3330B,18L3331B,18L3332B,18L3333B, 18L3334B,18L3335B,18L3336B,18L3337B,18L3338B,18L3339B,18L3340B, 18L3341B,18L3342B,18L3343B,18L3344B,18L3345B,18L3346B,18L3347B, 18L3348B,18L3349B,18L3350B,18L3351B,18L3352B,18L3353B,18L3354B, 18L3355B,18L3356B,18L3357B,18L3358B,18L3359B,18L3360B,18L3361B, 18L3362B,18L3363B,18L3364B,18L3365B,18L3366B,18L3367B,18L3368B, 18L3369B,18L3370B,18L3371B,18L3372B,18L3373B,18L3374B,18L3375B, 18L3376B,18L3377B,18L3378B,18L3379B,18L3380B,18L3381B,18L3382B, 18L3383B,18L3384B,18L3385B,18L3386B,18L3387B,18L3388B,18L3389B, 18L3390B,18L3391B,18L3392B,18L3393B,18L3394B,18L3395B,18L3396B, 18L3397B,18L3398B,18L3399B,18L3400B,18L3401B,18L3402B,18L3403B, 18L3404B,18L3405B,18L3406B,18L3407B,18L3408B,18L3409B,18L3410B, 18L3411B,18L3412B,18L3413B,18L3414B,18L3415B,18L3416B,18L3417B, 18L3418B,18L3419B,18L3420B,18L3421B,18L3422B,18L3423B,18L3424B, 18L3425B,18L3426B,18L3427B,18L3428B,18L3429B,18L3430B,18L3431B, 18L3432B,18L3433B,18L3434B,18L3435B,18L3436B,18L3437B,18L3438B, 18L3439B,18L3440B,18L3441B,18L3442B,18L3443B,18L3444B,18L3445B, 18L3446B,18L3447B,18L3448B,18L3449B,18L3450B,18L3451B,18L3452B, 18L3453B,18L3454B,18L3455B,18L3456B,18L3457B,18L3458B,18L3459B, 18L3460B,18L3461B,18L3462B,18L3463B,18L3464B,18L3465B,18L3466B, 18L3467B,18L3468B,18L3469B,18L3470B,18L3471B,18L3472B,18L3473B, 18L3474B,18L3475B,18L3476B,18L3477B,18L3478B,18L3479B,18L3480B, 18L3481B,18L3482B,18L3483B,18L3484B,18L3485B,18L3486B,18L3487B, 18L3488B,18L3489B,18L3490B,18L3491B,18L3492B,18L3493B,18L3494B, 18L3495B,18L3496B,18L3497B,18L3498B,18L3499B,18L3500B,18L3501B, 18L3502B,18L3503B,18L3504B,18L3505B,18L3506B,18L3507B,18L3508B, 18L3509B,18L3510B,18L3511B,18L3512B,18L3513B,18L3514B,18L3515B, 18L3516B,18L3517B,18L3518B,18L3519B,18L3520B,18L3521B,18L3522B, 18L3523B,18L3524B,18L3525B,18L3526B,18L3527B,18L3528B,18L3529B, 18L3530B,18L3531B,18L3532B,18L3533B,18L3534B,18L3535B,18L3536B, 18L3537B,18L3538B,18L3539B,18L3540B,18L3541B,18L3542B,18L3543B, 18L3544B,18L3545B,18L3546B,18L3547B,18L3548B,18L3549B,18L3551B, 18L3552B,18L3553B,18L3554B,18L3555B,18L3556B,18L3557B,18L3558B, 18L3559B,18L3560B,18L3561B,18L3562B,18L3563B,18L3564B,18L3565B, 18L3566B,18L3567B,18L3568B,18L3569B,18L3570B,18L3571B,18L3572B, 18L3573B,18L3574B,18L3575B,18L3576B,18L3577B,18L3578B,18L3579B, 18L3580B,18L3581B,18L3582B,18L3583B,18L3584B,18L3585B,18L3586B, 18L3587B,18L3588B,18L3589B,18L3590B,18L3591B,18L3592B,18L3593B, 18L3594B,18L3595B,18L3596B,18L3597B,18L3598B,18L3599B,18L3600B, 18L3601B,18L3602B,18L3603B,18L3604B,18L3605B,18L3606B,18L3607B, 18L3608B,18L3609B,18L3610B,18L3611B,18L3612B,18L3613B,18L3614B, 18L3615B,18L3616B,18L3617B,18L3618B,18L3619B,18L3620B,18L3621B, 18L3622B,18L3623B,18L3624B,18L3625B,18L3626B,18L3627B,18L3628B, 18L3629B,18L3630B,18L3631B,18L3632B,18L3633B,18L3634B,18L3635B, 18L3636B,18L3637B,18L3638B,18L3639B,18L3640B,18L3641B,18L3642B, 18L3643B,18L3644B,18L3645B,18L3646B,18L3647B,18L3648B,18L3649B, 18L3650B,18L3651B,18L3652B,18L3653B,18L3654B,18L3655B,18L3656B, 18L3657B,18L3658B,18L3659B,18L3660B,18L3661B,18L3662B,18L3663B, 18L3664B,18L3665B,18L3666B,18L3667B,18L3668B,18L3669B,18L3670B, 18L3671B,18L3672B,18L3673B,18L3674B,18L3675B,18L3676B,18L3677B, 18L3678B,18L3679B,18L3680B,18L3681B,18L3682B,18L3683B,18L3684B, 18L3685B,18L3686B,18L3687B,18L3688B,18L3689B,18L3690B,18L3691B, 18L3692B,18L3693B,18L3694B,18L3695B,18L3696B,18L3697B,18L3698B, 18L3699B,18L3700B,18L3701B,18L3702B,18L3703B,18L3704B,18L3705B, 18L3706B,18L3707B,18L3708B,18L3709B,18L3710B,18L3711B,18L3712B, 18L3713B,18L3714B,18L3715B,18L3716B,18L3717B,18L3718B,18L3719B, 18L3720B,18L3721B,18L3722B,18L3723B,18L3724B,18L3725B,18L3726B, 18L3727B,18L3728B,18L3729B,18L3730B,18L3731B,18L3732B,18L3733B, 18L3734B,18L3735B,18L3736B,18L3737B,18L3738B,18L3739B,18L3740B, 18L3741B,18L3742B,18L3743B,18L3744B,18L3745B,18L3746B,18L3747B, 18L3748B,18L3749B,18L3750B,18L3751B,18L3752B,18L3753B,18L3754B, 18L3755B,18L3756B,18L3757B,18L3758B,18L3759B,19A3760B,19A3761B, 19A3762B,19A3763B,19A3764B,19A3765B,19A3766B,19A3767B,19A3768B, 19A3769B,19A3770B,19A3771B,19A3772B,19A3773B,19A3774B,19A3775B, 19A3776B,19A3777B,19A3778B,19A3779B,19A3780B,19A3781B,19A3782B, 19A3783B,19A3784B,19A3785B,19A3786B,19A3787B,19A3788B,19A3789B, 19A3790B,19A3791B,19A3792B,19A3793B,19A3794B,19A3795B,19A3796B, 19A3797B,19A3798B,19A3799B,19A3800B,19A3801B,19A3802B,19A3803B, 19A3804B,19A3805B,19A3806B,19A3807B,19A3808B,19A3809B,19A3810B, 19A3811B,19A3812B,19A3813B,19A3814B,19A3815B,19A3816B,19A3817B, 19A3818B,19A3819B,19A3820B,19A3821B,19A3822B,19A3823B,19A3824B, 19A3825B,19A3826B,19A3827B,19A3828B,19A3829B,19A3830B,19A3831B, 19A3832B,19A3833B,19A3834B,19A3835B,19A3836B,19A3837B,19A3838B, 19A3839B,19A3840B,19A3841B,19A3842B,19A3843B,19A3844B,19A3845B, 19A3846B,19A3847B,19A3848B,19A3849B,19A3850B,19A3851B,19A3852B, 19A3853B,19A3854B,19A3855B,19A3856B,19A3857B,19A3858B,19A3859B, 19A3860B,19A3861B,19A3862B,19A3863B,19A3864B,19A3865B,19A3866B, 19A3867B,19A3868B,19A3869B,19A3870B,19A3871B,19A3872B,19A3873B, 19A3874B,19A3875B,19A3876B,19A3877B,19A3878B,19A3879B,19B3880B, 19B3881B,19B3882B,19B3883B,19B3884B,19B3885B,19B3886B,19B3887B, 19B3888B,19B3889B,19B3890B,19B3891B,19B3892B,19B3893B,19B3894B, 19B3895B,19B3896B,19B3897B,19B3898B,19B3899B,19B3900B,19B3901B, 19B3902B,19B3903B,19B3904B,19B3905B,19B3906B,19B3907B,19B3908B, 19B3909B,19B3910B,19B3911B,19B3912B,19B3913B,19B3914B,19B3915B, 19B3916B,19B3917B,19B3918B,19B3919B,19B3920B,19B3921B,19B3922B, 19B3923B,19B3924B,19B3925B,19B3926B,19B3927B,19B3928B,19B3929B, 19B3930B,19B3931B,19B3932B,19B3933B,19B3934B,19B3935B,19B3936B, 19B3937B,19B3938B,19B3939B,19B3940B,19B3941B,19B3942B,19B3943B, 19B3944B,19B3945B,19B3946B,19B3947B,19B3948B,19B3949B,19B3950B, 19B3951B,19B3952B,19B3953B,19B3954B,19B3955B,19B3956B,19B3957B, 19B3958B,19B3959B,19B3960B,19B3961B,19B3962B,19B3963B,19B3964B, 19B3965B,19B3966B,19B3967B,19B3968B,19B3969B,19B3970B,19B3971B, 19B3972B,19B3973B,19B3974B,19B3975B,19B3976B,19B3977B,19B3978B, 19B3979B,19C3980B,19C3981B,19C3982B,19C3983B,19C3984B,19C3985B, 19C3986B,19C3987B,19C3988B,19C3989B,19C3990B,19C3991B,19C3992B, 19C3993B,19C3994B,19C3995B,19C3996B,19C3997B,19C3998B,19C3999B, 19C4000B,19C4001B,19C4002B,19C4003B,19C4004B,19C4005B,19C4006B, 19C4007B,19C4008B,19C4009B,19C4010B,19C4011B,19C4012B,19C4013B, 19C4014B,19C4015B,19C4016B,19C4017B,19C4018B,19C4019B,19E4020B, 19E4021B,19E4022B,19E4023B,19E4024B,19E4025B,19E4026B,19E4027B, 19E4028B,19E4029B,19E4030B,19E4031B,19E4032B,19E4033B,19E4034B, 19E4035B,19E4036B,19E4037B,19E4038B,19E4039B,19E4040B,19E4041B, 19E4042B,19E4043B,19E4044B,19E4045B,19E4046B,19E4047B,19E4048B, 19E4049B,19E4050B,19E4051B,19E4052B,19E4053B,19E4054B,19E4055B, 19E4056B,19E4057B,19E4058B,19E4059B,19E4060B,19E4061B,19E4062B, 19E4063B,19E4064B,19E4065B,19E4066B,19E4067B,19E4068B,19E4069B, 19E4070B,19E4071B,19E4072B 数量 :892台 出荷時期 :平成31年4月9日〜令和2年12月17日 (2) 品番:JM-SP120 製造ロット:18091367,18091368,18091369,18091370,18091371,18091373,18091374, 18091375,18091376,18091377,18091378,18091379,18091380,18091381, 18091382,18091383,18091384,18091385,18091386,18091387,18091388, 18091389,18091390,18091391,18091392,18091393,18091394,18091395, 18091396,18091397,18091398,18091399,18091400,18091401,18091402, 18091403,18091404,18091405,18091406,18091407,18091408,18091409, 18091410,18091411,18091412,18091413,18091414,18091415,18091416, 18091417,18091418,18091419,18091420,18091421,18091422,18091423, 18091424,18091425,18091426,18091427,18091428,18091429,18091430, 18091431,18091432,18091433,18091434,18091435,18091436,18091437, 18091438,18091439,18091440,18091441,18091442,18091443,18091444, 18091445,18091446,18091447,18091448,18091449,18091450,19091451, 19091452,19091453,19091454,19091455,19091456,19091457,19091458, 19091459,19091460,19091461,19091462,19091463,19091464,19091465, 19091466,19091467,19091468,19091469,19091470,19091471,19091472, 19091473,19091474,19091475,19091476,19091477,19091478,19091479, 19091480,19091481,19091482,19091483,19091484,19091485,19091486, 19091491,19091492,19091493,19091494,19091495,19091496,19091497, 19091498,19091499,19091500,19121501,19121502,19121503,19121504, 19121505,19121506,19121507,19121508,19121509,19121510,19121511, 19121512,19121513,19121514,19121515,19121516,19121517,19121518, 19121519,19121520,19121521,19121522,19121523,19121524,19121525, 19121526,19121527,19121528,19121529,19121530,19121531,19121542, 19121543,19121544,19121545,19121546,19121547,19121548,19121549, 19121550,20021552,20021553,20021554,20021555,20021556,20021557, 20021558,20021559,20021560,20021561,20021562,20021563,20021564, 20021565,20021566,20021567,20021568,20021569,20021570,20021571, 20021572,20021573,20021574,20021575,20021576,20021577,20021578, 20021579,20021580,20021581,20021582,20021583,20021584,20021585, 20021586,20021587,20021588,20021589,20021590,20021591,20021592, 20021593,20021594,20021595,20021596,20021597,20021598,20021599, 20021600,20041601,20041602,20041603,20041604,20041605,20041606, 20041607,20041608,20041609,20041610,20041611,20041612,20041613, 20041614,20041615,20041616,20041617,20041618,20041619,20041620, 20041621,20041622,20041623,20041624,20041625,20041626,20041627, 20041628,20041629,20041630,20041631,20041632,20041633,20041634, 20041635,20041636,20041637,20041638,20041639,20041640,20041641, 20041642,20041643,20041644,20041645,20041646,20041647,20041648, 20041649,20041650,20041651,20041652,20041653,20041654,20041655, 20041656,20041657,20041658,20041659,20041660,20041661,20041662, 20041663,20041664,20041665,20041666,20041667,20041668,20041669, 20041670,20041671,20041672,20041673,20041674,20041675,20041676, 20041677,20041678,20041679,20041680,20041681,20041682,20041683, 20041684,20041685,20041686,20041687,20041688,20041689,20041690, 20041691,20041692,20041693,20041694,20041695,20041696,20041697, 20041698,20041699,20041700,20051701,20051702,20051703,20051704, 20051705,20051706,20051707,20051708,20051709,20051710,20051711, 20051712,20051713,20051714,20051715,20051716,20051717,20051718, 20051719,20051720,20051721,20051722,20051723,20051724,20051725, 20051726,20051727,20051728,20051729,20051730,20051731,20051732, 20051733,20051734,20051735,20051736,20051737,20051738,20051739, 20051740,20051741,20051742,20051743,20051744,20051745,20051746, 20051747,20051748,20051749,20051750,20051751,20051752,20051753, 20051754,20051755,20051756,20051757 数量 :375台 出荷時期 :平成31年2月22日〜令和2年12月18日 (B)対象バッテリに交換した装置(弊社に保守・修理を依頼された装置) (1) 品番:JM-SP520 製造ロット:14Z0357B,1510465B,1510466B,1510467B,1510468B,1510490B,1520569B, 1590690B,1590705B,1590760B,1590761B,1590776B,1590783B,1590789B, 16B0927B,16B0928B,16B1006B,16B1042B,16B1058B,16B1059B,16B1067B, 16B1074B,16B1080B,16B1082B,16C1172B,16C1181B,16C1217B,16C1220B, 16F1237B,16F1255B,16F1278B,16F1309B,16F1342B,16F1343B,16F1346B, 16F1361B,16F1362B,16F1370B,16F1391B,16F1401B,16F1407B,16H1606B, 16H1607B,16H1609B,17C1872B,17C1878B,17D1992B,17D1993B,17D2028B, 17D2037B,17D2083B,17E2105B,17F2191B,17I2278B,17J2312B,17J2319B, 17K2392B,18D2652B,18D2653B,18D2654B,18D2655B,18D2656B,18D2657B, 18D2658B,18D2659B,18D2660B,18D2661B,18D2664B,18D2674B,18D2726B, 18E2787B,4090051B,4090052B,4100097B,4100119B,4100218B,4100219B, 4100222B,4100230B,4100231B,4100232B,4110246B,4110268B,4110284B, 4110314B,4110318B,4110334B 数量 :87台 出荷時期 :平成26年9月30日〜平成30年8月21日 (2) 品番:JM-SP520P 製造ロット:1510078P,1510079P,1520149P,16A0169P,16A0174P,16A0183P,16A0201P, 16D0251P,16J0294P,17E0415P,17E0417P, 数量 :11台 出荷時期 :平成27年12月11日〜平成30年3月8日 (3) 品番:JM-SP120 製造ロット:15100036,15110153,15110190,15120331,15120332,15120334,15120354, 15120368,16030406,16040436,16040441,16040450,16040488,16050602, 16050603,16050611,16050636,16050638,16060672,16060678,16060679, 16060692,16060696,16060718,17030806,17030807,17030808,17030809, 17030818,17070964,17121073,18051115,18051116,18051117 数量 :34台 出荷時期 :平成27年10月30日〜平成31年1月23日 (C)対象バッテリに交換した可能性のある装置 (1) 品番:JM-SP520 製造ロット:1510443B,1510444B,1510501B,1510525B,1510526B,1520578B,1520579B, 1520580B,1560581B,1560582B,1560583B,1560584B,1560585B,1560586B, 1560587B,1560588B,1560589B,1560590B,1560607B,1560608B,1590639B, 1590640B,1590641B,1590642B,1590643B,1590656B,1590657B,1590658B, 1590659B,1590660B,1590661B,1590662B,1590663B,1590664B,1590665B, 1590682B,1590683B,1590684B,1590685B,1590686B,1590689B,1590694B, 1590695B,1590696B,1590697B,1590698B,1590699B,1590700B,1590701B, 1590702B,1590703B,1590704B,1590706B,1590716B,1590717B,1590718B, 1590719B,1590720B,1590721B,1590722B,1590723B,1590724B,1590725B, 1590726B,1590731B,1590749B,1590750B,1590752B,1590753B,1590754B, 1590756B,1590757B,1590758B,1590759B,1590763B,1590764B,1590765B, 1590766B,1590769B,1590770B,1590771B,1590772B,1590773B,1590774B, 1590812B,1590813B,1590816B,1590817B,15J0825B,15J0839B,16B0942B, 16B0943B,16B0949B,16B0950B,16B0951B,16B0952B,16B0953B,16B0954B, 16B0955B,16B0956B,16B0957B,16B0958B,16B0970B,16B0993B,16B0994B, 16B0995B,16B0996B,16B0997B,16B0998B,16B0999B,16B1000B,16B1001B, 16B1002B,16B1003B,16B1004B,16B1005B,16B1007B,16B1034B,16B1035B, 16B1036B,16B1037B,16B1038B,16B1039B,16B1040B,16B1050B,16B1061B, 16B1062B,16B1063B,16B1064B,16B1065B,16B1066B,16B1068B,16B1069B, 16B1070B,16B1071B,16B1072B,16B1075B,16B1076B,16B1077B,16B1078B, 16B1079B,16B1081B,16B1083B,16B1084B,16B1085B,16B1086B,16B1087B, 16B1088B,16B1089B,16B1090B,16B1091B,16C1136B,16C1137B,16C1155B, 16C1185B,16C1186B,16C1193B,16C1194B,16C1195B,16C1196B,16C1197B, 16C1198B,16C1199B,16C1200B,16C1201B,16C1202B,16C1219B,16C1221B, 16F1235B,16F1236B,16F1320B,16F1321B,16F1337B,16F1338B,16F1339B, 16F1340B,16F1341B,16F1351B,16F1352B,16F1364B,16F1365B,16F1386B, 16F1387B,16F1388B,16F1390B,16F1393B,16F1395B,16F1396B,16F1397B, 16F1398B,16F1399B,16F1408B,16F1411B,16F1412B,16F1427B,16G1491B, 16G1492B,16G1493B,16G1494B,16G1500B,16G1501B,16G1502B,16G1503B, 16G1504B,16G1508B,16G1509B,16G1512B,16G1513B,16G1514B,16G1515B, 16G1516B,16H1533B,16H1534B,16H1535B,16H1536B,16H1537B,16H1538B, 16H1539B,16H1540B,16H1541B,16H1542B,16H1543B,16H1544B,16H1545B, 16H1546B,16H1547B,16H1548B,16H1549B,16H1550B,16H1551B,16H1552B, 16H1553B,16H1554B,16H1555B,16H1556B,16H1557B,16H1558B,16H1559B, 16H1560B,16H1561B,16H1562B,16H1563B,16H1564B,16H1565B,16H1566B, 16H1567B,16H1568B,16H1569B,16H1570B,16H1571B,16H1572B,16H1573B, 16H1574B,16H1575B,16H1576B,16H1577B,16H1578B,16H1579B,16H1580B, 16H1581B,16H1582B,16H1583B,16H1584B,16H1585B,16H1586B,16H1587B, 16H1588B,16H1589B,16H1590B,16H1591B,16H1592B,16L1641B,16L1642B, 16L1643B,16L1644B,16L1645B,16L1646B,16L1647B,16L1648B,16L1649B, 16L1650B,16L1651B,16L1652B,16L1653B,16L1654B,16L1655B,16L1656B, 16L1657B,16L1658B,16L1659B,16L1660B,16L1661B,16L1662B,17B1695B, 17B1696B,17B1697B,17B1698B,17B1699B,17B1700B,17B1701B,17B1702B, 17B1703B,17B1704B,17B1705B,17B1706B,17B1707B,17B1708B,17B1709B, 17B1710B,17B1711B,17B1712B,17B1713B,17B1714B,17B1715B,17B1716B, 17B1717B,17B1718B,17B1724B,17B1725B,17B1726B,17B1727B,17B1728B, 17B1729B,17B1730B,17B1731B,17B1732B,17B1733B,17B1734B,17B1735B, 17B1736B,17B1737B,17B1738B,17B1739B,17B1740B,17B1741B,17B1742B, 17B1743B,17B1744B,17B1745B,17B1746B,17B1747B,17B1748B,17B1749B, 17B1750B,17B1751B,17B1752B,17B1753B,17B1754B,17B1755B,17B1756B, 17B1757B,17C1838B,17C1839B,17C1840B,17C1841B,17C1900B,17C1901B, 17C1902B,17C1903B,17C1904B,17C1926B,17C1927B,17C1928B,17C1929B, 17C1930B,17C1931B,17C1933B,17C1934B,17C1935B,17C1936B,17C1937B, 17C1968B,17C1969B,17C1970B,17C1971B,17C1972B,17C1973B,17D1994B, 17D1995B,17D2005B,17D2006B,17D2007B,17D2008B,17D2009B,17D2010B, 17D2011B,17D2012B,17D2039B,17D2040B,17D2041B,17D2042B,17D2043B, 17D2047B,17D2051B,17D2086B,17E2090B,17E2091B,17E2092B,17E2093B, 17E2094B,17E2119B,17F2141B,17F2167B,17F2168B,17F2202B,17F2203B, 17F2204B,17F2205B,17F2206B,17F2209B,17F2210B,17F2211B,17F2212B, 17F2213B,17F2214B,17F2215B,17F2216B,17F2217B,17F2218B,17F2219B, 17F2220B,17F2221B,17I2260B,17I2261B,17I2264B,17I2265B,17I2266B, 17I2267B,17I2268B,17I2269B,17I2270B,17I2271B,17I2272B,17I2273B, 17I2274B,17I2275B,17I2276B,17I2277B,17I2279B,17I2280B,17I2281B, 17I2282B,17I2283B,17I2291B,17I2292B,17I2293B,17I2294B,17I2295B, 17J2323B,17J2324B,17J2325B,17J2326B,17J2327B,17J2328B,17J2329B, 17J2330B,17J2331B,17J2332B,17J2333B,17J2334B,17J2335B,17J2336B, 17J2337B,17J2346B,17J2347B,17J2348B,17J2349B,17K2412B,17K2413B, 17K2414B,17K2415B,17K2416B,17K2417B,17K2418B,17K2442B,17K2443B, 17K2444B,17K2445B,17K2446B,17K2447B,17K2448B,17K2449B,17K2450B, 17K2451B,17K2452B,17K2453B,17K2454B,17K2455B,17K2456B,17K2457B, 17K2458B,17K2459B,17K2460B,17K2461B,17K2462B,17K2463B,17K2464B, 17K2465B,17K2466B,17K2467B,17K2468B,17K2469B,17K2470B,17K2471B, 17K2472B,17K2473B,17K2474B,17K2475B,17K2476B,17K2477B,17K2478B, 17K2479B,17K2480B,17K2481B,17K2482B,17K2483B,17K2484B,17K2485B, 17K2486B,17K2487B,17K2488B,17K2489B,17K2490B,17K2491B,17K2492B, 17K2493B,17K2494B,17K2495B,17K2496B,17K2497B,17K2498B,17K2499B, 18A2500B,18A2501B,18A2502B,18A2503B,18A2504B,18A2505B,18A2506B, 18A2507B,18A2508B,18A2509B,18A2510B,18A2511B,18A2527B,18A2528B, 18A2529B,18A2530B,18A2531B,18A2532B,18A2533B,18A2534B,18A2535B, 18A2536B,18A2537B,18A2538B,18A2539B,18A2540B,18A2541B,18A2542B, 18A2543B,18A2544B,18A2648B,18A2649B,18D2650B,18E2738B,18E2739B, 18E2740B,18E2741B,18E2742B,18E2743B,18E2744B,18E2745B,18E2746B, 18E2747B,18E2748B,18E2749B,18E2750B,18E2751B,18E2752B,18E2766B, 18E2767B,18E2768B,18E2776B,18E2811B,18E2812B,18E2813B,18E2814B, 18E2815B,18F2877B,18F2878B,18F2879B,18F2880B,18F2881B,18F2882B, 18F2883B,18F2884B,18F2885B,18F2886B,18F2887B,18F2888B,18F2889B, 18F2890B,18F2891B,18F2892B,18F2893B,18F2894B,18F2895B,18F2896B, 18F2897B,18F2898B,18F2899B,18F2900B,18F2901B,18F2902B,18F2903B, 18F2904B,18F2905B,18F2906B,18F2907B,18F2908B,18F2909B,18F2910B, 18F2911B,18F2912B,18F2913B,18F2914B,18F2915B,18F2916B,18F2917B, 18F2918B,18F2919B,18F2920B,18F2921B,18F2922B,18F2923B,18F2924B, 18F2925B,18F2926B,18G2932B,18G2943B,18G2944B,18G2945B,18G2946B, 18G2947B,18G2957B,18G2958B,18G2959B,18G2960B,18G2961B,18G2962B, 18G2963B,18G2964B,18G2965B,18G2966B,18G2967B,18G2968B,18G2971B, 18G2972B,18G2973B,18G2974B,18G2975B,18G2976B,18G2977B,18G2978B, 18G2979B,18G2980B,18G2981B,18G2982B,18G2983B,18G2984B,18G2985B, 18G2986B,18G2987B,18G2988B,18G2989B,18G2990B,18G2991B,18G2992B, 18G2993B,18G2994B,18G2995B,18G2996B,18G2997B,18G3023B,18G3024B, 18G3026B,18G3034B,18G3036B,18G3037B,18G3038B,18G3039B,18G3040B, 18H3085B,18H3090B,18H3159B,18H3160B,18H3161B,18H3162B,18H3163B, 18H3164B,18H3165B,18H3166B,19E4102B,19E4103B,19E4104B,19E4105B, 19E4106B,19E4107B,19E4108B,19E4109B,19E4110B,19E4111B,19E4112B, 19E4113B,19E4114B,19E4115B,19E4116B,19E4117B,19E4118B,19E4119B, 19E4120B,19E4121B,19E4122B,19E4123B,19E4124B,19E4125B,19E4126B, 19E4127B,19E4128B,19E4129B,19E4130B,19E4131B,19E4141B,19E4142B, 19E4152B,19E4153B,4090046B,4090064B,4090065B,4090066B,4090067B, 4090069B,4090070B,4090071B,4090072B,4090073B,4090074B,4090075B, 4090076B,4090077B,4090078B,4090079B,4090081B,4090082B,4090083B, 4100094B,4100095B,4100096B,4100098B,4100099B,4100100B,4100101B, 4100102B,4100103B,4100104B,4100105B,4100106B,4100107B,4100108B, 4100109B,4100110B,4100111B,4100112B,4100113B,4100114B,4100115B, 4100116B,4100117B,4100118B,4100120B,4100121B,4100122B,4100123B, 4100192B,4100193B,4100194B,4100195B,4100196B,4100197B,4100198B, 4100199B,4100200B,4100216B,4100220B,4100221B,4110261B,4110262B, 4110263B,4110281B,4110282B,4110283B,4110285B,4110286B,4110287B, 4110288B,4110289B,4110291B,4110292B,4110343B,4110344B,4110345B, 4110346B,4110347B <追加>(*) 14Z0388B,14Z0389B,14Z0390B,1590762B,1590767B,1590768B,1590778B, 1590779B,1590780B,1590781B,1590782B,1590784B,1590785B,1590786B, 1590787B,1590788B,1590790B,1590791B,1590792B,1590793B,1590794B, 1590795B,1590796B,1590797B,1590798B,1590799B,1590800B,1590801B, 1590803B,1590804B,1590805B,1590806B,1590807B,1590808B,1590809B, 1590810B,1590811B,15J0832B,16F1267B,16F1363B,16F1374B,16F1389B, 16F1392B,16F1394B,16G1510B,16G1511B,17C1955B,17C1956B,17C1957B, 17C1958B,17C1959B,4100229B 数量 :866台(*) 出荷時期 :平成27年1月14日〜令和3年1月6日 (2) 品番:JM-SP520P 製造ロット:14Z0006P,14Z0007P,14Z0008P,14Z0043P,14Z0047P,1510056P,1510058P, 1510059P,1510062P,1510065P,1510068P,1510069P,1510070P,1510071P, 1510072P,1510073P,1510074P,1510075P,1510076P,1510077P,1520134P, 1520135P,1520146P,16A0166P,16A0167P,16A0168P,16A0170P,16A0171P, 16A0172P,16A0173P,16A0175P,16A0176P,16A0177P,16A0178P,16A0179P, 16A0180P,16A0181P,16A0182P,16A0184P,16A0185P,16A0186P,16A0187P, 16A0188P,16A0189P,16A0190P,16A0191P,16A0192P,16A0193P,16A0194P, 16A0195P,16A0196P,16A0197P,16A0198P,16A0199P,16A0200P,16A0203P, 16A0204P,16A0205P,16A0206P,16A0207P,16A0208P,16A0209P,16A0210P, 16A0211P,16A0212P,16A0213P,16A0214P,16D0215P,16D0216P,16D0217P, 16D0218P,16D0219P,16D0220P,16D0221P,16D0222P,16D0223P,16D0224P, 16D0225P,16D0226P,16D0227P,16D0228P,16D0229P,16D0230P,16D0231P, 16D0232P,16D0233P,16D0234P,16D0235P,16D0236P,16D0237P,16D0238P, 16D0239P,16D0240P,16D0241P,16D0242P,16D0243P,16D0244P,16D0245P, 16D0246P,16D0247P,16D0248P,16D0249P,16D0250P,16D0252P,16D0253P, 16D0254P,16J0255P,16J0256P,16J0257P,16J0258P,16J0259P,16J0260P, 16J0261P,16J0262P,16J0263P,16J0264P,16J0265P,16J0266P,16J0267P, 16J0268P,16J0269P,16J0270P,16J0271P,16J0272P,16J0273P,16J0274P, 16J0275P,16J0276P,16J0277P,16J0278P,16J0279P,16J0280P,16J0281P, 16J0282P,16J0283P,16J0284P,16J0285P,16J0286P,16J0287P,16J0288P, 16J0289P,16J0290P,16J0291P,16J0292P,16J0293P,16J0295P,16J0296P, 16J0297P,16J0298P,16J0299P,16J0300P,16J0301P,16J0302P,16J0303P, 16J0304P,16J0305P,16J0306P,16J0307P,16J0308P,16J0309P,16J0310P, 16J0311P,16J0312P,16J0313P,16J0314P,16J0315P,16J0316P,16J0317P, 16J0318P,16J0319P,16J0320P,16J0321P,16J0322P,16J0323P,16J0324P, 16J0325P,16J0326P,16J0327P,16J0328P,16J0329P,16J0330P,16J0331P, 16J0332P,16J0333P,16J0334P,16J0335P,16J0336P,16J0337P,16J0338P, 16J0339P,16J0340P,16J0341P,16J0342P,16J0343P,16J0344P,16J0345P, 16J0346P,16J0347P,16J0348P,16J0349P,16J0350P,16J0351P,16J0352P, 16K0369P,16K0370P,16K0371P,16K0387P,16K0388P,16K0389P,16K0390P, 16K0391P,16K0392P,16K0393P,16K0394P,16K0395P,16K0396P,16K0397P, 16K0398P,16K0399P,16K0400P,16K0401P,16L0402P,17G0455P,17G0456P, 17K0489P,17K0490P,17K0491P,17K0492P,17K0493P,17K0494P,17K0495P, 17K0496P,17K0497P,17K0498P,17K0499P,17K0500P,17K0501P,17K0502P, 17K0503P,17K0504P,17K0505P,17K0506P,17K0507P,17K0508P 数量 :244台 出荷時期 :平成26年12月25日〜平成31年4月17日 (3) 品番:JM-SP120 製造ロット:15100111,15100112,15110166,15110167,15110168,15110169,15110170, 15110171,15110172,15110173,15110174,15110175,15110176,15110177, 15110178,15110179,15110180,15110181,15110182,15110183,15110184, 15110185,15110186,15110187,15110188,15110189,15110191,15110192, 15110193,15120288,15120289,15120290,15120291,15120292,15120293, 15120308,15120310,15120311,15120312,15120313,15120318,15120321, 15120322,15120323,15120324,15120325,15120326,15120356,15120357, 15120358,15120359,15120360,15120361,15120362,15120363,15120364, 15120365,15120366,15120367,15120369,15120370,15120371,15120372, 15120373,15120374,15120375,15120376,15120377,15120378,15120379, 15120380,16030381,16030382,16030383,16030384,16030385,16030386, 16030395,16030409,16030427,16040451,16040452,16040468,16040469, 16040470,16040471,16040472,16040473,16040474,16040475,16040476, 16040477,16040478,16040479,16040480,16040481,16040482,16040483, 16040484,16040485,16040486,16040569,16040570,16040571,16040572, 16040573,16040574,16040575,16040576,16040577,16040578,16040579, 16040580,16040581,16040582,16040583,16040584,16040585,16040586, 16040587,16040588,16040589,16040590,16050591,16050592,16050593, 16050594,16050604,16050620,16050621,16050622,16050623,16050624, 16050625,16050626,16050627,16050628,16050629,16050630,16050631, 16050632,16050633,16050637,16050639,16050640,16050641,16050642, 16050643,16050644,16050645,16050646,16050647,16050648,16050649, 16050650,16050651,16050652,16050653,16050654,16050655,16050656, 16050657,16050658,16050659,16050660,16060661,16060664,16060665, 16060666,16060670,16060671,16060673,16060674,16060675,16060676, 16060677,16060680,16060681,16060682,16060683,16060684,16060714, 16060715,16060719,16060720,16060721,16060730,16060732,16060743, 17030784,17030787,17030814,17060909,17060910,17060911,17060912, 17060913,17060914,17070940,17070941,17070942,17070945,17070946, 17070947,17070948,17070949,17070950,17070951,17070952,17070953, 17070954,17070955,17070956,17070957,17070958,17070959,17070962, 17070963,17111005,17111006,17111007,17111008,17111009,17111010, 17111011,17111012,17111013,17111014,17111015,17111016,17111026, 17111030,17121035,18051096,18061158,18061159,18061160,18061161, 18061162,18061163,18061164,18061165,18061166,18061167,18061168, 18061169,18061170,18061171,18061172,18071188,18071201,18071202, 18071203,18071205,18071206,18071218,18071220,18071221,18071222, 18071223,18071227,18071239,18071240,18071248,18091320,18091358, 20061801,20061803 <追加>(*) 16030408,16030410,16030411,16030412,16030413,16030414,16030415, 20061808 数量 :276台(*) 出荷時期 :平成27年11月26日〜令和3年1月21日(*) なお、保守用バッテリの情報は以下のとおりです。 品名 :バッテリ SP120/520 製造番号 :(1)180815 (2)181022 (3)190726 (4)190924 (5)200113 出荷数量 :(1)216個 (2)473個 (3)303個 (4)220個 (5)220個 計:1,432個
製造販売業者等名称
製造販売業者の名称 : 株式会社ジェイ・エム・エス 製造販売業者の所在地: 広島県広島市中区加古町12番17号 許可の種類 : 第一種医療機器製造販売業 許可番号 : 34B1X00001
回収理由
当該製品において、医療機関より「バッテリ点検」を実施した際に、「バッテリが劣化しています バッテリ を交換してください」と表示されるとの報告を受けました。 調査の結果、バッテリ変更により、一定の使用条件下において、容量を誤認識する場合があることが判明した 為、容量の誤認識が生じないバッテリに交換する自主回収を行います。
危惧される具体的な健康被害
バッテリ容量を誤認識した場合、当該装置の連続駆動時間が約半分(4時間程度)に低下しますが、バッテリ 残量警報及びバッテリ空警報は正常に機能することから、重篤な健康被害が発生する可能性はないものと考えま す。なお、現在までに当該事象に起因する健康被害の報告はありません。
回収開始年月日
令和3年3月26日
効能・効果又は用途等
医薬品及び溶液等をポンプによって発生した陽圧により患者に注入することを目的とし、あらかじめ設定され た投与速度又は投与量に従って連続(持続)注入、非連続(間欠)注入又はボーラスを制御するポンプである。
その他
出荷先は全て把握しており、該当する代理店、修理業者、医療機関に情報提供を行い、自主回収を行います。 なお、保守用の対象バッテリについては、修理業者、医療機関での保守状況を確認したうえで、対象装置の自 主回収を行います。