一般的名称及び販売名
一般的名称: 植込み能動型機器用プログラマ 販売名 : (1)ケアリンク SmartSync Device Manager (2)ケアリンク 2090W
対象ロット、数量及び出荷時期
(1)ケアリンク SmartSync Device Manager 対象製品のモバイルアプリケーション(以下、ソフトウェアともいう)がダウンロード可能なタブレット端末の 製造番号: DLXY8165LMW0, DLXY90PSLMW0, DLXY91RALMW0, DLXY92SMLMW0, DLXYC03PLMW0, DLXYC0C0LMW0, DLXYC0R8LMW0, DLXYC4BKLMW0, DLXYC4D1LMW0, DLXYC57CLMW0, DLXYC5N0LMW0, DLXYC5R3LMW0, DLXYD0ZNLMW0, DLXYF04MLMW0, DLXYF0C6LMW0, DLXYF0LLLMW0, DLXYF171LMW0, DLXYH0EGLMW0, DLXYH283LMW0, DLXYH338LMW0, DMPX703UJ2D2, DMPX7046J2D2, DMPX7052J2D2, DMPX706XJ2D2, DMPX708AJ2D2, DMPX70A6J2D2, DMPX70C4J2D2, DMPX70CQJ2D2, DMPX70D8J2D2, DMPX70G4J2D2, DMPX70K1J2D2, DMPX70LHJ2D2, DMPX70LWJ2D2, DMPX70P5J2D2, DMPX70VNJ2D2, DMPX82VDJ2D2, DMPX82YZJ2D2, DMPX9079J2D2, DMPX907GJ2D2, DMPX907WJ2D2, DMPX908GJ2D2, DMPX9092J2D2, DMPX90ADJ2D2, DMPX90AXJ2D2, DMPX90BNJ2D2, DMPX90CXJ2D2, DMPX90D7J2D2, DMPX90DFJ2D2, DMPX90DMJ2D2, DMPX90EAJ2D2, DMPX90EFJ2D2, DMPX90ETJ2D2, DMPX90F5J2D2, DMPX90FPJ2D2, DMPX90GUJ2D2, DMPX90HYJ2D2, DMPX90J9J2D2, DMPX90JJJ2D2, DMPX90KEJ2D2, DMPX90Q2J2D2, DMPX90QMJ2D2, DMPX90WTJ2D2, DMPX910TJ2D2, DMPX92WNJ2D2, DMPX9338J2D2, DMPX9350J2D2, DMPXC0YLJ2D2, DMPXC17NJ2D2, DMPXC17PJ2D2, DMPXC186J2D2, DMPXC18ZJ2D2, DMPXC19MJ2D2, DMPXC19XJ2D2, DMPXC1BVJ2D2, DMPXC1CLJ2D2, DMPXC1D5J2D2, DMPXC1DCJ2D2, DMPXC1DJJ2D2, DMPXC1DKJ2D2, DMPXC1E8J2D2, DMPXC1F7J2D2, DMPXC1GAJ2D2, DMPXC1GNJ2D2, DMPXC1GWJ2D2, DMPXC1JXJ2D2, DMPXC1K4J2D2, DMPXC1MZJ2D2, DMPXC1NNJ2D2, DMPXC1NPJ2D2, DMPXC1P5J2D2, DMPXC1SBJ2D2, DMPXC1TRJ2D2, DMPXC1U1J2D2, DMPXC1U9J2D2, DMPXD1WQJ2D2, DMPXF03WJ2D2, DMPXF0FYJ2D2, DMPXF0HBJ2D2, DMPXF0RRJ2D2, DMPXF0SNJ2D2, DMPXF0SXJ2D2, DMPXF0TXJ2D2, DMPXF0VGJ2D2, DMPXF16KJ2D2, DMPY403VLMW0, DMPY4077LMW0, DMPY40D1LMW0, DMPY40F1LMW0, DMPY40F5LMW0, DMPY40JBLMW0, DMPY40JVLMW0, DMPY40LALMW0, DMPY40MVLMW0, DMPY40NJLMW0, DMPY40PNLMW0, DMPY40SDLMW0, DMPY40SFLMW0, DMPY40TQLMW0, DMPY40ULLMW0, DMPY40UPLMW0, DMPY5014LMW0, DMPY503JLMW0, DMPY504LLMW0, DMPY504RLMW0, DMPY5060LMW0, DMPY5062LMW0, DMPY506ALMW0, DMPY506BLMW0, DMPY506CLMW0, DMPY506JLMW0, DMPY506KLMW0, DMPY506SLMW0, DMPY507VLMW0, DMPY508LLMW0, DMPY508NLMW0, DMPY509LLMW0, DMPY50AGLMW0, DMPY50APLMW0, DMPY50B2LMW0, DMPY50BHLMW0, DMPY50CRLMW0, DMPY50CYLMW0, DMPY50D3LMW0, DMPY50EULMW0, DMPY50FXLMW0, DMPY50GBLMW0, DMPY50GKLMW0, DMPY50H1LMW0, DMPY50JSLMW0, DMPY50JULMW0, DMPY50JZLMW0, DMPY50K3LMW0, DMPY50LHLMW0, DMPY50LTLMW0, DMPY50N2LMW0, DMPY50PBLMW0, DMPY50QVLMW0, DMPY50UHLMW0, DMPY50VWLMW0, DMPY511FLMW0, DMPY517GLMW0, DMPY517TLMW0, DMPY518FLMW0, DMPY5192LMW0, DMPY5199LMW0, DMPY519FLMW0, DMPY51AQLMW0, DMPY51AWLMW0, DMPY51AYLMW0, DMPY51B1LMW0, DMPY51BVLMW0, DMPY600CLMW0, DMPY602RLMW0, DMPY603DLMW0, DMPY6047LMW0, DMPY6066LMW0, DMPY609JLMW0, DMPY60ABLMW0, DMPY60DWLMW0, DMPY60FELMW0, DMPY60HCLMW0, DMPY60HXLMW0, DMPY60J6LMW0, DMPY60JFLMW0, DMPY60JTLMW0, DMPY60M4LMW0, DMPY60PQLMW0, DMPY60PRLMW0, DMPY60QKLMW0, DMPY701PLMW0, DMPY7033LMW0, DMPY7056LMW0, DMPY709LLMW0, DMPY70ACLMW0, DMPY70LWLMW0, DMPY70NLLMW0, DMPY70VQLMW0, DMPY70VSLMW0, DMPY70YWLMW0, DMPY70Z8LMW0, DMPY7103LMW0, DMPY710ZLMW0, DMPY7132LMW0, DMPY713CLMW0, DMPY7156LMW0, DMPY715QLMW0, DMPY7164LMW0, DMPY716ELMW0, DMPY717WLMW0, DMPY7187LMW0, DMPY718PLMW0, DMPY71BDLMW0, DMPY71BFLMW0, DMPY71BPLMW0, DMPY71BULMW0, DMPY71DALMW0, DMPY71FCLMW0, DMPY71FELMW0, DMPY71FRLMW0, DMPY71FWLMW0, DMPY71GWLMW0, DMPY71H1LMW0, DMPY71H9LMW0, DMPY71HTLMW0, DMPY71HYLMW0, DMPY71J5LMW0, DMPY71JNLMW0, DMPY71JPLMW0, DMPY71KELMW0, DMPY71KTLMW0, DMPY71LCLMW0, DMPY71LMLMW0, DMPY71M0LMW0, DMPY71M2LMW0, DMPY71MBLMW0, DMPY71MHLMW0, DMPY71N7LMW0, DMPY71NPLMW0, DMPY71P3LMW0, DMPY71P4LMW0, DMPY71PQLMW0, DMPY71PZLMW0, DMPY71Q4LMW0, DMPY71QRLMW0, DMPY71QTLMW0, DMPY71QZLMW0, DMPY71R1LMW0, DMPY71RCLMW0, DMPY71RHLMW0, DMPY71RLLMW0, DMPY71SBLMW0, DMPY71SFLMW0, DMPY71SRLMW0, DMPY71SWLMW0, DMPY71TCLMW0, DMPY71TDLMW0, DMPY71U9LMW0, DMPY71UPLMW0, DMPY71V5LMW0, DMPY71VPLMW0, DMPY71VTLMW0, DMPY71VWLMW0, DMPY71X3LMW0, DMPY71XELMW0, DMPY71XFLMW0, DMPY71XGLMW0, DMPY71XPLMW0, DMPY71Y6LMW0, DMPY71YDLMW0, DMPY71Z0LMW0, DMPY71Z2LMW0, DMPY71Z9LMW0, DMPY71ZDLMW0, DMPY71ZXLMW0, DMPY720HLMW0, DMPY720LLMW0, DMPY720QLMW0, DMPY720SLMW0, DMPY7217LMW0, DMPY721DLMW0, DMPY721QLMW0, DMPY721SLMW0, DMPY721ZLMW0, DMPY7220LMW0, DMPY7222LMW0, DMPY7226LMW0, DMPY722RLMW0, DMPY722ULMW0, DMPY722WLMW0, DMPY723CLMW0, DMPY723DLMW0, DMPY723LLMW0, DMPY7249LMW0, DMPY724CLMW0, DMPY724JLMW0, DMPY724MLMW0, DMPY724XLMW0, DMPY7259LMW0, DMPY725PLMW0, DMPY7262LMW0, DMPY7265LMW0, DMPY7269LMW0, DMPY726CLMW0, DMPY726JLMW0, DMPY727YLMW0, DMPY7288LMW0, DMPY728JLMW0, DMPY728LLMW0, DMPY728SLMW0, DMPY729BLMW0, DMPY72A0LMW0, DMPY72ALLMW0, DMPY72APLMW0, DMPY72BDLMW0, DMPY72C0LMW0, DMPY72C8LMW0, DMPY72CRLMW0, DMPY72DALMW0, DMPY72DFLMW0, DMPY72DLLMW0, DMPY72DNLMW0, DMPY72E5LMW0, DMPY72EFLMW0, DMPY72EGLMW0, DMPY72F8LMW0, DMPY72F9LMW0, DMPY72FELMW0, DMPY72FSLMW0, DMPY72FXLMW0, DMPY72GKLMW0, DMPY72GQLMW0, DMPY72HKLMW0, DMPY72HLLMW0, DMPY72JJLMW0, DMPY72JMLMW0, DMPY72JNLMW0, DMPY72KULMW0, DMPY72KXLMW0, DMPY72LTLMW0, DMPY72M0LMW0, DMPY72M4LMW0, DMPY72N3LMW0, DMPY72P5LMW0, DMPY72PMLMW0, DMPY72PYLMW0, DMPY72R8LMW0, DMPY72RFLMW0, DMPY72RGLMW0, DMPY72RLLMW0, DMPY72RQLMW0, DMPY72SELMW0, DMPY72SGLMW0, DMPY72SSLMW0, DMPY72STLMW0, DMPY72T8LMW0, DMPY72UALMW0, DMPY72VLLMW0, DMPY72VYLMW0, DMPY72WCLMW0, DMPY72WDLMW0, DMPY72WKLMW0, DMPY72WXLMW0, DMPY72X7LMW0, DMPY72XCLMW0, DMPY72XMLMW0, DMPY72Y6LMW0, DMPY72YHLMW0, DMPY72ZNLMW0, DMPY72ZPLMW0, DMPY7300LMW0, DMPY8003LMW0, DMPY8008LMW0, DMPY800LLMW0, DMPY8018LMW0, DMPY8019LMW0, DMPY801LLMW0, DMPY801NLMW0, DMPY801ULMW0, DMPY801YLMW0, DMPY802ALMW0, DMPY802VLMW0, DMPY8032LMW0, DMPY803ALMW0, DMPY804CLMW0, DMPY804QLMW0, DMPY804VLMW0, DMPY804YLMW0, DMPY805NLMW0, DMPY806GLMW0, DMPY807PLMW0, DMPY808WLMW0, DMPY809LLMW0, DMPY80A4LMW0, DMPY80AHLMW0, DMPY80AXLMW0, DMPY80AZLMW0, DMPY80B1LMW0, DMPY80CALMW0, DMPY80CFLMW0, DMPY80D7LMW0, DMPY80DBLMW0, DMPY80DYLMW0, DMPY80E0LMW0, DMPY80EGLMW0, DMPY80EKLMW0, DMPY80G5LMW0, DMPY80GWLMW0, DMPY80J0LMW0, DMPY80K8LMW0, DMPY80MNLMW0, DMPY80NALMW0, DMPY80NZLMW0, DMPY80PDLMW0, DMPY80PHLMW0, DMPY80Q9LMW0, DMPY80RYLMW0, DMPY80SDLMW0, DMPY80TULMW0, DMPY80UDLMW0, DMPY80UTLMW0, DMPY80UWLMW0, DMPY80V2LMW0, DMPY80WCLMW0, DMPY80WHLMW0, DMPY80X7LMW0, DMPY80XGLMW0, DMPY80XMLMW0, DMPY80YHLMW0, DMPY80Z4LMW0, DMPY80Z6LMW0, DMPY80ZCLMW0, DMPY80ZRLMW0, DMPY80ZYLMW0, DMPY8101LMW0, DMPY810FLMW0, DMPY810KLMW0, DMPY8115LMW0, DMPY811MLMW0, DMPY8125LMW0, DMPY812GLMW0, DMPY812HLMW0, DMPY812PLMW0, DMPY819DLMW0, DMPY81ALLMW0, DMPY81APLMW0, DMPY81BZLMW0, DMPY81C4LMW0, DMPY81CFLMW0, DMPY81D7LMW0, DMPY81DULMW0, DMPY81DXLMW0, DMPY81FFLMW0, DMPY81FLLMW0, DMPY81GNLMW0, DMPY81GWLMW0, DMPY81GXLMW0, DMPY81JHLMW0, DMPY81JSLMW0, DMPY81L1LMW0, DMPY81LJLMW0, DMPY81LKLMW0, DMPY81LNLMW0, DMPY81MNLMW0, DMPY81PBLMW0, DMPY81RZLMW0, DMPY81SELMW0, DMPY81SULMW0, DMPY81SWLMW0, DMPY81T2LMW0, DMPY81UELMW0, DMPY81VKLMW0, DMPY81Y5LMW0, DMPY81YELMW0, DMPY81YMLMW0, DMPY820NLMW0, DMPY823QLMW0, DMPY8258LMW0, DMPY826KLMW0, DMPY827SLMW0, DMPY82E2LMW0, DMPY82GBLMW0, DMPY82KHLMW0, DMPY82KZLMW0, DMPY82MJLMW0, DMPY82PYLMW0, DMPY82Q3LMW0, DMPY82QRLMW0, DMPY82R4LMW0, DMPY82TKLMW0, DMPY82UGLMW0, DMPY82UWLMW0, DMPY82VALMW0, DMPY82VCLMW0, DMPY82VFLMW0, DMPY82VLLMW0, DMPY82WKLMW0, DMPY82WRLMW0, DMPY82XKLMW0, DMPY82XTLMW0, DMPY82XXLMW0, DMPY82XYLMW0, DMPY82YDLMW0, DMPY82ZCLMW0, DMPY82ZFLMW0, DMPY82ZPLMW0, DMPY8301LMW0, DMPY8303LMW0, DMPY8304LMW0, DMPY830GLMW0, DMPY832ALMW0, DMPY833LLMW0, DMPY8347LMW0, DMPY834MLMW0, DMPY834NLMW0, DMPY835DLMW0, DMPY835HLMW0, DMPY836ALMW0, DMPY836MLMW0, DMPY839ALMW0, DMPY839GLMW0, DMPY839PLMW0, DMPY839QLMW0, DMPY83AELMW0, DMPY83AJLMW0, DMPY83AQLMW0, DMPY83BALMW0, DMPY83BELMW0, DMPY83BQLMW0, DMPY83C1LMW0, DMPY83CQLMW0, DMPY83EULMW0, DMPY83FMLMW0, DMPY83HXLMW0, DMPY83JJLMW0, DMPY83JMLMW0, DMPY83LALMW0, DMPY83LBLMW0, DMPY83LULMW0, DMPY83MPLMW0, DMPY83NFLMW0, DMPY83P8LMW0, DMPY83PJLMW0, DMPY83PKLMW0, DMPY83PLLMW0, DMPY83QHLMW0, DMPY83R4LMW0, DMPY83R8LMW0, DMPY83RULMW0, DMPY83S0LMW0, DMPY83SZLMW0, DMPY83T5LMW0, DMPY83U0LMW0, DMPY83UGLMW0, DMPY83UHLMW0, DMPY83USLMW0, DMPY83V0LMW0, DMPY83VHLMW0, DMPY83W3LMW0, DMPY83WELMW0, DMPY83WRLMW0, DMPY83WVLMW0, DMPY83XXLMW0, DMPY83YALMW0, DMPY83Z7LMW0, DMPY840KLMW0, DMPY840QLMW0, DMPY840VLMW0, DMPY840ZLMW0, DMPY842HLMW0, DMPY842KLMW0, DMPY8445LMW0, DMPY8457LMW0, DMPY845PLMW0, DMPY8467LMW0, DMPY8468LMW0, DMPY846VLMW0, DMPY8475LMW0, DMPY847ALMW0, DMPY847TLMW0, DMPY847YLMW0, DMPY8481LMW0, DMPY8483LMW0, DMPY8485LMW0, DMPY848DLMW0, DMPY849SLMW0, DMPY849VLMW0, DMPY849WLMW0, DMPY849XLMW0, DMPY849YLMW0, DMPY84A7LMW0, DMPY84AALMW0, DMPY84AGLMW0, DMPY84B9LMW0, DMPY84BMLMW0, DMPY84BRLMW0, DMPY84BULMW0, DMPY84BWLMW0, DMPY84C0LMW0, DMPY84CLLMW0, DMPY84CULMW0, DMPY84D5LMW0, DMPY84D8LMW0, DMPY84DKLMW0, DMPY84DXLMW0, DMPY84E2LMW0, DMPY84E5LMW0, DMPY84EFLMW0, DMPY84ETLMW0, DMPY84FALMW0, DMPY84FPLMW0, DMPY84FTLMW0, DMPY84GBLMW0, DMPY84GCLMW0, DMPY84GTLMW0, DMPY84GYLMW0, DMPY9004LMW0, DMPY9010LMW0, DMPY901FLMW0, DMPY901ULMW0, DMPY902DLMW0, DMPY902LLMW0, DMPY902XLMW0, DMPY9030LMW0, DMPY9033LMW0, DMPY9036LMW0, DMPY903DLMW0, DMPY903ELMW0, DMPY9058LMW0, DMPY905DLMW0, DMPY9067LMW0, DMPY906GLMW0, DMPY906VLMW0, DMPY9082LMW0, DMPY908PLMW0, DMPY9093LMW0, DMPY909ALMW0, DMPY909LLMW0, DMPY909VLMW0, DMPY90BJLMW0, DMPY90BTLMW0, DMPY90D7LMW0, DMPY90FFLMW0, DMPY90G6LMW0, DMPY90H4LMW0, DMPY90HKLMW0, DMPY90HTLMW0, DMPY90HWLMW0, DMPY90J2LMW0, DMPY90JELMW0, DMPY90JJLMW0, DMPY90JVLMW0, DMPY90JYLMW0, DMPY90KGLMW0, DMPY90KLLMW0, DMPY90KMLMW0, DMPY90KRLMW0, DMPY90L4LMW0, DMPY90MCLMW0, DMPY90N1LMW0, DMPY90PRLMW0, DMPY90PSLMW0, DMPY90PZLMW0, DMPY90Q6LMW0, DMPY90QRLMW0, DMPY90R7LMW0, DMPY90R9LMW0, DMPY90RULMW0, DMPY90TCLMW0, DMPY90TVLMW0, DMPY90UTLMW0, DMPY90V0LMW0, DMPY90VZLMW0, DMPY90W0LMW0, DMPY90WCLMW0, DMPY90X1LMW0, DMPY90YFLMW0, DMPY90Z6LMW0, DMPY90ZWLMW0, DMPY910XLMW0, DMPY9112LMW0, DMPY911BLMW0, DMPY911FLMW0, DMPY911MLMW0, DMPY9124LMW0, DMPY9130LMW0, DMPY913JLMW0, DMPY913ULMW0, DMPY9143LMW0, DMPY9151LMW0, DMPY9157LMW0, DMPY916NLMW0, DMPY916XLMW0, DMPY916ZLMW0, DMPY918ELMW0, DMPY9191LMW0, DMPY91A0LMW0, DMPY91ACLMW0, DMPY91B4LMW0, DMPY91BGLMW0, DMPY91CYLMW0, DMPY91ECLMW0, DMPY91GXLMW0, DMPY91HRLMW0, DMPY91K6LMW0, DMPY91KHLMW0, DMPY91NRLMW0, DMPY91P9LMW0, DMPY91SPLMW0, DMPY91STLMW0, DMPY91Y7LMW0, DMPY923ZLMW0, DMPY9243LMW0, DMPY925GLMW0, DMPY927HLMW0, DMPY92BTLMW0, DMPY92EJLMW0, DMPY92ESLMW0, DMPY92G1LMW0, DMPY92JSLMW0, DMPY92SBLMW0, DMPY92WCLMW0, DMPY934ULMW0, DMPY934XLMW0, DMPY934YLMW0, DMPY935DLMW0, DMPY935TLMW0, DMPY936HLMW0, DMPY936NLMW0, DMPY9382LMW0, DMPY938GLMW0, DMPY938JLMW0, DMPY938VLMW0, DMPY93A8LMW0, DMPY93A9LMW0, DMPY93ADLMW0, DMPY93BALMW0, DMPY93BJLMW0, DMPY93CHLMW0, DMPY93CJLMW0, DMPY93EZLMW0, DMPY93FPLMW0, DMPY93GDLMW0, DMPY93H0LMW0, DMPY93H2LMW0, DMPY93HVLMW0, DMPY93JRLMW0, DMPY93K2LMW0, DMPY93KVLMW0, DMPY93LXLMW0, DMPY93M5LMW0, DMPY93MULMW0, DMPY93NDLMW0, DMPY93PELMW0, DMPY93QTLMW0, DMPY93R2LMW0, DMPY93UYLMW0, DMPY93XGLMW0, DMPY93ZWLMW0, DMPY941FLMW0, DMPY9423LMW0, DMPY9427LMW0, DMPY942ULMW0, DMPY9433LMW0, DMPY944PLMW0, DMPY946YLMW0, DMPY9486LMW0, DMPY94A1LMW0, DMPYF0ESLMW0, DMPYF12ALMW0, DMPYF23WLMW0, DMPYF2LDLMW0, DMPYF2PLLMW0, DMPYF2SJLMW0, DMPYF2TRLMW0, DMPYF2Z0LMW0, DMPYG007LMW0, DMPYG00HLMW0, DMPYG00PLMW0, DMPYG01ZLMW0, DMPYG028LMW0, DMPYG03ELMW0, DMPYG03JLMW0, DMPYG03NLMW0, DMPYG03TLMW0, DMPYG04TLMW0, DMPYG04XLMW0, DMPYG060LMW0, DMPYG06BLMW0, DMPYG06HLMW0, DMPYG06SLMW0, DMPYG08DLMW0, DMPYG08HLMW0, DMPYG08MLMW0, DMPYG08VLMW0, DMPYG0A0LMW0, DMPYG0A1LMW0, DMPYG0A4LMW0, DMPYG0A6LMW0, DMPYG0AGLMW0, DMPYG0BMLMW0, DMPYG0BQLMW0, DMPYG0BSLMW0, DMPYG0BULMW0, DMPYG0CELMW0, DMPYG0CRLMW0, DMPYG0D2LMW0, DMPYG0D5LMW0, DMPYG0EXLMW0, DMPYG0JDLMW0, DMPYG0JWLMW0, DMPYG0JYLMW0, DMPYG0K5LMW0, DMPYG0KKLMW0, DMPYG0KSLMW0, DMPYG0KULMW0, DMPYG0LPLMW0, DMPYG0M5LMW0, DMPYG0MFLMW0, DMPYG0MULMW0, DMPYG0MWLMW0, DMPYG0N1LMW0, DMPYG0N2LMW0, DMPYG0N4LMW0, DMPYG0N7LMW0, DMPYG0NDLMW0, DMPYG0NLLMW0, DMPYG0P1LMW0, DMPYG0PRLMW0, DMPYG0Q1LMW0, DMPYG0Q7LMW0, DMPYG0QKLMW0, DMPYG0RNLMW0, DMPYG0RQLMW0, DMPYG0RWLMW0, DMPYG0RZLMW0, DMPYG0SLLMW0, DMPYG0STLMW0, DMPYG0SYLMW0, DMPYG0T3LMW0, DMPYG0T7LMW0, DMPYG0TGLMW0, DMPYG0TJLMW0, DMPYG0TYLMW0, DMPYG0UTLMW0, DMPYG0VMLMW0, DMPYG0VNLMW0, DMPYG0WCLMW0, DMPYG0X7LMW0, DMPYG0XALMW0, DMPYG0XDLMW0, DMPYG0XPLMW0, DMPYG0Y6LMW0, DMPYG0YCLMW0, DMPYG0ZQLMW0, DMPYG0ZTLMW0, DMPYG109LMW0, DMPYG119LMW0, DMPYG11RLMW0, DMPYG124LMW0, DMPYG148LMW0, DMPYG15BLMW0, DMPYG197LMW0, DMPYG19BLMW0, DMPYG1KBLMW0, DMPYG1Q9LMW0, DMPYG1R1LMW0, DMPYG1T0LMW0, DMPYG1T4LMW0, DMPYG2FJLMW0, DMPYG2FKLMW0, DMPYG2G3LMW0, DMPYG2J0LMW0, DMPYG2STLMW0, DMPYG342LMW0, DMPYG343LMW0, DMPYG344LMW0, DMPYG34ULMW0, DMPYG34XLMW0, DMPYG34ZLMW0, DMPYG365LMW0, DMPYG367LMW0, DMPYG37KLMW0, DMPYG37YLMW0, DMPYG38BLMW0, DMPYG38DLMW0, DMPYG39BLMW0, DMPYG39PLMW0, DMPYG39QLMW0, DMPYG3A1LMW0, DMPYG3ALLMW0, DMPYG3BULMW0, DMPYG3BXLMW0, DMPYG3CGLMW0, DMPYG3CXLMW0, DMPYG3D5LMW0, DMPYG3DDLMW0, DMPYG3E2LMW0, DMPYG3EHLMW0, DMPYG3EKLMW0, DMPYG3ELLMW0, DMPYG3F9LMW0, DMPYG3G7LMW0, DMPYG3HRLMW0, DMPYG3J7LMW0, DMPYG3JLLMW0, DMPYG3JRLMW0, DMPYG3JWLMW0, DMPYG3K1LMW0, DMPYG3LTLMW0, DMPYG3LVLMW0, DMPYG3NFLMW0, DMPYG3QHLMW0, DMPYG3QQLMW0, DMPYG3RELMW0, DMPYG3RRLMW0, DMPYG3RWLMW0, DMPYG3SLLMW0, DMPYG3SMLMW0, DMPYG3SPLMW0, DMPYG3SULMW0, DMPYG3U1LMW0, DMPYG3U3LMW0, DMPYG3U6LMW0, DMPYG3ULLMW0, DMPYG3URLMW0, DMPYG3V7LMW0, DMPYG3VFLMW0, DMPYG3VGLMW0, DMPYG3VYLMW0, DMPYG3W9LMW0, DMPYG3WWLMW0, DMPYG40PLMW0, DMPYG40ULMW0, DMPYG41MLMW0, DMPYG425LMW0, DMPYG42GLMW0, DMPYG42KLMW0, DMPYG42RLMW0, DMPYG43MLMW0, DMPYG45ULMW0, DMPYG46ELMW0, DMPYG46VLMW0, DMPYG474LMW0, DMPYG47ULMW0, DMPYG48ELMW0, DMPYG4MGLMW0, DMPYG4MNLMW0, DMPYG4QVLMW0, DMPYH03NLMW0, DMPYH04FLMW0, DMPYH05LLMW0, DMPYH05PLMW0, DMPYH06LLMW0, DMPYH07YLMW0, DMPYH09BLMW0, DMPYH0A0LMW0, DMPYH0B0LMW0, DMPYH0B1LMW0, DMPYH0B5LMW0, DMPYH0BCLMW0, DMPYH0C6LMW0, DMPYH0CQLMW0, DMPYH0EGLMW0, DMPYH0F7LMW0, DMPYH0FRLMW0, DMPYH0G1LMW0, DMPYH0G6LMW0, DMPYH0GMLMW0, DMPYH0GNLMW0, DMPYH0GQLMW0, DMPYH0GULMW0, DMPYH0GZLMW0, DMPYH0KELMW0, DMPYH0N0LMW0, DMPYH0YJLMW0, DMPYH0ZELMW0, DMPYH12ALMW0, DMPYH160LMW0, DMPYH1EVLMW0, DMPYH1QLLMW0, DMPYJ00ULMW0, DMPYJ034LMW0, DMPYJ07XLMW0, DMPYJ0CGLMW0, DMPYJ0DFLMW0, DMPYJ0L3LMW0, DMPYJ1CTLMW0, DMPYJ1D8LMW0, DMPYJ1LNLMW0, DMPYJ1MWLMW0, DMPYJ1PGLMW0, DMPYJ1RULMW0, DMPYJ1UDLMW0, DMPYJ1XALMW0, DMPYJ1YMLMW0, DMPYJ1YXLMW0, DMPYJ20ULMW0, DMPYJ23HLMW0, DMPYJ258LMW0, DMPYJ25HLMW0, DMPYJ26MLMW0, DMPYJ27CLMW0, DMPYJ27TLMW0, DMPYJ2ADLMW0, DMPYJ2BALMW0, DMPYJ2CYLMW0, DMPYJ2DTLMW0, DMPYJ2DWLMW0, DMPYJ2EFLMW0, DMPYJ2EULMW0, DMPYJ2EVLMW0, DMPYJ2F3LMW0, DMPYJ2FYLMW0, DMPYJ2K8LMW0, DMPYJ2LHLMW0, DMPYJ2LMLMW0, DMPYJ2M0LMW0, DMPYJ2NDLMW0, DMPYJ2NULMW0, DMPYJ2P4LMW0, DMPYJ2Q5LMW0, DMPYJ2Q9LMW0, DMPYJ2RJLMW0, DMPYJ2STLMW0, DMPYJ2W4LMW0, DMPYJ2ZELMW0, DMPYJ31BLMW0, DMPYJ33DLMW0, DMPYJ35GLMW0, DMPYJ35LLMW0, DMPYJ38FLMW0, DMPYJ39QLMW0, DMPYJ3ABLMW0, DMPYJ3AJLMW0, DMPYJ3ALLMW0, DMPYJ3AYLMW0, DMPYJ3BTLMW0, DMPYJ3C7LMW0, DMPYJ3D7LMW0, DMPYJ3E0LMW0, DMPYJ3E1LMW0, DMPYJ3EXLMW0, DMPYJ3G3LMW0, DMPYJ3H0LMW0, DMPYJ3JFLMW0, DMPYJ3MXLMW0, DMPYJ3NBLMW0, DMPYJ3PGLMW0, DMPYJ3QSLMW0, DMPYJ3TALMW0, DMPYJ3TSLMW0, DMPYJ3UWLMW0, DMPYJ3XDLMW0, DMPYJ3Y6LMW0, DMPYJ3YKLMW0, DMPYJ3YNLMW0, DMPYJ416LMW0, DMPYJ43VLMW0, DMPYJ457LMW0, DMPYJ48DLMW0, DMPYJ4DDLMW0, DMPYJ4EHLMW0, DMPYJ4FRLMW0, DMPYJ4FSLMW0, DMPYJ4GALMW0, DMPYJ4K2LMW0, DMPYJ4K7LMW0, DMPYJ4MBLMW0, DMPYJ4NLLMW0, DMPYJ4S0LMW0, DMPYJ50XLMW0, DMPYJ581LMW0, DMPYJ5CDLMW0, DMPYJ5DBLMW0, DMPYJ5E4LMW0, DMPYJ5MFLMW0, DMPYJ5PULMW0, DMPYJ5QKLMW0, DMPYJ5QXLMW0, DMPYJ5SVLMW0, DMPYJ5U6LMW0, DMPYJ5U8LMW0, DMPYJ5YPLMW0, DMPYJ60CLMW0, DMPYJ683LMW0, DMPYJ6C6LMW0, DMPYJ6CELMW0, DMPYJ6DTLMW0, DMPYJ6KALMW0, DMPYJ6L8LMW0, DMPYJ6N6LMW0, DMPYJ6NYLMW0, DMPYJ6RSLMW0, DMPZN1JKLMW0, DMPZN1Q7LMW0, DMPZN1RJLMW0, DMPZN1SSLMW0, DMPZN1T7LMW0, DMPZN1UJLMW0, DMPZN2RMLMW0, DMPZQ00BLMW0, DMPZQ011LMW0, DMPZQ01PLMW0, DMPZQ05RLMW0, F9FZ900TLMW0, F9FZ905PLMW0, F9FZ9099LMW0, F9FZ90GQLMW0, F9FZ90KCLMW0, F9FZ90MELMW0, F9FZ90SWLMW0, F9FZ9168LMW0, F9FZ919HLMW0, F9FZD033LMW0, F9FZD0ZKLMW0, F9FZD0ZYLMW0, F9FZD10WLMW0, F9FZD125LMW0, F9FZD13RLMW0, F9FZD142LMW0, F9FZD150LMW0, F9FZD15ULMW0, F9FZD17WLMW0, F9FZD18MLMW0, F9FZD1BYLMW0, F9FZD1G0LMW0, F9FZD1H9LMW0, F9FZD1HFLMW0, F9FZD1KGLMW0, F9FZD1Q5LMW0, F9FZD1QXLMW0, F9FZD1R7LMW0, F9FZD1UGLMW0, F9FZD1UYLMW0, F9FZD1VBLMW0, F9FZD1X5LMW0, F9FZD1XCLMW0, F9FZD1XWLMW0, F9FZD1YCLMW0, F9FZD1YELMW0, F9FZD1ZTLMW0, F9FZD1ZWLMW0, F9FZD202LMW0, F9FZD203LMW0, F9FZD22BLMW0, F9FZD22FLMW0, F9FZD22JLMW0, F9FZD24XLMW0, F9FZD258LMW0, F9FZD25BLMW0, F9FZD26JLMW0, F9FZD2A1LMW0, F9FZD2A7LMW0, F9FZD2ASLMW0, F9FZD2ATLMW0, F9FZD2BELMW0, F9FZD2BGLMW0, F9FZD2BULMW0, F9FZD2C0LMW0, F9FZD2C8LMW0, F9FZD2C9LMW0, F9FZD2CALMW0, F9FZD2CHLMW0, F9FZD2CULMW0, F9FZD2CYLMW0, F9FZD2DELMW0, F9FZD2DJLMW0, F9FZD2DTLMW0, F9FZD2ECLMW0, F9FZD2F5LMW0, F9FZD2FZLMW0, F9FZD2HDLMW0, F9FZD2K9LMW0, F9FZD2KBLMW0, F9FZD2N6LMW0, F9FZD2NVLMW0, F9FZD2PLLMW0, F9FZD2PSLMW0, F9FZD2QCLMW0, F9FZD2QRLMW0, F9FZD2QTLMW0, F9FZD2R7LMW0, F9FZD2RFLMW0, F9FZD2RLLMW0, F9FZD2RVLMW0, F9FZD2SALMW0, F9FZD2SKLMW0, F9FZD2SXLMW0, F9FZD2T3LMW0, F9FZD2TBLMW0, F9FZD2TMLMW0, F9FZD2U8LMW0, F9FZD2UALMW0, F9FZD2UBLMW0, F9FZD2UPLMW0, F9FZD2VTLMW0, F9FZD2WCLMW0, F9FZD2WLLMW0, F9FZD2X2LMW0, F9FZD2X4LMW0, F9FZD2Y3LMW0, F9FZD2Y8LMW0, F9FZD2YKLMW0, F9FZD2YMLMW0, F9FZD30CLMW0, F9FZD30DLMW0, F9FZD30ELMW0, F9FZD30FLMW0, F9FZD313LMW0, F9FZD31FLMW0, F9FZD33VLMW0, F9FZD34CLMW0, F9FZD351LMW0, F9FZD35ALMW0, F9FZF01TLMW0, F9FZF01XLMW0, F9FZF02KLMW0, F9FZF032LMW0, F9FZF04ULMW0, F9FZF04WLMW0, F9FZF053LMW0, F9FZF09FLMW0, F9FZF0A5LMW0, F9FZF0B5LMW0, F9FZF0BKLMW0, F9FZF0H2LMW0, F9FZF0H3LMW0, F9FZF0W0LMW0, F9FZF1RNLMW0, F9FZF1SALMW0, F9FZF1UVLMW0, F9FZF1V9LMW0, F9FZF1WFLMW0, F9FZF1Z4LMW0, F9FZF23DLMW0, F9FZF26ELMW0, F9FZF2CTLMW0, F9FZF2FPLMW0, F9FZG0NULMW0, F9FZJ093LMW0, F9FZJ0CJLMW0, F9FZJ0CXLMW0, F9FZJ0Q3LMW0, F9FZJ0T2LMW0, F9FZJ0UCLMW0, FPLDF001LMW0, GCGVCHL0HLJK, GCTV637QHLJK, GCTV646UHLJK, GCTV64ASHLJK, GCTVXCKYHLJK, GCTVXCLHHLJK, GCTVXCQEHLJK, GCTVXDC9HLJK, GCTVXDQYHLJK, GCTVXE2GHLJK, GCTVXEV4HLJK, GCTVXEV9HLJK, GCTVXFY4HLJK, GCTW10DRHLJK, GCTW10DUHLJK, GCTW10X3HLJK, GCTW10X6HLJK, GCTW10Y1HLJK, GCTW118SHLJK, GCTW119ZHLJK, GCTW11AHHLJK, GCTW11DFHLJK, GCTW1274HLJK, GCTW139DHLJK 数量 1,270台 出荷時期 平成30年6月15日から令和3年3月24日 (2)ケアリンク 2090W 製造番号: PKK001057R, PKK001069R, PKK002306R, PKK003166R, PKK003387R, PKK003426R, PKK003898R, PKK004251R, PKK004356R, PKK004800R, PKK005316R, PKK005459R, PKK005503R, PKK005530R, PKK005547R, PKK005948R, PKK006185R, PKK006205R, PKK006217R, PKK006238R, PKK006240R, PKK006244R, PKK006246R, PKK006247R, PKK006248R, PKK006253R, PKK006256R, PKK006262R, PKK006268R, PKK006271R, PKK006273R, PKK006282R, PKK006297R, PKK006299R, PKK006305R, PKK006308R, PKK006310R, PKK006318R, PKK006319R, PKK006327R, PKK006343R, PKK006346R, PKK006352R, PKK006353R, PKK006358R, PKK006360R, PKK006387R, PKK006400R, PKK006617R, PKK006619R, PKK006627R, PKK006632R, PKK006636R, PKK006811R, PKK007067R, PKK007103R, PKK007116R, PKK007121R, PKK007365R, PKK007605R, PKK007621R, PKK007629R, PKK007632R, PKK007641R, PKK007644R, PKK007648R, PKK007652R, PKK007656R, PKK007657R, PKK007658R, PKK007660R, PKK007661R, PKK007662R, PKK007676R, PKK007680R, PKK007687R, PKK007688R, PKK007691R, PKK007692R, PKK007693R, PKK007694R, PKK007695R, PKK007696R, PKK007698R, PKK007699R, PKK007702R, PKK007703R, PKK007704R, PKK007705R, PKK007707R, PKK007711R, PKK007712R, PKK007726R, PKK007728R, PKK007753R, PKK007762R, PKK007767R, PKK007775R, PKK007780R, PKK007795R, PKK007825R, PKK012012R, PKK012095R, PKK012463R, PKK013039R, PKK013096R, PKK013106R, PKK013134R, PKK013135R, PKK013138R, PKK013139R, PKK013140R, PKK013142R, PKK013143R, PKK013147R, PKK013148R, PKK013149R, PKK013150R, PKK013151R, PKK013153R, PKK013158R, PKK013159R, PKK013161R, PKK013163R, PKK013164R, PKK013165R, PKK013166R, PKK013168R, PKK013170R, PKK013171R, PKK013175R, PKK013176R, PKK013177R, PKK013179R, PKK013180R, PKK013181R, PKK013182R, PKK013183R, PKK013184R, PKK013185R, PKK013187R, PKK013189R, PKK013190R, PKK013193R, PKK013194R, PKK013195R, PKK013198R, PKK013238R, PKK013322R, PKK013502R, PKK013559R, PKK013562R, PKK013564R, PKK013572R, PKK013577R, PKK013578R, PKK013592R, PKK013595R, PKK013596R, PKK013600R, PKK013610R, PKK013614R, PKK013616R, PKK013628R, PKK013632R, PKK013635R, PKK013637R, PKK013639R, PKK013641R, PKK013643R, PKK013644R, PKK013652R, PKK013658R, PKK013661R, PKK013662R, PKK013672R, PKK013678R, PKK013781R, PKK013837R, PKK013854R, PKK014063R, PKK014313R, PKK014439R, PKK014457R, PKK014474R, PKK014987R, PKK015488R, PKK015531R, PKK015644R, PKK015654R, PKK015681R, PKK015686R, PKK015719R, PKK015722R, PKK015796R, PKK015822R, PKK015832R, PKK015836R, PKK015846R, PKK015847R, PKK015851R, PKK015854R, PKK015855R, PKK015856R, PKK015857R, PKK015864R, PKK015865R, PKK015866R, PKK015867R, PKK015868R, PKK015870R, PKK015874R, PKK015875R, PKK015876R, PKK015877R, PKK015878R, PKK015879R, PKK015880R, PKK015881R, PKK015882R, PKK015884R, PKK015886R, PKK015887R, PKK015888R, PKK015889R, PKK015892R, PKK015893R, PKK015894R, PKK015895R, PKK015896R, PKK015897R, PKK015898R, PKK015899R, PKK015900R, PKK015901R, PKK015902R, PKK015903R, PKK015905R, PKK015906R, PKK015908R, PKK015909R, PKK015910R, PKK015911R, PKK015912R, PKK015913R, PKK015914R, PKK015915R, PKK015916R, PKK015917R, PKK015918R, PKK015919R, PKK015920R, PKK015921R, PKK015922R, PKK015923R, PKK015924R, PKK015925R, PKK015926R, PKK015927R, PKK015928R, PKK015930R, PKK015931R, PKK015933R, PKK015934R, PKK015935R, PKK015936R, PKK015937R, PKK015938R, PKK015939R, PKK015942R, PKK015945R, PKK015946R, PKK015947R, PKK015948R, PKK015950R, PKK015951R, PKK015952R, PKK015953R, PKK015955R, PKK015957R, PKK015958R, PKK015959R, PKK015962R, PKK015963R, PKK015964R, PKK015966R, PKK015967R, PKK015969R, PKK015971R, PKK015972R, PKK015973R, PKK015975R, PKK015976R, PKK015977R, PKK015978R, PKK015979R, PKK015980R, PKK015981R, PKK015982R, PKK015983R, PKK015984R, PKK015985R, PKK015986R, PKK015987R, PKK015988R, PKK015989R, PKK015990R, PKK015991R, PKK015993R, PKK015994R, PKK015997R, PKK015998R, PKK015999R, PKK016001R, PKK016002R, PKK016003R, PKK016004R, PKK016006R, PKK016007R, PKK016008R, PKK016011R, PKK016012R, PKK016013R, PKK016016R, PKK016018R, PKK016019R, PKK016020R, PKK016021R, PKK016022R, PKK016025R, PKK016026R, PKK016027R, PKK016028R, PKK016029R, PKK016031R, PKK016032R, PKK016033R, PKK016034R, PKK016035R, PKK016036R, PKK016037R, PKK016043R, PKK016046R, PKK016047R, PKK016048R, PKK016050R, PKK016051R, PKK016054R, PKK016056R, PKK016060R, PKK016078R, PKK016081R, PKK016084R, PKK016093R, PKK016097R, PKK016101R, PKK016102R, PKK016103R, PKK016104R, PKK016105R, PKK016115R, PKK016215R, PKK016242R, PKK016281R, PKK016293R, PKK016318R, PKK016325R, PKK016329R, PKK016352R, PKK016366R, PKK016391R, PKK016397R, PKK016434R, PKK016443R, PKK016450R, PKK016455R, PKK016457R, PKK016459R, PKK016474R, PKK016476R, PKK016479R, PKK016485R, PKK016486R, PKK016487R, PKK016488R, PKK016490R, PKK016492R, PKK016494R, PKK016495R, PKK016496R, PKK016497R, PKK016498R, PKK016499R, PKK016500R, PKK016501R, PKK016502R, PKK016503R, PKK016504R, PKK016505R, PKK016506R, PKK016507R, PKK016508R, PKK016509R, PKK016511R, PKK016512R, PKK016514R, PKK016516R, PKK016517R, PKK016518R, PKK016520R, PKK016522R, PKK016523R, PKK016524R, PKK016525R, PKK016526R, PKK016527R, PKK016528R, PKK016529R, PKK016531R, PKK016533R, PKK016534R, PKK016535R, PKK016536R, PKK016537R, PKK016539R, PKK016549R, PKK016550R, PKK016551R, PKK016552R, PKK016555R, PKK016558R, PKK016559R, PKK016560R, PKK016565R, PKK016566R, PKK016568R, PKK016570R, PKK016571R, PKK016572R, PKK016573R, PKK016576R, PKK016577R, PKK016578R, PKK016581R, PKK016584R, PKK016585R, PKK016588R, PKK016589R, PKK016590R, PKK016591R, PKK016592R, PKK016593R, PKK016595R, PKK016596R, PKK016597R, PKK016598R, PKK016601R, PKK016604R, PKK016607R, PKK016608R, PKK016609R, PKK016662R, PKK016674R, PKK016679R, PKK016787R, PKK016849R, PKK016870R, PKK017011R, PKK017157R, PKK017168R, PKK017172R, PKK017190R, PKK017212R, PKK017218R, PKK017223R, PKK017228R, PKK017230R, PKK017231R, PKK017232R, PKK017236R, PKK017239R, PKK017240R, PKK017241R, PKK017250R, PKK017253R, PKK017256R, PKK017257R, PKK017259R, PKK017261R, PKK017262R, PKK017267R, PKK017269R, PKK017272R, PKK017276R, PKK017277R, PKK017278R, PKK017283R, PKK017285R, PKK017286R, PKK017287R, PKK017288R, PKK017289R, PKK017291R, PKK017292R, PKK017293R, PKK017294R, PKK017296R, PKK017297R, PKK017299R, PKK017300R, PKK017302R, PKK017303R, PKK017304R, PKK017306R, PKK017310R, PKK017312R, PKK017332R, PKK017545R, PKK017605R, PKK017633R, PKK017672R, PKK017773R, PKK017903R, PKK017924R, PKK017955R, PKK017981R, PKK018004R, PKK018022R, PKK018062R, PKK018081R, PKK018093R, PKK018096R, PKK018111R, PKK018114R, PKK018117R, PKK018123R, PKK018131R, PKK018137R, PKK018146R, PKK018149R, PKK018159R, PKK018162R, PKK018163R, PKK018165R, PKK018166R, PKK018167R, PKK018169R, PKK018173R, PKK018174R, PKK018175R, PKK018176R, PKK018177R, PKK018178R, PKK018179R, PKK018180R, PKK018181R, PKK018182R, PKK018183R, PKK018184R, PKK018186R, PKK018187R, PKK018188R, PKK018189R, PKK018191R, PKK018192R, PKK018194R, PKK018197R, PKK018199R, PKK018201R, PKK018202R, PKK018204R, PKK018205R, PKK018206R, PKK018207R, PKK018208R, PKK018209R, PKK018213R, PKK018214R, PKK018216R, PKK018217R, PKK018218R, PKK018219R, PKK018220R, PKK018221R, PKK018222R, PKK018223R, PKK018226R, PKK018227R, PKK018228R, PKK018230R, PKK018231R, PKK018233R, PKK018234R, PKK018235R, PKK018236R, PKK018237R, PKK018238R, PKK018239R, PKK018240R, PKK018242R, PKK018244R, PKK018247R, PKK018248R, PKK018249R, PKK018252R, PKK018253R, PKK018257R, PKK018259R, PKK018262R, PKK018263R, PKK018265R, PKK018266R, PKK018268R, PKK018270R, PKK018272R, PKK018273R, PKK018275R, PKK018277R, PKK018278R, PKK018279R, PKK018281R, PKK018282R, PKK018283R, PKK018284R, PKK018285R, PKK018287R, PKK018288R, PKK018289R, PKK018291R, PKK018292R, PKK018293R, PKK018294R, PKK018295R, PKK018296R, PKK018297R, PKK018298R, PKK018299R, PKK018300R, PKK018301R, PKK018302R, PKK018303R, PKK018304R, PKK018305R, PKK018306R, PKK018307R, PKK018309R, PKK018310R, PKK018311R, PKK018312R, PKK018313R, PKK018314R, PKK018316R, PKK018317R, PKK018318R, PKK018320R, PKK018322R, PKK018323R, PKK018324R, PKK018325R, PKK018326R, PKK018327R, PKK018328R, PKK018329R, PKK018330R, PKK018331R, PKK018332R, PKK018333R, PKK018334R, PKK018336R, PKK018337R, PKK018338R, PKK018339R, PKK018340R, PKK018341R, PKK018342R, PKK018343R, PKK018344R, PKK018345R, PKK018346R, PKK018347R, PKK018348R, PKK018350R, PKK018351R, PKK018352R, PKK018353R, PKK018354R, PKK018355R, PKK018356R, PKK018357R, PKK018358R, PKK018360R, PKK018361R, PKK018362R, PKK018363R, PKK018364R, PKK018365R, PKK018366R, PKK018367R, PKK018368R, PKK018369R, PKK018370R, PKK018371R, PKK018372R, PKK018373R, PKK018374R, PKK018375R, PKK018376R, PKK018377R, PKK018378R, PKK018379R, PKK018380R, PKK018381R, PKK018382R, PKK018383R, PKK018385R, PKK018386R, PKK018387R, PKK018388R, PKK018389R, PKK018390R, PKK018391R, PKK018392R, PKK018393R, PKK018394R, PKK018395R, PKK018396R, PKK018397R, PKK018398R, PKK018399R, PKK018400R, PKK018401R, PKK018402R, PKK018403R, PKK018404R, PKK018405R, PKK018406R, PKK018407R, PKK018408R, PKK018409R, PKK018410R, PKK018411R, PKK018412R, PKK018414R, PKK018415R, PKK018416R, PKK018417R, PKK018419R, PKK018420R, PKK018424R, PKK018425R, PKK018426R, PKK018427R, PKK018428R, PKK018429R, PKK018430R, PKK018433R, PKK018434R, PKK018436R, PKK018437R, PKK018438R, PKK018440R, PKK018442R, PKK018444R, PKK018447R, PKK018448R, PKK018449R, PKK018451R, PKK018452R, PKK018457R, PKK018458R, PKK018460R, PKK018461R, PKK018462R, PKK018464R, PKK018470R, PKK018477R, PKK018482R, PKK018483R, PKK018484R, PKK018491R, PKK018494R, PKK018495R, PKK018496R, PKK018498R, PKK018499R, PKK018501R, PKK018507R, PKK018509R, PKK018510R, PKK018511R, PKK018513R, PKK018514R, PKK018516R, PKK018517R, PKK018520R, PKK018521R, PKK018523R, PKK018526R, PKK018527R, PKK018532R, PKK018534R, PKK018535R, PKK018537R, PKK018541R, PKK018542R, PKK018543R, PKK018548R, PKK018549R, PKK018550R, PKK018551R, PKK018553R, PKK018557R, PKK018558R, PKK018559R, PKK018560R, PKK018561R, PKK018562R, PKK018565R, PKK018566R, PKK018567R, PKK018568R, PKK018571R, PKK018573R, PKK018719R, PKK018771R, PKK018796R, PKK018823R, PKK018827R, PKK018893R, PKK019245R, PKK019283R, PKK019302R, PKK019303R, PKK019373R, PKK019409R, PKK019415R, PKK019438R, PKK019466R, PKK019471R, PKK019476R, PKK019483R, PKK019493R, PKK019494R, PKK019510R, PKK019517R, PKK019519R, PKK019523R, PKK019526R, PKK019527R, PKK019530R, PKK019532R, PKK019533R, PKK019536R, PKK019539R, PKK019540R, PKK019541R, PKK019542R, PKK019543R, PKK019544R, PKK019545R, PKK019547R, PKK019549R, PKK019551R, PKK019552R, PKK019553R, PKK019554R, PKK019555R, PKK019556R, PKK019557R, PKK019559R, PKK019560R, PKK019561R, PKK019566R, PKK019567R, PKK019569R, PKK019571R, PKK019572R, PKK019573R, PKK019574R, PKK019575R, PKK019576R, PKK019577R, PKK019578R, PKK019579R, PKK019580R, PKK019584R, PKK019585R, PKK019586R, PKK019587R, PKK019588R, PKK019591R, PKK019592R, PKK019593R, PKK019594R, PKK019595R, PKK019597R, PKK019598R, PKK019600R, PKK019601R, PKK019603R, PKK019605R, PKK019613R, PKK019616R, PKK019617R, PKK019618R, PKK019619R, PKK019620R, PKK019621R, PKK019625R, PKK019632R, PKK019634R, PKK019642R, PKK019653R, PKK019655R, PKK019657R, PKK019670R, PKK019672R, PKK019673R, PKK019674R, PKK019675R, PKK019676R, PKK019677R, PKK019678R, PKK019681R, PKK019682R, PKK019683R, PKK019685R, PKK019686R, PKK019687R, PKK019688R, PKK019689R, PKK019690R, PKK019692R, PKK019693R, PKK019695R, PKK019697R, PKK019698R, PKK019700R, PKK019702R, PKK019703R, PKK019704R, PKK019705R, PKK019706R, PKK019707R, PKK019709R, PKK019711R, PKK019712R, PKK019716R, PKK019717R, PKK019719R, PKK019720R, PKK019721R, PKK019722R, PKK019723R, PKK019724R, PKK019725R, PKK019726R, PKK019727R, PKK019728R, PKK019729R, PKK019730R, PKK019731R, PKK019732R, PKK019733R, PKK019735R, PKK019738R, PKK019740R, PKK019741R, PKK019747R, PKK019749R, PKK019760R, PKK019766R, PKK019768R, PKK019769R, PKK019770R, PKK019771R, PKK019772R, PKK019773R, PKK019776R, PKK019780R, PKK019785R, PKK019787R, PKK019788R, PKK019789R, PKK019790R, PKK019793R, PKK019795R, PKK019798R, PKK019799R, PKK019800R, PKK019801R, PKK019802R, PKK019803R, PKK019806R, PKK019808R, PKK019809R, PKK019810R, PKK019811R, PKK019812R, PKK019813R, PKK019815R, PKK019816R, PKK019817R, PKK019818R, PKK019820R, PKK019821R, PKK019822R, PKK019824R, PKK019825R, PKK019827R, PKK019828R, PKK019830R, PKK019832R, PKK019834R, PKK019835R, PKK019836R, PKK019841R, PKK019842R, PKK019843R, PKK019844R, PKK019845R, PKK019846R, PKK019849R, PKK019851R, PKK019852R, PKK019853R, PKK019855R, PKK019856R, PKK019857R, PKK019859R, PKK019860R, PKK019862R, PKK019863R, PKK019864R, PKK019866R, PKK019867R, PKK019869R, PKK019871R, PKK019872R, PKK019873R, PKK019874R, PKK019876R, PKK019877R, PKK019878R, PKK019879R, PKK019880R, PKK019882R, PKK019889R, PKK019909R, PKK019910R, PKK019920R, PKK019929R, PKK019930R, PKK019939R, PKK019940R, PKK019944R, PKK019945R, PKK019947R, PKK019950R, PKK019951R, PKK020083R, PKK020084R, PKK020085R, PKK020092R, PKK020099R, PKK020101R, PKK020103R, PKK020108R, PKK020115R, PKK020117R, PKK020120R, PKK020129R, PKK020138R, PKK020139R, PKK020146R, PKK020147R, PKK020152R, PKK020153R, PKK020165R, PKK020168R, PKK020169R, PKK020178R, PKK020182R, PKK020184R, PKK020185R, PKK020192R, PKK020194R, PKK020195R, PKK020197R, PKK020198R, PKK020199R, PKK020200R, PKK020201R, PKK020202R, PKK020203R, PKK020204R, PKK020205R, PKK020206R, PKK020207R, PKK020208R, PKK020209R, PKK020210R, PKK020211R, PKK020215R, PKK020247R, PKK020251R, PKK020256R, PKK020259R, PKK020262R, PKK020263R, PKK020268R, PKK020278R, PKK020282R, PKK020290R, PKK020292R, PKK020294R, PKK020304R, PKK020316R, PKK020323R, PKK020329R, PKK020330R, PKK020338R, PKK020344R, PKK020349R, PKK020359R, PKK020361R, PKK020362R, PKK020380R, PKK020404R, PKK020549R, PKK020550R, PKK020551R, PKK020552R, PKK020564R, PKK020577R, PKK020582R, PKK020587R, PKK020589R, PKK020590R, PKK020593R, PKK020594R, PKK020601R, PKK020602R, PKK020603R, PKK020605R, PKK020606R, PKK020607R, PKK020608R, PKK020609R, PKK020610R, PKK020611R, PKK020612R, PKK020613R, PKK020614R, PKK020616R, PKK020617R, PKK020618R, PKK020619R, PKK020620R, PKK020621R, PKK020622R, PKK020623R, PKK020624R, PKK020625R, PKK020626R, PKK020627R, PKK020628R, PKK020629R, PKK020630R, PKK020631R, PKK020632R, PKK020633R, PKK020634R, PKK020635R, PKK020637R, PKK020638R, PKK020639R, PKK020640R, PKK020641R, PKK020642R, PKK020643R, PKK020644R, PKK020645R, PKK020646R, PKK020647R, PKK020648R, PKK020650R, PKK020651R, PKK020652R, PKK020653R, PKK020654R, PKK020655R, PKK020656R, PKK020657R, PKK020658R, PKK020659R, PKK020660R, PKK020661R, PKK020662R, PKK020663R, PKK020664R, PKK020665R, PKK020666R, PKK020667R, PKK020668R, PKK020669R, PKK020670R, PKK020672R, PKK020673R, PKK020674R, PKK020675R, PKK020676R, PKK020677R, PKK020678R, PKK020679R, PKK020680R, PKK020681R, PKK020684R, PKK020685R, PKK020687R, PKK020688R, PKK020690R, PKK020691R, PKK020693R, PKK020694R, PKK020695R, PKK020696R, PKK020697R, PKK020698R, PKK020699R, PKK020700R, PKK020701R, PKK020702R, PKK020703R, PKK020704R, PKK020705R, PKK020707R, PKK020708R, PKK020709R, PKK020710R, PKK020712R, PKK020713R, PKK020716R, PKK020719R, PKK020722R, PKK020723R, PKK020726R, PKK020727R, PKK020728R, PKK020729R, PKK020731R, PKK020732R, PKK020734R, PKK020735R, PKK020736R, PKK020738R, PKK020740R, PKK020742R, PKK020744R, PKK020745R, PKK020746R, PKK020747R, PKK020748R, PKK020749R, PKK020750R, PKK020752R, PKK020754R, PKK020755R, PKK020756R, PKK020757R, PKK020758R, PKK020759R, PKK020760R, PKK020761R, PKK020762R, PKK020763R, PKK020764R, PKK020765R, PKK020766R, PKK020767R, PKK020768R, PKK020769R, PKK020770R, PKK020771R, PKK020772R, PKK020773R, PKK020774R, PKK020775R, PKK020776R, PKK020779R, PKK020780R, PKK020781R, PKK020782R, PKK020783R, PKK020784R, PKK020785R, PKK020786R, PKK020787R, PKK020788R, PKK020789R, PKK020790R, PKK020791R, PKK020792R, PKK020793R, PKK020794R, PKK020795R, PKK020796R, PKK020797R, PKK020798R, PKK020799R, PKK020800R, PKK020801R, PKK020802R, PKK020803R, PKK020804R, PKK020805R, PKK020806R, PKK020807R, PKK020808R, PKK020810R, PKK020811R, PKK020812R, PKK020813R, PKK020814R, PKK020815R, PKK020816R, PKK020817R, PKK020818R, PKK020819R, PKK020820R, PKK020821R, PKK020822R, PKK050311R, PKK050393R, PKK050558R, PKK050581R, PKK050780R, PKK051047R, PKK051127R, PKK051134R, PKK051137R, PKK051138R, PKK051155R, PKK051166R, PKK051174R, PKK051180R, PKK051181R, PKK051191R, PKK051192R, PKK051194R, PKK051201R, PKK051204R, PKK051208R, PKK051209R, PKK051210R, PKK051211R, PKK051213R, PKK051214R, PKK051216R, PKK051218R, PKK051219R, PKK051220R, PKK051221R, PKK051233R, PKK051250R, PKK051331R, PKK051373R, PKK051389R, PKK051400R, PKK051420R, PKK051432R, PKK051433R, PKK051436R, PKK051444R, PKK051445R, PKK051447R, PKK051448R, PKK051449R, PKK051450R, PKK051452R, PKK051454R, PKK051455R, PKK051456R, PKK051457R, PKK051461R, PKK051462R, PKK051463R, PKK051464R, PKK051465R, PKK051466R, PKK051467R, PKK051468R, PKK051469R, PKK051471R, PKK051473R, PKK051474R, PKK051475R, PKK051476R, PKK051477R, PKK051479R, PKK051480R, PKK051484R, PKK051485R, PKK051486R, PKK051487R, PKK051488R, PKK051491R, PKK051492R, PKK051493R, PKK051494R, PKK051495R, PKK051496R, PKK051497R, PKK051498R, PKK051499R, PKK051500R, PKK051501R, PKK051502R, PKK051503R, PKK051504R, PKK051505R, PKK051507R, PKK051509R, PKK051510R, PKK051512R, PKK051513R, PKK051517R, PKK051518R, PKK051520R, PKK051522R, PKK051523R, PKK051524R, PKK051525R, PKK051526R, PKK051527R, PKK051529R, PKK051530R, PKK051532R, PKK051536R, PKK051537R, PKK051538R, PKK051539R, PKK051540R, PKK051542R, PKK051544R, PKK051545R, PKK051547R, PKK051548R, PKK051550R, PKK051551R, PKK051552R, PKK051553R, PKK051554R, PKK051555R, PKK051556R, PKK051558R, PKK051560R, PKK051563R, PKK051564R, PKK051565R, PKK051568R, PKK051569R, PKK051572R, PKK051573R, PKK051575R, PKK051576R, PKK051581R, PKK051582R, PKK051588R, PKK051589R, PKK051593R, PKK051623R, PKK051628R, PKK052166R, PKK052251R, PKK052253R, PKK052264R, PKK052278R, PKK052284R, PKK052285R, PKK052286R, PKK052292R, PKK052293R, PKK052477R, PKK052479R, PKK052484R, PKK052487R, PKK052493R, PKK052533R, PKK052788R, PKK052789R, PKK052792R, PKK052797R, PKK052798R, PKK052817R, PKK052818R, PKK052819R, PKK052820R, PKK052822R, PKK052826R, PKK052833R, PKK052834R, PKK052835R, PKK052949R, PKK053059R, PKK053062R, PKK053086R, PKK053109R, PKK053117R, PKK053118R, PKK053123R, PKK053126R, PKK053130R, PKK053137R, PKK053140R, PKK053144R, PKK053166R, PKK053171R, PKK053174R, PKK053240R, PKK053246R, PKK053563R, PKK053564R, PKK053608R, PKK053698R, PKK053748R, PKK053782R, PKK053906R, PKK053974R, PKK054222R, PKK054299R, PKK054333R, PKK055139R, PKK055190R, PKK055223R, PKK055260R, PKK055446R, PKK055500R, PKK055501R, PKK055502R, PKK055503R, PKK055504R, PKK055505R, PKK055510R, PKK055515R, PKK055631R, PKK055689R, PKK055711R, PKK055999R, PKK056363R, PKK056370R, PKK056372R, PKK056373R, PKK056382R, PKK056396R, PKK056400R, PKK056595R, PKK056598R, PKK056615R, PKK056633R, PKK056636R, PKK056641R, PKK056653R, PKK056664R, PKK056665R, PKK056667R, PKK056669R, PKK056672R, PKK056673R, PKK056675R, PKK056676R, PKK056678R, PKK056680R, PKK056682R, PKK056684R, PKK056685R, PKK056688R, PKK056689R, PKK056690R, PKK056691R, PKK056692R, PKK056693R, PKK056694R, PKK056698R, PKK056699R, PKK056703R, PKK056706R, PKK056708R, PKK056709R, PKK056710R, PKK056713R, PKK056714R, PKK056716R, PKK057042R, PKK057049R, PKK057050R, PKK057051R, PKK057052R, PKK057053R, PKK057054R, PKK057055R, PKK057056R, PKK057058R, PKK057059R, PKK057061R, PKK057062R, PKK057063R, PKK057064R, PKK057065R, PKK057066R, PKK057067R, PKK057068R, PKK057069R, PKK057072R, PKK057073R, PKK057074R, PKK057075R, PKK057076R, PKK057077R, PKK057078R, PKK057079R, PKK057080R, PKK057086R, PKK057091R, PKK057105R, PKK057108R, PKK057109R, PKK057110R, PKK057111R, PKK057112R, PKK057113R, PKK057114R, PKK057116R, PKK057117R, PKK057119R, PKK057121R, PKK057122R, PKK057123R, PKK057124R, PKK057125R, PKK057126R, PKK057127R, PKK057128R, PKK057129R, PKK057130R, PKK057131R, PKK057132R, PKK057133R, PKK057134R, PKK057136R, PKK057139R, PKK057140R, PKK057141R, PKK057143R, PKK057144R, PKK057148R, PKK057149R, PKK057153R, PKK057158R, PKK057161R, PKK057171R, PKK057174R, PKK057175R, PKK057177R, PKK057178R, PKK057179R, PKK057180R, PKK057181R, PKK057182R, PKK057183R, PKK057184R, PKK057185R, PKK057186R, PKK057187R, PKK057189R, PKK057190R, PKK057191R, PKK057194R, PKK057195R, PKK057197R, PKK057198R, PKK057200R, PKK057201R, PKK057202R, PKK057203R, PKK057204R, PKK057205R, PKK057206R, PKK057208R, PKK057209R, PKK057210R, PKK057211R, PKK057212R, PKK057213R, PKK057214R, PKK057215R, PKK057216R, PKK057217R, PKK057218R, PKK057219R, PKK057220R, PKK057221R, PKK057222R, PKK057223R, PKK057224R, PKK057225R, PKK057226R, PKK057227R, PKK057228R, PKK057229R, PKK057230R, PKK057232R, PKK057233R, PKK057234R, PKK057236R, PKK057237R, PKK057239R, PKK101143R, PKK102460R, PKK102461R, PKK102464R, PKK102478R, PKK102482R, PKK102485R, PKK102486R, PKK102487R, PKK103561R, PKK103586R, PKK104160R, PKK105556R, PKK105920R, PKK106546R, PKK106596R, PKK106665R, PKK107715R, PKK107855R, PKK107982R, PKK107984R, PKK107989R, PKK107991R, PKK107995R, PKK108008R, PKK108009R, PKK108011R, PKK108016R, PKK108019R, PKK108021R, PKK108024R, PKK108025R, PKK108029R, PKK108039R, PKK108040R, PKK108043R, PKK108046R, PKK108047R, PKK108048R, PKK108049R, PKK108057R, PKK108058R, PKK108060R, PKK108062R, PKK108063R, PKK108069R, PKK108083R, PKK108085R, PKK108117R, PKK108121R, PKK108137R, PKK108138R, PKK108139R, PKK108140R, PKK108142R, PKK108143R, PKK108146R, PKK108148R, PKK108152R, PKK108153R, PKK108156R, PKK108157R, PKK108158R, PKK108159R, PKK108163R, PKK108164R, PKK108165R, PKK108167R, PKK108168R, PKK108172R, PKK108173R, PKK108176R, PKK108178R, PKK108180R, PKK108333R, PKK108370R, PKK108391R, PKK108397R, PKK108404R, PKK108406R, PKK108424R, PKK108427R, PKK108428R, PKK108431R, PKK108432R, PKK108433R, PKK108436R, PKK108441R, PKK108444R, PKK108448R, PKK108462R, PKK108463R, PKK108468R, PKK108856R, PKK108875R, PKK108998R, PKK109018R, PKK109022R, PKK109026R, PKK109040R, PKK109048R, PKK109054R, PKK109066R, PKK109077R, PKK109079R, PKK109081R, PKK109083R, PKK109084R, PKK109088R, PKK109095R, PKK109097R, PKK109098R, PKK109099R, PKK109100R, PKK109655R, PKK109687R, PKK109703R, PKK109716R, PKK109717R, PKK109721R, PKK109723R, PKK109724R, PKK109726R, PKK109729R, PKK109731R, PKK109732R, PKK109733R, PKK109734R, PKK109735R, PKK109736R, PKK109737R, PKK109738R, PKK109740R, PKK109741R, PKK109745R, PKK109746R, PKK109747R, PKK109748R, PKK109750R, PKK109752R, PKK109753R, PKK109754R, PKK109755R, PKK109756R, PKK109757R, PKK109758R, PKK109759R, PKK109761R, PKK109762R, PKK109763R, PKK109764R, PKK109765R, PKK109768R, PKK109770R, PKK109771R, PKK109772R, PKK109773R, PKK109774R, PKK109776R, PKK202624R, PKK202680R, PKK202685R, PKK202686R, PKK202687R, PKK202688R, PKK202690R, PKK202694R, PKK202698R, PKK202701R, PKK202705R, PKK202712R, PKK202715R, PKK202716R, PKK202719R, PKK202720R, PKK202721R, PKK202727R, PKK202729R, PKK202730R, PKK202737R, PKK202740R, PKK202742R, PKK202743R, PKK202744R, PKK202748R, PKK202749R, PKK202752R, PKK202754R, PKK202755R, PKK202757R, PKK202759R, PKK202760R, PKK202761R, PKK202762R, PKK202764R, PKK202765R, PKK202766R, PKK202767R, PKK202768R, PKK202769R, PKK202771R, PKK202772R, PKK202773R, PKK202775R, PKK202777R, PKK202779R, PKK202780R, PKK202781R, PKK202783R, PKK202784R, PKK202785R, PKK202787R, PKK202791R, PKK202792R, PKK202794R, PKK202796R, PKK202798R, PKK202815R, PKK202821R, PKK202836R, PKK202840R, PKK202842R, PKK202845R, PKK202846R, PKK202847R, PKK202848R, PKK202849R, PKK202850R, PKK202851R, PKK202852R, PKK202855R, PKK202856R, PKK202857R, PKK202858R, PKK202859R, PKK202860R, PKK202861R, PKK202864R, PKK202868R, PKK202870R, PKK202871R, PKK202872R, PKK202873R, PKK202875R, PKK202877R, PKK202878R, PKK202879R, PKK202883R, PKK202887R, PKK202888R, PKK202890R, PKK202892R, PKK202895R, PKK202897R, PKK202898R, PKK202899R, PKK202900R, PKK202901R, PKK202904R, PKK202905R, PKK202906R, PKK202908R, PKK202909R, PKK202910R, PKK202911R, PKK202912R, PKK202913R, PKK202916R, PKK202917R, PKK202919R, PKK202924R, PKK202926R, PKK202927R, PKK202928R, PKK202929R, PKK202938R, PKK202951R, PKK202970R, PKK202981R, PKK202985R, PKK202988R, PKK203165R, PKK203166R, PKK203168R, PKK203173R, PKK203176R, PKK203178R, PKK203181R, PKK203183R, PKK203184R, PKK203187R, PKK203197R, PKK203201R, PKK203207R, PKK203211R, PKK203212R, PKK203213R, PKK203223R, PKK203224R, PKK203228R, PKK203396R, PKK203400R, PKK203402R, PKK203403R, PKK203404R, PKK203405R, PKK203407R, PKK203409R, PKK203411R, PKK203412R, PKK203416R, PKK203417R, PKK203420R, PKK203426R, PKK203427R, PKK203434R, PKK203436R, PKK203439R, PKK203441R, PKK203442R, PKK203444R, PKK203448R, PKK203452R, PKK203456R, PKK203462R, PKK203464R, PKK203469R, PKK203471R, PKK203472R, PKK203478R, PKK203479R, PKK203481R, PKK203483R, PKK203485R, PKK203488R, PKK203489R, PKK203491R, PKK203492R, PKK203493R, PKK203495R, PKK203496R, PKK203497R, PKK203498R, PKK203499R, PKK203500R, PKK203502R, PKK203503R, PKK203504R, PKK203506R, PKK203508R, PKK203511R, PKK203516R, PKK203517R, PKK203518R, PKK203520R, PKK203521R, PKK203522R, PKK203523R, PKK203525R, PKK203526R, PKK203527R, PKK203528R, PKK203529R, PKK203530R, PKK203533R, PKK203535R, PKK203541R, PKK203545R, PKK203546R, PKK203547R, PKK203549R, PKK203551R, PKK203552R, PKK203554R, PKK203555R, PKK203558R, PKK203559R, PKK203560R, PKK203561R, PKK203562R, PKK203563R, PKK203564R, PKK203565R, PKK203566R, PKK203567R, PKK203571R, PKK203572R, PKK203573R, PKK203574R, PKK203575R, PKK203576R, PKK203577R, PKK203578R, PKK203579R, PKK203580R, PKK203581R, PKK203582R, PKK203583R, PKK203585R, PKK203591R, PKK203597R, PKK203599R, PKK203600R, PKK203605R, PKK203610R, PKK203611R, PKK203612R, PKK203614R, PKK203615R, PKK203617R, PKK203621R, PKK203622R, PKK203623R, PKK203624R, PKK203625R, PKK203626R, PKK203627R, PKK203628R, PKK203629R, PKK203630R, PKK203631R, PKK203632R, PKK203633R, PKK203634R, PKK203635R, PKK203745R, PKK203746R, PKK203747R, PKK203748R, PKK203749R, PKK203750R, PKK203751R, PKK203753R, PKK203754R, PKK203755R, PKK203756R, PKK203757R, PKK203759R, PKK203760R, PKK203761R, PKK203763R, PKK203764R, PKK203767R, PKK203768R, PKK203769R, PKK203770R, PKK203771R, PKK203772R, PKK203773R, PKK203774R, PKK203775R, PKK203776R, PKK203777R, PKK203778R, PKK203779R, PKK203780R, PKK203781R, PKK203782R, PKK203783R, PKK203784R, PKK203785R, PKK203786R, PKK203787R, PKK203788R, PKK203789R, PKK203791R, PKK203793R, PKK203794R, PKK203795R, PKK203796R, PKK203797R, PKK203798R, PKK203800R, PKK203801R, PKK203804R, PKK203806R, PKK203807R, PKK203808R, PKK203809R, PKK203811R, PKK203812R, PKK203813R, PKK203814R, PKK203815R, PKK203816R, PKK203818R, PKK203819R, PKK203820R, PKK203821R, PKK203822R, PKK203823R, PKK203825R, PKK203826R, PKK203828R, PKK203829R, PKK203831R, PKK203832R, PKK203833R, PKK203834R, PKK203835R, PKK203837R, PKK203838R, PKK203839R, PKK203840R, PKK203841R, PKK203842R, PKK203843R, PKK203844R, PKK203845R, PKK203846R, PKK203847R, PKK203849R, PKK203850R, PKK203851R, PKK203852R, PKK203853R, PKK203854R, PKK203856R, PKK203857R, PKK203858R, PKK203859R, PKK203860R, PKK203861R, PKK203862R, PKK203863R, PKK203864R, PKK203865R, PKK203866R, PKK203867R, PKK203868R, PKK203869R, PKK203870R, PKK203872R, PKK203873R, PKK203874R, PKK203875R, PKK203876R, PKK203877R, PKK204230R, PKK204234R, PKK204240R, PKK204258R, PKK204260R, PKK204261R, PKK204264R, PKK204267R, PKK204269R, PKK204277R, PKK204288R, PKK204289R, PKK204293R, PKK204299R, PKK204300R, PKK204301R, PKK204369R, PKK204474R, PKK204477R, PKK204479R, PKK204480R, PKK204483R, PKK204486R, PKK204491R, PKK204494R, PKK204495R, PKK204499R, PKK204504R, PKK204511R, PKK204513R, PKK204514R, PKK204515R, PKK204523R, PKK204538R, PKK204548R, PKK204549R, PKK204550R, PKK204551R, PKK204556R, PKK204558R, PKK204559R, PKK204561R, PKK204562R, PKK204563R, PKK204564R, PKK204565R, PKK204566R, PKK204567R, PKK204568R, PKK204570R, PKK204571R, PKK204572R, PKK204573R, PKK204574R, PKK204575R, PKK204576R, PKK204577R, PKK204578R, PKK204579R, PKK204580R, PKK204581R, PKK204582R, PKK204583R, PKK204584R, PKK204585R, PKK204587R, PKK204588R, PKK204589R, PKK204590R, PKK204591R, PKK204592R, PKK204593R, PKK204594R, PKK204595R, PKK204596R, PKK204597R, PKK204598R, PKK204599R, PKK204600R, PKK204601R, PKK204602R, PKK204603R, PKK204604R, PKK204605R, PKK204606R, PKK204607R, PKK204608R, PKK204610R, PKK204611R, PKK204612R, PKK204614R, PKK204615R, PKK204616R, PKK204620R, PKK204624R, PKK204627R, PKK204633R, PKK204634R, PKK204635R, PKK204637R, PKK204638R, PKK204640R, PKK204641R, PKK204646R, PKK204647R, PKK204648R, PKK204649R, PKK204652R, PKK204657R, PKK204658R, PKK204659R, PKK204660R, PKK204664R, PKK204679R, PKK204680R, PKK204684R, PKK204685R, PKK204690R, PKK204691R, PKK204694R, PKK204696R, PKK204697R, PKK204698R, PKK204699R, PKK204700R, PKK204701R, PKK204704R, PKK204705R, PKK204706R, PKK204707R, PKK204708R, PKK204710R, PKK204712R, PKK204713R, PKK204717R, PKK204718R, PKK204719R, PKK204720R, PKK204721R, PKK204722R, PKK204725R, PKK204726R, PKK204727R, PKK204729R, PKK204730R, PKK204731R, PKK204732R, PKK204733R, PKK204734R, PKK204735R, PKK204736R, PKK204738R, PKK204739R, PKK204740R, PKK204741R, PKK204742R, PKK204745R, PKK204747R, PKK204748R, PKK204750R, PKK204751R, PKK204752R, PKK204753R, PKK204754R, PKK204755R, PKK204757R, PKK204758R, PKK204759R, PKK204761R, PKK204762R, PKK204764R, PKK204765R, PKK204766R, PKK204767R, PKK204768R, PKK204769R, PKK204770R, PKK204772R, PKK204773R, PKK204774R, PKK204775R, PKK204776R, PKK204777R, PKK204778R, PKK204779R, PKK204780R, PKK204783R, PKK204784R, PKK204785R, PKK204787R, PKK204789R, PKK204790R, PKK204791R, PKK204792R, PKK204793R, PKK204794R, PKK204795R, PKK204796R, PKK204798R, PKK204799R, PKK204800R, PKK204801R, PKK204802R, PKK204804R, PKK204806R, PKK204808R, PKK204809R 数量 2,462台 出荷時期 平成19年7月4日から令和2年8月28日
製造販売業者等名称
製造販売業者の名称 : 日本メドトロニック株式会社 製造販売業者の所在地: 東京都港区港南一丁目2番70号 許可の種類 : 第一種医療機器製造販売業 許可番号 : 13B1X00261
改修理由
海外製造元より、対象製品を使用して、メドトロニック社製の植込み能動型機器(以下、「植込み機器」 という。) であるMedtronic Azure MRIシリーズ(医療機器承認番号:23000BZX00027000)、 Percepta MRI CRT-Pシリーズ(医療機器承認番号:23000BZX00230000)およびSerena MRI CRT-Pシリーズ (医療機器承認番号:23000BZX00139000)と通信を行った際に、実際の製品寿命よりも予想寿命が 長く表示されるとの連絡を受領しました。 この事象は、心房リードと右心室リードのどちらか一方、もしくは両方をユニポーラ設定で使用されている 場合において、バイポーラのリードインピーダンス値を用いて予想寿命を算出することにより発生します。 このため、対象製品のソフトウェアのアップデートにより本事象を修正する自主改修を実施します。
危惧される具体的な健康被害
ユニポーラ設定に関連する予想寿命の算出エラーは、製品寿命の前期にのみ発生する事象です。 植込み機器の機能や推奨交換時期(RRT)、選択的交換指標(ERI)および電池寿命終了(EOS)は本事象による 影響を受けません。 植込み機器を植え込んだ後は少なくとも3〜4か月ごとに、対象製品を用いた対面もしくは遠隔モニタリング にてフォローアップを実施し、電池の消耗、作動状況、患者診断情報等を確認すること、また交換時期は、 予想残存寿命ではなく推奨交換時期(RRT)に基づき判断することと添付文書にて注意喚起を行っております。 そのため、本事象が発生した場合でも植込み機器の交換時期の検討は適切に行われることから、重篤な健康 被害が発生する事はまず考えられません。 なお、2021年3月5日時点で、国内外において本事象に起因する不可逆的な健康被害は報告されておりません。
改修開始年月日
令和3年4月7日 情報提供の開始
効能・効果又は用途等
(1)ケアリンク SmartSync Device Manager 本品は、併用可能なメドトロニック社製の植込み機器をインテロゲート及びプログラムするために用います。 また、本品は、植込み機器の植え込み時におけるリードシステムのペーシング及びセンシング性能を評価する ために用います。 さらに、本品は、併用可能な植込み機器において承認された機能に限り、植込み機器に対して非侵襲的に外部 からその機能を追加することもできます。 (2)ケアリンク 2090W 本品は、メドトロニック社製の植込み型心臓ペースメーカ、除細動機能付き植込み型両心室 ペーシングパルスジェネレータ、除細動機能なし植込み型両心室ペーシングパルスジェネレータ、 自動植込み型除細動器、デュアルチャンバ植込み型除細動器、植込み型心電用データレコーダ、 並びにビタトロン社製の植込み型心臓ペースメーカの1つ以上の動作モード、パラメータ等を 非侵襲的に外部からコントロールすることにより、植込み時及び植込み後のペーシング状態の 測定又はプログラムを行うことを目的とします。 また、上記の植込み型機器において承認を取得した機能に限り、本品は植込み型機器に非侵襲的に 外部から追加することもできます。
その他
対象製品および対象製品を使用した場合に不正確な予想寿命が表示される植込み機器を使用中の医療機関等は 全て把握しておりますので、該当する医療機関および代理店に対して情報提供の上、ソフトウェアの供給が 可能となりましたら改修を実施します。 (1)ケアリンク SmartSync Device Manager インターネット回線を通じて修正されたソフトウェアを対象製品に配信し、改修を行います。 (2)ケアリンク 2090W 対象製品のソフトウェアをアップデートし、改修を行います。
担当者及び連絡先
担当者 : 品質・市販後安全管理 嶋田 稔之/高城 義雄(高はハシゴダカ) 連絡先 : 日本メドトロニック株式会社 東京都港区港南一丁目2番70号 電話番号: 03-6776-0040 FAX番号 : 03-6774-4675